Medical Device Recalls
Class 2 Recall
St. Jude Medical CPS Direct SL
|
|
Date Recall Initiated |
June 01, 2006
|
Date Posted |
June 24, 2006
|
Recall Number |
Z-1166-06
|
Product |
CPS Direct SL, Models 410110, 410111, 410112, 410113, 410114, 410115, 410116, 410120, 410121, 410122, 410123, 410124, 410125, & 410126
|
Code Information |
All lots with a manufacturing date of May 30, 2006 or earlier
|
Recalling Firm/ Manufacturer |
St Jude Medical CRMD
15900 Valley View Ct
Sylmar
, California
91342-3577
|
For Addition Information Contact |
Nestor Kusnierz
818-493-2587
|
Reason For Recall |
Sterility-Some of the sterile pouches in which these products are packaged may not have been properly sealed during their packaging process.
|
Action |
The attached 'Dear Hospital Administrator' letter was sent out to our US sales representatives via FedEx on June 1, 2006. The sales representatives have been instructed to send back all of the devices in their inventory and to retrieve and return any remaining devices that were placed into hospital inventory. They are also instructed to deliver the attached 'Dear Hospital Administrator' letter to the appropriate person at each hospital where product is to be removed from the hospital shelf.
.
|
Quantity in Commerce |
6290
|
Distribution |
Hospitals/Medical Centers in AZ, CA, MS, NY, & PA
|
|
|
|
|
|