FDA

Enforcement Report

The FDA Enforcement Report is published weekly by the Food and Drug Administration, U.S. Public Health Service, Department of Health and Human Services. It contains information on actions taken in connection with agency regulatory activities.
 ENFORCE
03/03/1993

Recalls and Field Corrections:  Foods -- Class II -- 03/03/1993


MARCH 3, 1993                                              93-9


RECALLS AND FIELD CORRECTIONS:  FOODS -- CLASS II
              
PRODUCT        Jalapeno Hot Sauce, in 6 ounce glass bottles.  Recall #F-225-3.
CODE           None on container.
MANUFACTURER   Panola Pepper Corporation, Lake Providence, Louisiana.
RECALLED BY    Manufacturer, by telephone and by fax on or about December 10,   
               1992.  Firm-initiated recall ongoing.
DISTRIBUTION   Virginia, Minnesota, Texas, Tennessee, Louisiana, Florida.
QUANTITY       4,034 cases (24 bottles per case) were distributed; firm         
               estimates 947 cases remain on the market.
REASON         Product contains undeclared FD&C Yellow #5.


RECALLS AND FIELD CORRECTION:  FOODS -- CLASS III
              
PRODUCT        L-Carnitine 250 mg Tablets, under the Rugby and Barth's label.   
               Recall #F-226-3.
CODE           3841 0921 EXP 12/93 (Rugby & Barth's), 2301-1901 EXP 10/94       
               (Barth's).
MANUFACTURER   Tishcon Corporation, Westbury, New York.
RECALLED BY    Manufacturer, by letter November 16, 1992.  Firm-initiated       
               recall ongoing.
DISTRIBUTION   New York.
                                       
QUANTITY       1,365 units of lot 3841-0921 and 1,017 units of lot 2301-1901    
               were distributed; firm estimates no product remains on the       
               market from lot 3841-0921 and 50 units remain on the market      
               from lot 2301-1901.
REASON         Product is subpotent in L-Carnitine, 67.9% of label declaration.


RECALLS AND FIELD CORRECTIONS:  DRUGS -- CLASS III
              
PRODUCT        Laxative products under The Rx Place label:
               (a) Stool Softener Docusate Sodium 100 mg, 100 softgels, OTC;
               (b) Laxative with Stool Softener, Casanthranol and Docusate      
               Sodium, 100 softgels, OTC.  Recall #D-145/146-3.
CODE           Lot numbers:  (a) 1063-2 EXP 12/93; (b) 1063-4 EXP 1/94.
MANUFACTURER   Pharmacaps, Inc., Elizabeth, New Jersey.
RECALLED BY    G&W Labs, S. Plainfield, New Jersey (responsible firm), by       
               letter dated October 27, 1992.  Firm-initiated recall complete.
DISTRIBUTION   Massachusetts, Rhode Island, New Jersey, New York, Pennsylvania.
QUANTITY       (a) 1,512 containers; (b) 1,440 containers were distributed.
REASON         Main label panels correctly declare ingredients.  Side panels    
               reversed the ingredients of the two products.

              
PRODUCT        Nitrotan, germicidal wound spray, in 4 ounce and 8 ounce         
               aerosol cans, an OTC for topical use.  Recall #D-160-3.
CODE           Lots 1000 through 2200.
MANUFACTURER   Cramer Products, Inc., Gardner, Kansas.
RECALLED BY    Manufacturer, by letter during the first week of November        
               1992.  Firm-initiated field correction complete.
DISTRIBUTION   Nationwide.
QUANTITY       52,980 cans were distributed.
REASON         Internal mechanism of the aerosol valve may fail, allowing       
               product leakage.


RECALLS AND FIELD CORRECTIONS:  BIOLOGICS -- CLASS II
              
PRODUCT        Red Blood Cells.  Recall #B-149-3.
CODE           Unit #1093765.
MANUFACTURER   Hunter Blood Center, Inc., Clearwater, Florida.
RECALLED BY    Manufacturer, by letter March 18, 1992.  Firm-initiated recall   
               complete.
DISTRIBUTION   Florida.
QUANTITY       1 unit.
REASON         Blood product, which tested negative for all required and        
               recommended viral markers, but was collected from a donor who    
               experienced an occupational needle stick exposure approximately  
               seven months prior to donation, was distributed.

