Class 2 Recall
CONTAK RENEWAL 4 AVT

See Related Information
Date Recall
Initiated
March 23, 2007
Date Posted July 11, 2007
Recall Number Z-1051-2007
Product These models are not available in the US. CONTACT RENEWAL 4 AVT HE (model number M170, M175, M177, M179). Cardiac Resynchronization Therapy Defibrillator. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.
Code Information Model M170, serial numbers: 100136, 100138, 100142, 100144, 100145, 200003, 200004, 200006, 200007, 200009, 200010, 200011, 200013, 200017, 200018, 200019, 200021, 200022, 200023, 200025, 200027, 200028, 200029, 200030, 200032, 200033, 200034, 200035, 200036, 200037, 200038, 200039, 200041, 200042, 200043, 200046, 200047, 200048, 200051, 200052, 200053, 200054, 200056, 200057, 200058, 200059, 200060, 200073, 200074, 200077, 200079, 200082, 200084, 200087, 200088, 200089, 200090, 200093, 200094, 200095, 200098, 200101, 200112. Model M175, serial numbers: 100202, 100207, 100209, 100210, 100211, 100212, 100213, 100214, 100215, 100216, 100219, 100221, 100222, 100224, 100251, 100277, 100282, 100283, 100284, 100298, 100299, 100336, 100341, 100344, 100345, 100349, 100350, 100353, 100354, 100356, 100358, 100361, 100362, 100369, 100370, 100372, 100384, 100401, 100404, 100407, 100409, 100410, 100411, 100418, 100421, 100433, 200002, 200003, 200004, 200005, 200007, 200009, 200010, 200011, 200012, 200014, 200015, 200016, 200017, 200019, 200022, 200025, 200027, 200028, 200029, 200030, 200032, 200033, 200034, 200036, 200038, 200039, 200040, 200042, 200044, 200047, 200048, 200051, 200052, 200053, 200054, 200056, 200057, 200058, 200060, 200062, 200067, 200071, 200074, 200075, 200079, 200080, 200081, 200083, 200084, 200085, 200086, 200088, 200089, 200091, 200092, 200093, 200094, 200095, 200108, 200112, 200113, 200115, 200124, 200143, 200153, 200156, 200157, 200158, 200162, 200163, 200166. Model M177, serial numbers: 100154, 100157, 100158, 100168, 100169, 100173, 100174, 100180, 100181, 100182, 100183, 100190, 100191, 100196, 100197, 100221, 200003, 200004, 200005, 200006, 200007, 200008, 200009, 200010, 200011, 200012, 200013, 200014, 200015, 200016, 200017, 200018, 200020, 200021, 200022, 200023, 200024, 200025, 200026, 200027, 200028, 200029, 200030, 200032, 200033, 200034, 200035, 200036, 200037, 200038, 200039, 200040, 200041, 200042, 200044, 200045, 200046, 200047, 200048, 200049, 200052, 200053, 200054, 200055, 200056, 200057, 200058, 200059, 200060, 200062, 200066, 200067, 200068, 200069, 200072, 200073, 200074, 200076, 200078, 200080, 200081, 200083, 200084, 200085, 200086, 200087, 200089, 200090, 200091, 200092, 200093, 200095, 200096, 200097, 200099, 200102, 200104, 200105, 200106, 200107, 200108, 200109, 200110, 200111, 200112, 200113, 200114, 200115, 200116, 200117, 200118, 200119, 200120, 200121, 200122, 200123, 200124, 200125, 200126, 200128, 200129, 200130, 200131, 200134, 200135, 200136, 200140, 200141, 200142, 200143, 200145, 200146, 200147, 200148, 200149, 200150, 200152, 200153, 200154, 200155, 200156, 200158, 200160, 200161, 200164, 200165, 200166, 200167, 200168, 200169, 200170, 200171, 200172, 200175, 200177, 200178, 200179, 200180, 200181, 200182. Model M179, serial numbers: 100216, 100217, 100219, 100223, 100227, 100228, 100229, 100230, 100231, 100232, 100233, 100235, 100237, 100238, 100239, 100240, 100243, 100244, 100247, 100250, 100253, 100256, 100257, 100263, 100266, 100270, 100271, 100272, 100274, 100275, 100276, 100278, 100279, 100280, 100281, 100282, 100283, 100285, 100287, 100290, 100291, 100293, 100294, 100295, 100297, 100298, 100300, 100302, 100303, 100304, 100305, 100315, 100316, 100317, 100319, 100321, 100322, 100323, 100324, 100325, 100326, 100327, 100338, 100342, 100343, 100349, 100353, 100354, 100391, 100430, 100435, 100436, 100437, 100439, 100440, 100441, 100442, 100448, 100449, 100450, 100452, 100467, 100469, 100470, 100473, 100475, 100479, 100480, 100482, 100488, 100489, 100500, 100502, 100503, 100515, 100516, 100517, 100520, 100522, 100533, 100543, 100545, 100547, 100549, 100552, 100555, 100556, 100557, 100558, 100563, 100566, 100567, 100568, 100577, 100579, 100581, 100584, 100587, 100589, 100590, 100591, 100614, 200002, 200003, 200004, 200005, 200006, 200007, 200008, 200009, 200010, 200013, 200014, 200015, 200016, 200017, 200018, 200019, 200020, 200021, 200022, 200023, 200024, 200026, 200027, 200028, 200029, 200030, 200031, 200032, 200035, 200036, 200037, 200039, 200040, 200042, 200043, 200044, 200046, 200047, 200049, 200051, 200052, 200053, 200054, 200055, 200056, 200058, 200059, 200060, 200061, 200062, 200063, 200064, 200065, 200066, 200068, 200070, 200071, 200072, 200075, 200077, 200078, 200079, 200080, 200081, 200083, 200084, 200086, 200088, 200089, 200090, 200092, 200093, 200094, 200096, 200097, 200100, 200101.
Recalling Firm/
Manufacturer
Boston Scientific CRM Corp
4100 Hamline Ave N
Saint Paul , Minnesota 55112-5700
Reason For
Recall
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective replacement indicator (ERI) and end of life (EOL) to less than 3 months.
Action An Urgent Medical Devivce Information letter dated April 05, 2007 was sent to implanting and following physicians. The letter describes the issue, identifies patients with affected devices and provides recommendations.
Quantity in Commerce 539
Distribution Worldwide, including USA, Argentina, Aruba, Australia, Austria, Barbados, Belgium, Bermuda, Brazil, Canada, Chile, Costa Rica, Cyprus, Czech Republic, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guam, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Pakistan, Polynesia, Portugal, Puerto Rico, Qatar, Saint Lucia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan R O C, Thailand, Tunisia, Turkey, U.S. Virgin Islands, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, and West Indies.
 