Skip NavigationFDA Logo links to FDA home pageCenter for Devices and Radiological Health, U.S. Food and Drug AdministrationHHS Logo links to Department of Health and Human Services website
FDA Home Page | CDRH Home Page | Search | A-Z Index U.S. Food and Drug AdministrationCenter for Devices and Radiological Health Questions?
horizonal rule
Super Search
510(k) | Registration & Listing | Adverse Events | PMA | Classification | CLIA
CFR Title 21 | Advisory Committees | Assembler | Recalls | Guidance | Standards
 

New Search Back To Search Results
 

Medical Device Recalls
Class 1 Recall
Welch Allyn PIC 50

See Related Information
Date Recall
Initiated
June 30, 2006
Date Posted July 08, 2006
Recall Number Z-1195-06
Product Welch Allyn PIC 50 Portable Intensive Care System Multi-Parameter Monitor/Defibrillator; a portable, 12V internal battery powered defibrillator; Manufactured by MRL Inc., A Welch Allyn Company, 1000 Asbury Drive, Buffalo Grove, Illinois 60089
Code Information Part numbers 971081, 971082, 971083 and 971084; serial numbers 80104, 80119, 80120, 80204, 80211, 80212, 80429, 80432, 80569, 80627, 80661, 80685, 80880, 80964, 81056, 81134, 81263, 81293/81295, 81303, 81445/ 81448, 81733, 81738, 81739, 81744/81752, 81757, 81761/81781, 81802/81807, 81824, 81825, 81835, 81845, 81846, 81852, 81864, 81865, 81868, 81869, 81884/81886, 81892/81897, 81918/81955, 81957, 81983/81990, 81996/82025, 82027/82033, 82035/82039, 82043, 82044, 82061/82080, 82084, 82097/82111, 82122/82128, 82133, 82136/82143, 82150/82169, 82172/82204, 82245, 82252/82273, 82283/82286, 82307/82330, 82340/82349, 82405/82426, 82430, 82460/82490, 82498/82553, 82555/82565, 82568/82570, 82612/82631, 82654/82673, 82689, 82690, 82773/82813, 82827/82830, 82852, 82856/82859, 82906, 82936, 82937, 82958/82977, 82983/83013, 83060/83071, 83083/83090, 83093, 83142/83147, 83153/83172, 83198/83239, 83255, 83368/83370, 83400/83407, 83414/83418, 83420/83429, 83440/83443, 83499/83517, 83578/83580, 83633, 83644, 83645, 83654, 85001/85005, 85009/85013, 85026/85029, 85032/85052, 85073/85079, 85093, 85094, 85110, 85111, 85143/85157, 85187/85201, 85223, 85224, 85226/85238, 85244/85249, 85329, 85332/85334, 85337/85350, 85364, 85365, 85368/85370, 85381, 85388/85398, 85402/85404, 85410/85414, 85420/85433, 85441, 85505/85515, 85517/85520, 85522/85530, 85534, 85552, 85554/85557, 85587/85589, 85602/85607, 85619/85623, 85630, 85645/85655, 85658/85667, 85686, 85728/85734, 85755/85757, 85761/85766, 85771/85785, 85787/85793, 85810, 85812/85823, 85825/85838, 85856/85872, 85877/85884, 85888/85902, 85931/85933, 85945, 85955, 85964, 85967/85970, 86017, 86018, 86024/86029, 86055/86057, 86103, 86104, 86118, 86141/86147, 86149, 86150, 86222/86228, 86230, 86245, 86256/86258, 86261/86268, 86281/86283, 86289/86293, 86295, 86296, 86299, 86305, 86306, 86314, 86315, 86320/86322, 86324/86328, 86334, 86335, 86358, 86359, 86385, 86387, 86389/86399, 86400, 86411, 86412, 86419, 86420, 86439, 86440
Recalling Firm/
Manufacturer
MRL, Inc., A Welch Allyn Company
1000 Asbury Dr Ste 17
Buffalo Grove , Illinois 60089-4551
For Addition Information Contact Jamie Arnold
315-685-4599
Reason For
Recall
The Welch Allyn PIC 50 Defibrillators may display a 'Defib Comm' or 'Pace Comm' error message on the device display during use which may result in a terminal failure of the device to analyze the patient's ECG and deliver the appropriate therapy.
Action MRL sent Urgent Medical Device Recall letters dated 6/30/06 via certified mail to the direct accounts informing them of an intermittent connection between a pin and socket on a circuit board inside of the device that could result in a malfunction that prevents or unacceptably delays the delivery of a defibrillating shock, which possibly could result in failure to resuscitate the patient. The accounts were provided a list of the affected units' parts/serial numbers and requested to inspect their PIC 50 units for the affected serial numbers and return the identified units for servicing. The accounts were also instructed to respond to the notification within 5 working days if the unit has displayed the Defib Comm error, and within 30 days if it has not displayed the Defib Comm error by fax or mail. A loaner PIC 50 will then be sent to the consignee with instructions on how to return their unit for correction.
Quantity in Commerce 673 units - USA, 511 units - International
Distribution Worldwide-including USA, Puerto Rico, Australia, Belgium, Canada, China, Cyprus, France, Germany, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Malaysia, Poland, Portugal, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sultanate of Oman, Switzerland, United Arab Emirates, United Kingdom and Vietnam.
 

horizonal rule

CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page

Center for Devices and Radiological Health / CDRH