FDA

Enforcement Report

The FDA Enforcement Report is published weekly by the Food and Drug Administration, U.S. Public Health Service, Department of Health and Human Services. It contains information on actions taken in connection with agency regulatory activities.
 ENFORCE
08/08/1995

 ENFORCEMENT REPORT FOR 08/09/1995 


August 9, 1995                                                  95-32

RECALLS AND FIELD CORRECTIONS:  FOODS -- CLASS I
==========================
_______________
PRODUCT        Cooked, Ready-To Eat Crabmeat in 8-16 ounce plastic
               containers.  Recall #F-717-5.
CODE           None.  Product manufactured between April 17 to June 27,
               1995.
MANUFACTURER   John W. Dunbar Company, West Tremont, Maine.
RECALLED BY    Manufacturer, by telephone June 27, 1995.  Firm-initiated
               recall complete.
DISTRIBUTION   Massachusetts, Rhode Island, Maine.
QUANTITY       Firm estimates none remains on the market.
REASON         The product is contaminated with Listeria monocytogenes.

_______________
PRODUCT        Kroger brand Deluxe Butter Pecan Ice Cream and Texas Gold
               Butter Pecan Ice Cream in 1/2 gallon cartons.  
               Recall #F-718-5.
CODE           NOV 95YE.
MANUFACTURER   The Kroger Company, Cincinnati, Ohio.
RECALLED BY    Manufacturer, by letters sent June 2 and 7, 1995.  Firm-
               initiated recall complete.
DISTRIBUTION   South Carolina, Georgia, Alabama, Tennessee, Kentucky.
QUANTITY       1,440 half gallon cartons were distributed.
REASON         The product contains undeclared peanuts.

_______________
PRODUCT        Fancy Extra Large Pecan Pieces BRS (Butter Roasted Salted)
               in bulk 30 pound cases.  Recall #F-719-5.
CODE           Lot 554.
MANUFACTURER   Young Pecan Company, Florence, South Carolina.
RECALLED BY    Manufacturer, by telephone.  Firm-initiated recall complete.
QUANTITY       551 30-pound cases were distributed; firm estimates none
               remains on the market.
REASON         Peanuts in pecan pieces.


RECALLS AND FIELD CORRECTIONS:  FOODS -- CLASS II
=========================
_______________
PRODUCT        Zetov brand Supersnacks in 2.5 ounce packages:  
               (a) Falafel Flavored Snacks;
               (b) BBQ Flavored Snacks;
               (c) Taco Flavored Snacks.  Recall #F-697/699-5.
CODE           Production code on cellophane bags:  5004 and 50043;
               Code on master shipping carton - 6253 or 6254.
MANUFACTURER   OSEM Food Industries, Ltd., Petach-Tikva, Israel.
RECALLED BY    Zetov, Inc., Brooklyn, New York, by telephone and by letter
               April 6, 1995.  Firm-initiated recall ongoing.
DISTRIBUTION   New York.
QUANTITY       2,202 cases were distributed; firm estimates none remains on
               the market.
REASON         Product is contaminated with Salmonella tennessee.

_______________
PRODUCT        Osem brand Bissli Snacks, in 2.5 ounce packages:
               (a) Falafel Flavored Snacks; (b) BBQ Flavored Snacks;
               (c) Taco Flavored Snacks.  Recall #F-710/712-5.
CODE           Production code on cellophane bags - 5005, 50043 or 41983;
               Code on master shipping carton - 6253, 6254, 5968, 6250.
MANUFACTURER   OSEM Food Industries, Ltd., Petach-Tikva, Israel.
RECALLED BY    OSEM USA, Inc., Englewood Cliffs, New Jersey, by telephone
               and by letter April 5, 1995.  Firm-initiated recall ongoing.
DISTRIBUTION   California, Florida, Illinois, Louisiana, Massachusetts,
               Maryland, Michigan, Minnesota, North Carolina, New Jersey,
               New York, Pennsylvania, Texas.
QUANTITY       2,654 cartons were distributed; firm estimates that little
               or no product remains on the market.
REASON         Product is contaminated with Salmonella. 
_______________
PRODUCT        Kellogg's brand Apple Jacks Cereal, in 15 ounce cartons. 
               Recall #F-715-5.
CODE           Carton #K-0292D with "better if used before" date of SEP 11,
               1995 MB009.
MANUFACTURER   Kellogg's, Inc., Memphis, Tennessee.
RECALLED BY    Kellogg's USA, Battle Creek, Michigan, by telephone October
               28, 1994, followed by fax and press release.  Firm-initiated
               recall complete.
DISTRIBUTION   Arkansas Georgia, Mississippi, Tennessee, Missouri, Texas.
QUANTITY       79,170 cartons were distributed.
REASON         Product contained undeclared almonds.