              
PRODUCT        Red Blood Cells.  Recall #B-156-3.
CODE           Unit numbers:  28025-8994, 28025-8997, 28025-8998, 28025-9000,   
               28025-1602, 28026-1604, 28026-1607, 28026-1608, 28026-1609.

                                      -2-
MANUFACTURER   United Blood Services Blood Systems, Inc., San Angelo, Texas.
RECALLED BY    Manufacturer, by visit April 2, 1992.  Firm-initiated recall     
               complete.
DISTRIBUTION   Texas.
QUANTITY       9 units.
REASON         Red Blood Cells, prepared more than eight hours after the        
               collection of the Whole Blood units, were distributed.


RECALLS AND FIELD CORRECTIONS:  DEVICES -- CLASS II
              
PRODUCT        (a) Stainless Steel Smoke Evacuation Vaginal Speculums,          
               Standard Size;
               (b) Stainless Steel Smoke Evacuation Vaginal Speculums, Large    
               Size.  Recall #Z-119/120-3.
CODE           All product distributed prior to April 21, 1992.
MANUFACTURER   Macho International, Pakistan.
RECALLED BY    Ellman International, Inc., Hewlett, New York, by letter April   
               21, 1992.  Firm-initiated recall complete.
DISTRIBUTION   Nationwide.
QUANTITY       120 units were distributed.
REASON         The underside of the arm of the speculum around the area of the  
               evacuation tube may oxidize/rust depending on the sterilization  
               technique used.

              
PRODUCT        I.V. Catheter Extension and Y-Type Extension Sets, single-use    
               devices which are used on both adults and children:
               Baxter brand Catheters:  (a) Catalog #2N1190; (b) Catalog        
               #2N1192; (c) 2N1194;
               Medex brand Catheters:  (d) MX451-FL, (e) MX453; (f) MX453-L;    
               (g) MX453-2L, (h) MX453-SL, (i) MX456-L; (j) M458L; (k) MX459;   
               (l) MX596-C; (m) MX613; (n) MX6251.
               Recall #Z-231/244-3.
CODE           Lot numbers:  (a) 22A07063 through and including 22I10155;
               (b) 22A21099 through and including 22I08183;
               (c) 22A02014 through and including 22I15051;
               (d) 21J23014 through 22E27028;
               (e) 21L7126 through 22A23112;
               (f) 21J08044 through 22E27173;
               (g) 22B18047 through 22E07052;
               (h) 21J16050 through 22C16041;
               (i) 22B27047 through 22F12040;
               (j) 22E21020 through 22E21020;
               (k) 22A07048 through 22E14053;
               (l) 21K20065 through 22C09109;
               (m) 22C04123 through 22D07085;
               (n) 22A31064 through 22D13060.
MANUFACTURER   Baxter Healthcare Corporation, Deerfield, Illinois.
RECALLED BY    Medex, Inc., Hillard, Ohio, by letter October 8, 1992.           
               Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide and international.

                                      -3-
QUANTITY       1,808,783 devices were distributed.
REASON         The female luer is cracking and leaking.

              
PRODUCT        Boston Envision Rigid Gas Permeable Daily Wear Contact Lenses    
               (blue), Part #80168.  Recall #Z-261-3.
CODE           Lot #0014230.
MANUFACTURER   Polymer Technology, Wilmington, Massachusetts.
RECALLED BY    Manufacturer, by letter or by telephone November 2, 1992.        
               Firm-initiated recall ongoing.
DISTRIBUTION   Ohio, Texas, France, Poland, Yugoslavia.
QUANTITY       51 lenses were distributed.
REASON         The lenses were cut as -3.25 D, but were labeled with a power    
               of -3.00 D.

