Event description cpi received information that the rate sensing portion of this endotak down-sized plus lead was capped.
The patient with an implantable cardioverter defibrillator (icd) and this lead was experiencing oversensing and inappropriate therapy.
The physician elected to perform an invasive procedure to replace the rate sensing portion, in order to remove the pin from the device the physician tried to loosen the screws on the header.
On one screw he broke off, another screw broke in such a way that is was impossible to remove the broken part of the screw driver from the inside of the screw.
A new icd was implanted.
|
Event conclusion cpi's analysis found: only the terminal pin section was returned.
Outer insulation of the rate sensing terminal pin section is torn apart at the distal end of the terminal pin barrel.
Rate sense positive conductor coil filars (4) are fractured at the distal end of the terminal pin barrel.
Fractured conductor coils and insulation damage are most likely due to a load being applied to the lead either at implant, or due to the position of the lead in the body, coupled with movement.
The icd was also returned and cpi's analysis found: the torque wrench tip was stuck in the set screw.
All other set screws passed the gauge pin test.
The icd was electrically in specification.
|