RECALLS AND FIELD CORRECTIONS:  FOODS -- CLASS III
========================
_______________
PRODUCT        Hi-C Fruit Punch in 8.45 ounce aseptic containers packaged
               in (a) Three pack containers; and (b) Variety pack
               containers, containing 3 fruit punch packages and six
               packages of 2 other flavors.  Recall #F-713/714-5.
CODE           (a) MAR1996HC (container), MAR1996HCD (case); (b) 1st line:
               01914XXXXXX; 2nd line:  Contains at some point MAR1996.
MANUFACTURER   Coca-Cola Foods, Highstown, New Jersey.
RECALLED BY    Coca-Cola Foods, Division of The Coca-Cola Company, Houston,
               Texas, by visit on June 29, 1995.  Firm-initiated recall
               complete.
DISTRIBUTION   Massachusetts, New York, Pennsylvania.
QUANTITY       (a) 1,680 packages;(b) 9,600 packages were distributed.
REASON         Product is contaminated with yeast.


RECALLS AND FIELD CORRECTIONS:  DRUGS -- CLASS II
=========================
_______________
PRODUCT        Just One Per Day, plus any other private labels listed on
               the carton of Phenylpropanolamine 75 mg, OTC, Time Released
               Caplets for Weight Loss, packed in cartons containing 24 or
               48 caplets each.  Recall #D-228-5.
CODE           Lot #53D16J4 EXP 6/96.
MANUFACTURER   Davis Pharmaceuticals, Norwich, Connecticut.
RECALLED BY    The Reese Chemical Company, Cleveland, Ohio, by letter July
               19, 1995.  Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide.
QUANTITY       112,292 individual caplets were distributed.
REASON         Product does not meet dissolution specifications.

_______________
PRODUCT        Dilantin Kapseals, 30 mg, Rx antiepileptic drug.
               Recall #D-229-5.
CODE           Lot #27224L EXP 2/96.
MANUFACTURER   Warner-lambert Company, Lititz, Pennsylvania.
RECALLED BY    Manufacturer, by letters dated June 23, 1995.  Firm-
               initiated recall complete.
DISTRIBUTION   Nationwide.
QUANTITY       Undetermined.
REASON         Product does not meet dissolution specifications.

_______________
PRODUCT        Bulk Medical Oxygen USP.  Recall #D-230-5.
CODE           Lot numbers: 141-0149-95-02-A and 145-0149-95-02-B.
MANUFACTURER   Air Liquide America Corporation, Ingleside, Texas.
RECALLED BY    Air Liquide America Corporation, Houston, Texas, by letters
               dated June 6, 1995, followed by visits.  Firm-initiated
               recall complete.
DISTRIBUTION   Texas.
QUANTITY       Firm estimates none remains on the market.
REASON         Current good manufacturing practice deficiencies.

_______________
PRODUCT        Rifadin I.V. (rifampin for injection), 600 mg, Rx for the
               treatment of tuberculosis.  Recall #D-231-5.
CODE           Lot 025A EXP 7/97.
MANUFACTURER   Gruppo Lepetit S.p.A., Lainate, Italy.
RECALLED BY    Marion Merrell Dow, Inc., Kansas City, Missouri, by
               telephone June 23, 1995.  Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide.
QUANTITY       18,237 vials were distributed.
REASON         Lack of assurance of sterility.