              
PRODUCT        Jude Medical 9 Fr RediGuard DL 40 cc Intra-Aortic Balloon        
               Catheter Kit, Catalog #060-0525, containing UMI Cath-Seal        
               Percutaneous Catheter Introducer (PCI).  Recall #Z-262-3.
CODE           UMI part #1205-60-3701, lot numbers:  138187, 138188, 138189,    
               137373, 136333, 137698, 137802, 136829, 137843.
MANUFACTURER   Universal Medical Instruments, Ballston, SPA, New York.
RECALLED BY    St. Jude Medical, Inc., Chelmsford, Massachusetts.  Salesforce   
               notified hospitals on December 25, 1992, followed by letter      
               dated January 7, 1993.  Firm-initiated recall ongoing.
DISTRIBUTION   Florida, Texas, Pennsylvania, Missouri, Michigan, Illinois.
QUANTITY       115 kits were distributed.
REASON         The bond between the sheath and hub assembly may separate due    
               to contact with aqueous solutions.

              
PRODUCT        TENS (Transcutaneous Electrical Nerv Stimulator), a Rx device    
               distributed under the Dial Away Pain 400 and Synaptic 1000       
               labels.  Recall #Z-327-3.
CODE           Serial numbers 101 through 146 (except unit 131).
MANUFACTURER   Halleck-Willard, Inc., Fredrick, Colorado.
RECALLED BY    Relief From Pain, Inc., Aurora, Colorado, by letter on or about  
               August 5, 1992.  Firm-initiated field correction ongoing.
DISTRIBUTION   Iowa, California, Michigan, North Carolina, Colorado, New York,  
               Washington state, Illinois, Nebraska, North Dakota.
QUANTITY       16 units were subject to field correction as of 8/3/92.
REASON         Device may deliver a small shock to the patient.

              
PRODUCT        Sterile Mayfield Disposable Skull Pins, used to keep a           
               patient's head stationary during brain surgery:
               (a) Stainless Steel; (b) Stainless Steel with a Polycarbonate    
               base  Recall Z-330/331-3.
CODE           Part Nos.     Lot Nos.
               A-1072        52921291, 52915092, 52929592, 52929592A, 52926692,
                             52927992, 52932492
               A-1072S       35505992, 32913292, 52913292, 52931092.
MANUFACTURER   Ohio Medical Instrument Company, Cincinnati, Ohio.

                                      -4-
RECALLED BY    Manufacturer, by letter January 22, 1993.  Firm-initiated        
               recall ongoing.
DISTRIBUTION   Nationwide, Germany, Canada, Japan, Spain.
QUANTITY       Approximately (a) 58; (b) 439 boxes (12 units per box) were      
               distributed.
REASON         Both products were marketed without an approved 510(k).

              
PRODUCT        Life Pulse High Frequency Ventilator Humidifier Cartridges,      
               also called patient breathing circuits.  Recall #Z-347-3.
CODE           Lots 133 - 179.
MANUFACTURER   Bunnell, Inc., Salt Lake City, Utah.
RECALLED BY    Manufacturer, by letter November 13, 1992.  Firm-initiated       
               recall ongoing.
DISTRIBUTION   Nationwide, Japan, Korea.
QUANTITY       Approximately 11,386 units were distributed.
REASON         The cartridges are susceptible to cracking at the hose ports,    
               which could result in breakage and failure.


RECALLS AND FIELD CORRECTIONS:  DEVICES -- CLASS III
              
PRODUCT        Nor-Partigen Control Set I, an in-vitro diagnostic kit, for use  
               with the monospecific, Nor-Partigen Plates for the quantitative  
               determination of immunoglobulins in human serum.
               Recall #Z-283-3.
CODE           Catalog #942001, lot numbers:  20282 EXP 8/90, 20807 EXP 11/90.
MANUFACTURER   Behringwerke, AG, Marburg, Germany.
RECALLED BY    Behring Diagnostics, Somerville, New Jersey, by letter July 17,  
               1990.  Firm-initiated field correction complete.
DISTRIBUTION   Nationwide.
QUANTITY       188 sets were distributed.  Firm estimates none remains on the   
               market, product is expired.
REASON         IqM constituent valve of these controls was reading greater      
               than 15% of its high value.  This value is outside its limit of  
               acceptability for this constituent.

