Class 2 Recall
Jostra

See Related Information
Date Recall
Initiated
October 03, 2006
Date Posted November 07, 2006
Recall Number Z-0159-2007
Product Jostra HL-20 Heart Lung Machine - guiding pins
Code Information All HL 20 systems Serial #s 1159,14031794,14031795, 14031796,14031797,14031798, 100155, 1345, 1363,1588, 14031765, 14031766, 1570, 1571,1593,1319,1320,1321,1322,14031774,14031775, 14031776, 1323,1324,1478,1479,14031067,14031069,10331385,1597,14021730,14031792,1459,1701,1512,1513,100035, 100036, 1034,1035, 14031045, 14031046, 14031047, 14031048, 1022,1023,1541,1559,100094,1025,1026,1032,1331,1332, 100011,100015,100016,100009,100063, 1668,1511, 1652, 1702,1454,1455, 1379,1380, 1381, 1272, 1417,1480, 1500, 1653, 14031054, 14031055,14031059, 14041074, 14041874, 1128, 1592, 100030, 100031, 100012, 100013, 1595, 1596, 14031039, 1510, 1105, 1433, 1434, 1456, 1457, 1577, 1578, 1579, 14021718, 14021719, 14041904, 100026, 14021706, 1420, 1421, 1398, 1418, 1598, J14041874,1584, 100108, 100109,14041074, JOSTRA, 1132, 100009, 100010,100130,100131, 1625, 1626, 1627, 1628, 1629, 100135, 100136, 100137, 1199, 1200, 1540, 1552, 1567, 1568, 1569, 1580, 1581, 1584, 1585, 1586, 14031040, 14031068, 14031761, 14031762, 1716, 100084, 14021716, 14041882, 100188, 1562, 1563, 100028, 100023, 14031831, 14031793, 1019, 1020, 1021, 1295, 100021, 100022, 14031777, 14031778, 14031779, 1523, 1524, 1017, 1442, 1443, 1465, 1466, 1635, 100116, 14031051, 14031052, 1507, 1508, 1509, 14031732, 100001, 100002, 100061, 100062, 14031734, 14031735, 14031736, 100017, 1483, 1481, 1482, 14031056, 14031057, 1535,1536, 1537, 1538, 1539, 1547, 1548, 1549, 1550, 1551, 14031060, 14031061, 14031062, 14031063, 100219,100220, 100221, 100222, 100223, 100224, 100225, 100226, 100227, 100228, 1642, 1643, 1494, 1666, 1667, 1618, 1619, 1517, 1518, 100027, 1345
Recalling Firm/
Manufacturer
MAQUET Inc.
1140 US Highway 22
Bridgewater , New Jersey 08807-2958
For Addition Information Contact Jamie Yieh
908-947-2311
Reason For
Recall
The HL-20's pump head on the roller pump has four pairs of tube guide rolls that keep the tube in place in the raceway and roll along the inserted tube as the pump head turns. If dirt and other particles accumulate in the tube guide rolls, they cannot roll smoothly and finally they can get stuck.
Action An 'Important Safety Notice - Please Read Immediately' letter was sent to all US customers on 10/3/06 by FedEx Return Receipt.
Quantity in Commerce 227 HL-20 base units with 545 single pumps
Distribution Worldwide
 