Pt information was sourced by the manufacturer.
Glucose result of 61 mg/dl repeated as 405 mg/dl upon redraw.
The original tube was re-tested with a result of 405 mg/dl.
Instrument precision was evaluated using control material and found to be within label claims.
The customer stated that the sample from the original run may have had a bubble or fibrin and that they felt the instrument was operating as required.
No additional probable cause was found for the discrepant result.
The application sheet for glucose (30-1476/r4/dec98) states that the presence of fibrin may cause erroneous results.
The areoset system operations manual (30-1677/r2/march 99, section 7, operational precautions and limitations, specimens - preparation and storage) also states that for optimal results, specimens should be free of fibrin.
Adequate labeling exists to minimize any associated risk to pt results.
A malfunction did not occur.
This is the final report.
|