              
PRODUCT        Vacutainer brand SST Blood Collection Tubes.  Recall #Z-291-3.
CODE           Catalog #6509, lot #2C914 EXP 4/93.
MANUFACTURER   Becton Dickinson Vacutainer Systems, Sumter, South Carolina.
RECALLED BY    Becton Dickinson Vacutainer Systems, Rutherford, New Jersey, by  
               telephone and letter dated July 20, 1992.  Firm-initiated        
               recall complete.
DISTRIBUTION   Nationwide, Canada, England, Japan.
QUANTITY       436,000 units were distributed.
REASON         Product may break in the centrifuge due to firecrack defects in  
               the bottom of the tube.

                                      -5-
              
PRODUCT        Coumatrak Protime Controls, lyophilized whole blood based        
               control materials for two levels of prothrombin time, normal     
               and abnormal, packaged in 1 ml vials, for in-vitro use only as   
               a check on the analytical performance of the Coumatrak Monitor   
               and Protime Test System.  Recall #Z-308-3.
CODE           Lot numbers:  CPC1L15, CPC1M16, CPC2A01, CPC2A02, CPC2A03,       
               CPC2B04, CPC2B05, CPC2B06, CPC2E11, CPC2F13.
MANUFACTURER   Consolidated Technologies, Inc., Austin, Texas.
RECALLED BY    Biotrack, Inc., Mountain View, California, by telephone          
               November 23, 1992, followed by Fax and letter.  Firm-initiated   
               recall ongoing.
DISTRIBUTION   Delaware.
QUANTITY       11,748 boxes (5 vials per box) were distributed.
REASON         The controls take longer to clot than the clotting time ranges   
               on the label.

              
PRODUCT        Ventricular Drainage System, intended for single use, to         
               facilitate the collection of cerebrospinal fluid (CSF), fluid    
               injection, CSF sampling and intracranial pressure monitoring:
               (a) Ventricular Drainage System, Model 041;
               (b) Ventricular Bolt Pressure Monitoring & Drainage Kit,
               Model 110-4HC;
               (c) Microventricular Bolt Pressure Monitoring & Drainage Kit,    
               Model 110-HMC.  Recall #Z-318/320-3.
CODE           Sterilization codes:  B2003, B2004, C2002, C2004, F2001, F2002,  
               G2004.
MANUFACTURER   Plasco, Inc., Gurnee, Illinois (component).
RECALLED BY    Camino Laboratories, Inc., San Diego, California, by letter      
               September 1, 1992.  Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide, Argentina, Italy.
QUANTITY       2,920 units were distributed.
REASON         The bushings of the ventricular tubing do not fit properly       
               which could lead to leakage at the bushing and tubing            
               connection. 

              
PRODUCT        Nova Stat Profile Control, Level 2:
               (a) Nova Catalog #13425 (Baxter Catalog #B6352-49) Stat Profile  
               Control Level 2;
               (b)  Nova Catalog #13427 (Baxter Catalog #B6352-51) Stat         
               Profile Control Multipak with Glucose;
               (c) Nova Catalog #13428 Stat Profile Control QAP Multipak;
               Recall #Z-321/323-3.
CODE           Catalog Nos.      Lot Nos.
               (a) 13425         2706 Level 2
               (b) 13427         2655 Level 1, 2706 Level 1, 2707 Level 3
               (c) 1324-QAP      2655 Level 1, 2706 Level 2, 2707 Level 3.
               Note:  Level 2 product only was mislabeled.
MANUFACTURER   Nova Bio Medical, Waltham, Massachusetts.
RECALLED BY    Manufacturer, by telephone during August 26-31, 1992, and        
               letter/fax August 31, 1992.  Firm-initiated recall complete.

                                      -6-
DISTRIBUTION   Nationwide and international.
QUANTITY       (a) 23 boxes of 30; (b) 508 boxes of 30; (c) 20 boxes of 30      
               ampoules were distributed.
REASON         Some ampoules labeled as containing Level 2 control actually     
               contained Level 1 control.

