Additional SIMS Map of MOS 1213 Chip F Crater 3

Impact feature information

Feature Image



SIMS Images

Below are the SIMS images of this feature. A thumbnail image overview is available, too. The analysis was performed on 00-00-1980. The following mass numbers were imaged:
27, 56, 56, 24, 12, 23, 39

Analytical Conditions:

These parameters directly affect spectrometer signal intensities.

Primary beam:

Secondary beam:


Image 1, Al (Mass 27)

signal intensity scale: 0 -> 232

NOTE: CHANNEL PLATE GAIN AND PRIMARY BEAM CURRENT WAS REDUCED SUBSTANTIALLY
      ON Aluminum SUBSTRATES.
--------------------------------------------------------------------------------
			% isotopic	Resolution
Element Interferents	abundance	  (m/Dm)	Comments
--------------------------------------------------------------------------------
27Al   *26MgH		 11.0		  +3042 	moderate abundance,
							idenitfied Al line if
							sample with low Al
	54Cr2+,*13C15N,*12C2H3,54Fe2+	   <2298	resolved
--------------------------------------------------------------------------------
NOTE: An asterisk (*) means that these species were in fact observed on Hi-Res
      Mass Spectra of a particular mass on the Standard.

Image Conditions

These parameters directly affect image intensities.

Image 2, Mass 56 (Si2)

signal intensity scale: 0 -> 466

NOTE: 28Si2 and 40CaO are not resolvable among themselves and appear as
      one peak to the right of the 56Fe peak.
--------------------------------------------------------------------------------
			% isotopic	Resolution
Element Interferents	abundance	  (m/Dm)	Comments
--------------------------------------------------------------------------------
56Fe
 --AND--
56Si2	55MnH		 99.9		  +5118 	not resolvable, unlikely
							to be present without Fe
       *28Si2,44CaC,*40CaO,29SiAl,	   <2956	resolved
	40KO,32SC2,24MgO2,C4H6
--------------------------------------------------------------------------------
NOTE: An asterisk (*) means that these species were in fact observed on Hi-Res
      Mass Spectra of a particular mass on the Standard.

Image Conditions

These parameters directly affect image intensities.

Image 3, Mass 56 (Fe)

signal intensity scale: 0 -> 570

NOTE: 28Si2 and 40CaO are not resolvable among themselves and appear as
      one peak to the right of the 56Fe peak.
--------------------------------------------------------------------------------
			% isotopic	Resolution
Element Interferents	abundance	  (m/Dm)	Comments
--------------------------------------------------------------------------------
56Fe
 --AND--
56Si2	55MnH		 99.9		  +5118 	not resolvable, unlikely
							to be present without Fe
       *28Si2,44CaC,*40CaO,29SiAl,	   <2956	resolved
	40KO,32SC2,24MgO2,C4H6
--------------------------------------------------------------------------------
NOTE: An asterisk (*) means that these species were in fact observed on Hi-Res
      Mass Spectra of a particular mass on the Standard.

Image Conditions

These parameters directly affect image intensities.

Image 4, Mass 24 (Mg)

signal intensity scale: 0 -> 3348

--------------------------------------------------------------------------------
					Resolution
Element Interferents			  (m/Dm)	Comments
--------------------------------------------------------------------------------
24Mg   *48Ti2+,48Ca2+,*23NaH+,12C2	   <2731	resolved
--------------------------------------------------------------------------------
NOTE: An asterisk (*) means that these species were in fact observed on Hi-Res
      Mass Spectra of a particular mass on the Standard.

Image Conditions

These parameters directly affect image intensities.

Image 5, Mass 12 (C)

signal intensity scale: 0 -> 169

--------------------------------------------------------------------------------
					Resolution
Element Interferents			  (m/Dm)	Comments
--------------------------------------------------------------------------------
12C	24Mg2+,*11BH			   <1604	resolved, non-coincident
							with Mg distribution
--------------------------------------------------------------------------------
NOTE: An asterisk (*) means that these species were in fact observed on Hi-Res
      Mass Spectra of a particular mass on the Standard.

Image Conditions

These parameters directly affect image intensities.

Image 6, Mass 23 (Na)

signal intensity scale: 0 -> 617

--------------------------------------------------------------------------------
			% isotopic	Resolution
Element Interferents	abundance	  (m/Dm)	Comments
--------------------------------------------------------------------------------
23Na	30SiO2+ 	  3.10		  -4222 	not resolvable, low
							abundance. Na distri-
							bution had to be non-co-
							incident with high con-
							centration of the Si
							distribution for a
							positive ID. Looked for
							NaH (not resolvable from
							12C2) and/or NaO (not
							resolvable from 27AlC)
							in samples with high Si.
	*46Ti2+,46Ca2+			   <1776	resolved
--------------------------------------------------------------------------------
NOTE: An asterisk (*) means that these species were in fact observed on Hi-Res
      Mass Spectra of a particular mass on the Standard.

Image Conditions

These parameters directly affect image intensities.

Image 7, Mass 39 (K)

signal intensity scale: 0 -> 241

--------------------------------------------------------------------------------
					Resolution
Element Interferents			  (m/Dm)	Comments
--------------------------------------------------------------------------------
39K    *27AlC,23NaO,12C3H3		   <2190	resolved
--------------------------------------------------------------------------------
NOTE: An asterisk (*) means that these species were in fact observed on Hi-Res
      Mass Spectra of a particular mass on the Standard.

Image Conditions

These parameters directly affect image intensities.

retrieve SIMS data fileget C source code
IDE database home page Back to surfaces page
Klaus G. Paul, 7-12-1994