              
PRODUCT        Concise Strep A Kit, an in-vitro diagnostic assay used for the   
               qualitative determination of Group A Streptococcus antigen from  
               throat swabs.  Recall #Z-324-3.
CODE           Catalog Nos.    Lot Nos.   EXP Date
               4512            290963     1/13/93
                               291003     1/20/93
                               291004     1/20/93
                               291006     1/20/93
                               291007     1/20/93
               4513            290792     12/16/92
MANUFACTURER   Pacific Biotech, Inc., San Diego, California.
RECALLED BY    Hybritech, Inc., San Diego, California, by letter December 2,    
               1992.  Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide.
QUANTITY       2,265 kits were distributed.
REASON         The product will produce a false positive result.

              
PRODUCT        Total Iron and Total Iron Binding Capacity (TIBC) Test Kit.      
               Recall #Z-325-3.
CODE           Kit lot numbers:  10318 (30 tests each kit) and 10319 (240       
               tests each kit), EXP 9/92.  Binding reagent lot #10335
               (code 1030).
MANUFACTURER   Bio-Analytic Laboratories, Inc., Palm City, Florida.
RECALLED BY    Manufacturer, by fax or by telephone May 22, 1991.               
               Firm-initiated recall complete.
DISTRIBUTION   Michigan, Mexico, Egypt, Turkey.
QUANTITY       10 kits were distributed.
REASON         One of the reagents in the kit is precipitating.

              
PRODUCT        Repliplate Cathra 256 mcg/ml Sulfisoxazole (SUL) antibiotic      
               susceptibility test plate, used to check the susceptibility of   
               disease-causing bacteria to antibiotics.  Recall #Z-328-3.
CODE           Lot 11102 SUL256 of plates in lot 11102 ABS3 of the sets of      
               plates.
MANUFACTURER   AutoMed, Inc., Arden Hills, Minnesota.
RECALLED BY    Manufacturer, by telephone December 11, 1992.  Firm-initiated    
               recall ongoing.
DISTRIBUTION   Canada.
QUANTITY       80 sets containing 1 sulfisoxazole plate per set were            
               distributed.
REASON         The plates were contaminated.

              
PRODUCT        Testosterone Test  Standard, a component of the RSL 1251         
               Testosterone Test Kit, which detects the total unconjugated      
               form of this steroid, for in-vitro diagnostic use only.
               Recall #Z-326-3.
                                      -7-
CODE           Catalog #07189130, lot #RTS9203, EXP 10/1/93.
MANUFACTURER   ICN Biomedicals, Inc., Costa Mesa, California.
RECALLED BY    Manufacturer, by letter October 12, 1992.  Firm-initiated        
               recall complete.
DISTRIBUTION   California, Illinois, Indiana, New Jersey, Pennsylvania, Texas,  
               Washington state, Wisconsin.
QUANTITY       228 sets were distributed.
REASON         Lack of product stability when stored above the recommended      
               -15C. 

              
PRODUCT        Abbott TestPack Plus hCG-Combo Test Kits, List #3A60, a          
               self-performing immunoassay in-vitro diagnostic designed for     
               the qualitative determination of human chorionic gonadotropin    
               (hCG) in urine and serum for early detection of pregnancy.
               RECALL #Z-329-3.
CODE           Lot numbers:  53136M100, 53591M200.
MANUFACTURER   Abbott Laboratories, North Chicago, Illinois.
RECALLED BY    Abbott Laboratories, Diagnostics Division, Abbott Park,          
               Illinois, by telephone June 25-July 1, 1991, followed by         
               letters dated June 25, 1991.  Firm-initiated recall complete.
DISTRIBUTION   Nationwide, Canada, Costa Rica, Dominican Republic, Venezuela,   
               Australia, Germany, Hong Kong.
QUANTITY       1,338 20-test kits and 1,868 40-test kits were distributed;      
               firm estimates none remains on the market.
REASON         Some clinical specimens could give false positive results.

              
PRODUCT        Concise Strep A In-Vitro Diagnostic Kits, intended for the       
               qualitative determination of group a streptococcus antigen from  
               throat swabs.  Recall #Z-334-3.
CODE           Lot numbers 291913 EXP 5/17/93, 291941 EXP 5/17/93,
               291943 EXP 4/28/93.
MANUFACTURER   Hybritech, Inc., San Diego, California.
RECALLED BY    Manufacturer, by letter October 7, 1992, and by telephone.       
               Firm-initiated field correction ongoing.
DISTRIBUTION   Nationwide, Canada.
QUANTITY       773 kits were distributed.
REASON         The wrong revision directions for use were packaged with the     
               kit. 


RECALLS AND FIELD CORRECTIONS:  VETERINARY -- CLASS III
              
PRODUCT        Super Pet Capsules (Pycnogenol for animals), an anti-aging       
               nutritional supplement, in bottles of 120.  Recall #V-018-3.
CODE           All lots.
MANUFACTURER   Norwest Marketing Company, Pacoima, California.
RECALLED BY    Reseau International, Cincinnati, Ohio, by letter sent during    
               the week of January 11, 1993.  Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide, Canada, Bermuda.
QUANTITY       933 bottles of tablets and 234 bottles of capsules were          
               distributed.
REASON         Product is an unapproved new animal drug.

                                      -8-
MEDICAL DEVICE SAFETY ALERT
              
PRODUCT        Therascan Beam Data Acquistion System, an accessory for          
               radiation treatment planning systems -- computerized             
               workstations used to calculate and determine appropriate         
               radiation treatment programs for doctors to administer to        
               cancer patients.  Safety Alert #N-025-3.
CODE           Serial numbers:  BDAS-004, BDAS-006, BDAS--007, BDAS-009,        
               BDAS-010, BDAS-011, BDAS-013, BDAS-014, BDAS-015, BDAS-016,      
               BDAS-019, BDAS-021, BDAS-022, BDAS-023, BDAS-025, BDAS-029,      
               BDAS-030, BDAS-033, BDAS-034, BDAS-035, BDAS-036, BDAS-038,      
               BDAS-039, BDAS-040, BDAS-042.
MANUFACTURER   Theratronics International, Ltd., (formerly Atomic Energy of     
               Canada, Ltd.,) Kanata, Ontario, Canada; Theratronics             
               International, Ltd., Carrollton, Texas.
ALERTED BY     Manufacturer, by User Bulletin BDAS UB 80-006, dated December    
               15, 1992.
DISTRIBUTION   Arkansas, California, Georgia, Florida, Michigan, Montana,       
               North Carolina, North Dakota, New York, Ohio, Texas, Virginia,   
               Wisconsin.
QUANTITY       25 units were distributed.
REASON         Radio-frequency interference produced by the magnetrons of some  
               linear accelerators may affect electrometer operation and give   
               distorted readings.

INJUNCTION:
              
PRODUCT        Various seafood products.
CHARGE         The seafood is adulterated and misbranded in that it is
               decomposed in whole or in part, or is otherwise unfit for
               food due to rancidity, dehydration, or freezer burn; it has
               been prepared, packed, or held under insanitary conditions
               whereby it may have become contaminated with filth, or
               whereby it may have been rendered injurious to health; and,
               it is offered for sale under the name of another food.
DEFENDANTS     Impex Shrimp & Fish Company, Inc., and Bernard M. Kane, Elk
               Grove Village, Illinois.
FILED          December 30, 1992, U.S. District Court for the Northern
               District of Illinois, Eastern Division; Civil #92C 8399,
               INJ 1312.

MASS SEIZURE:
              
PRODUCT        Oxygen, USP and Nitrous Oxide, USP (92-596-608/9).
CHARGE         Adulterated - The methods used in, and the facilities and
               controls used for, the articles' manufacture, processing,
               and holding do not conform to and are not operated and
               administered in conformity with current good manufacturing
               practice regulations.
FIRM           Grand Rapids Welding Supply Co., Grand Rapids, Michigan.
FILED          February 3, 1993; U.S. District Court for the Western
               District of Michigan; Civil #1:93CV88, FDC #66634.
SEIZED         February 3, 1993 - goods valued at approximately $28,035.

                                      -9-
SEIZURES:
              
PRODUCT        Frozen seafood - raw trim, Alaskan pollock portions, and
               surfburger patties (92-661-970/2).
CHARGE         Adulterated - Some of the articles consist in whole or in
               part of decomposed seafood, and are unfit for food because
               they are rancid.  Misbranded - Some of the articles' labels
               fail to bear the common or usual name of the foods; and one
               of the articles is fabricated from two or more ingredients
               and the label fails to bear the common or usual name of each
               ingredient.
FIRM           Americold Corporation, Chicago, Illinois.
FILED          December 23, 1992; U.S. District Court for the Northern
               District of Illinois, Eastern Division; Civil #92C 8400,
               FDC #66632.
SEIZED         December 30, 1992 - goods valued at approximately $5,600.

              
PRODUCT        Seafood - Pacific whiting and whiting portions (92-661-564).
CHARGE         Adulterated - Both articles consist in part of a
               decomposed substance, and one of the articles is otherwise
               unfit for food because it is rancid.
FIRM           Impex Shrimp and Fish Company, Elk Grove Village, Illinois.
FILED          December 23, 1992; U.S. District Court for the Northern
               District of Illinois, Eastern Division; Civil #92C 8398,
               FDC #66639.
SEIZED         December 30, 1992 - goods valued at approximately $11,400.

              
PRODUCT        Frozen peeled and deveined shrimp (92-609-381).
CHARGE         Adulterated - The article consists in whole or in
               part of decomposed shrimp.
FIRM           Ala-Maid Seafood Co., Inc., Whistler, Alabama.
FILED          December 11, 1992; U.S. District Court for the Southern
               District of Alabama, Southern Division; Civil #92-1038-RV-M,
               FDC #66638.
SEIZED         December 14, 1992 - goods valued at approximately $1,392.

              
PRODUCT        Frozen flounder fillets and cooked and peeled shrimp
               (93-650-993/4).
CHARGE         Adulterated - The articles consist in whole or in
               part of decomposed seafood, and one of the articles is
               otherwise unfit for food because it is rancid.
FIRM           Imperial Foods, Inc., Smyrna, Tennessee.
FILED          December 23, 1992; U.S. District Court for the Middle
               District of Tennessee, Nashville Division; Civil #3-02-1141,
               FDC #66647.
Seized:        December 30, 1992 - goods valued at approximately $4,433.

                                     -10-
              
PRODUCT        Various drugs (92-550-769 et al.).
CHARGE         New drug - The articles are unapproved new drugs.
               Misbranded - the articles' labeling is false and
               misleading, and fails to bear adequate directions for the
               user for which they were intended.
FIRM           Research Health Care, New Troy, Michigan.
FILED          January 15, 1993; U.S. District Court for the Western
               District of Michigan; Civil #1:93CV39, FDC #66613.
Seized:        February 1, 1993 - goods valued at approximately $4,100.

              
PRODUCT        Gentamicin Sulphate B.P. (for veterinary use) (93-661-157).
CHARGE         Misbranded - The article's label fails to bear adequate
               directions for the use for which it is intended.
FIRM           British Airways Cargo Terminal, Des Plaines, Illinois.
FILED          February 3, 1993; U.S. District Court for the Northern
               District of Illinois, Eastern Division; Civil #93C 0703,
               FDC #66674.
SEIZED         February 4, 1993 - goods valued at $6,500.

              
PRODUCT        Microwave hyperthermia units (92-663-235 & 92-684-342).
CHARGE         Adulterated - The articles are class III medical devices
               which do not have in effect the required approved
               premarket approval applications.
               Misbranded - The articles' labeling is false and misleading,
               and does not bear adequate directions for use for the
               purposes for which they were intended; and, notice or other
               information respecting the articles was not filed as
               required.
FIRM           HBCI, Inc., doing business as Valley Cancer Institute,
               Los Angeles, California, and stored to the account of HBCI
               at E-Z Storage, Culver City, California.
FILED          September 4, 1992 - U.S. District Court for the Central
               District of California; Civil ##CV925435 HLH(Dx) &
               CV925436 HLH(Dx), FDC #66485 & #66497.
SEIZED         September 23, 1992 - six units valued at $1.5 million.


                                     -11-


END OF ENFORCEMENT REPORT FOR MARCH 3, 1993.  BLANK PAGES MAY
FOLLOW.