RECALLS AND FIELD CORRECTIONS:  DRUGS -- CLASS III
========================
_______________
PRODUCT        Hydrosol Anesthetic Medicated Ointment, packed in 1 ounce
               tubes and in 1 pound jars, OTC used for the treatment of
               minor skin irritations.  Recall #D-227-5.
CODE           RAES, WCJA, WBGA on 1 pound jars; P-3, P-4, P-5 on 1 ounce
               tubes.
MANUFACTURER   Hydrosol Manufacturing Company, Miamitown, Ohio.
RECALLED BY    Manufacturer, by letter July 25, 1995.  Firm-initiated
               recall ongoing.
DISTRIBUTION   Nationwide.
QUANTITY       Approximately 7,452 tubes and 348 jars of the product were
               distributed.
REASON         Current good manufacturing practice deficiencies.


RECALLS AND FIELD CORRECTIONS:  BIOLOGICS -- CLASS II
=====================
_______________
PRODUCT        Red Blood Cells.  Recall #B-492-5.
CODE           Unit #12FZ16639.
MANUFACTURER   American National Red Cross, Charlotte, North Carolina.
RECALLED BY    Manufacturer, by letters dated February 24, 1995.  Firm-
               initiated recall complete.
DISTRIBUTION   North Carolina, Switzerland.
QUANTITY       1 unit.
REASON         Blood products, which were collected from a donor with a
               medical history of cancer, were distributed.

_______________
PRODUCT        Red Blood Cells.  Recall #B-494-5.
CODE           Unit #32GQ91533.
MANUFACTURER   American National Red Cross, Madison, Wisconsin.
RECALLED BY    Manufacturer, by letters dated July 13 or 14, 1994.  Firm-
               initiated recall complete.
DISTRIBUTION   Missouri, Switzerland.
QUANTITY       1 unit.
REASON         Blood products, which were collected from a donor with a
               medical history of cancer, were distributed.

_______________
PRODUCT        (a) Red Blood Cells; (b) Platelets.  Recall #B-498/499-5.
CODE           Unit #s (a) 53LZ18012, (b) 50LZ18012.
MANUFACTURER   American National Red Cross, Baltimore, Maryland.
RECALLED BY    Manufacturer, by letters dated February 18, 1994.  Firm-
               initiated recall complete.
DISTRIBUTION   Maryland.
QUANTITY       1 unit of each component.
REASON         Blood products, which were collected from a donor who had
               traveled to an area designated as endemic for malaria and
               had taken prophylactic antimalarial medications, were
               distributed.

_______________
PRODUCT        Source Plasma.  Recall #B-500-5.
CODE           Unit #23921214.
MANUFACTURER   Binary Associates, Inc., Colorado Springs, Colorado.
RECALLED BY    Manufacturer, by fax October 28, 1992.  Firm-initiated
               recall complete.
DISTRIBUTION   California.
QUANTITY       1 unit.
REASON         A unit of Source Plasma, which tested repeatedly reactive
               for antibody to the hepatitis C virus encoded antigen (anti-
               HCV), was distributed.

_______________
PRODUCT        (a) Red Blood Cells; (b) Platelets; (c) Recovered Plasma. 
               Recall #B-501/503-5.
CODE           Unit numbers:  (a) 29107-9179, 29116-7728, 29113-7652, 9106-
               3643; (b) 29106-3643, 29116-7728; (c) 29107-9179, 29116-
               7728, 29113-7652, 29106-3643.
MANUFACTURER   United Blood Services, Chicago, Illinois.
RECALLED BY    Manufacturer, by letters dated May 24, 25, 31, 1995.  Firm-
               initiated recall ongoing.
DISTRIBUTION   Illinois, Nevada, Switzerland.
QUANTITY       (a) 4 units; (b) 2 units; (c) 4 units.
REASON         Blood products, which were incorrectly tested for antibody
               to human immunodeficiency virus types 1 and 2 (anti-HIV) or
               hepatitis B surface antigen (HBsAg), were distributed.

_______________
PRODUCT        Platelets.  Recall #B-508-5.
CODE           Unit #11FG16601.
MANUFACTURER   American National Red Cross, St. Louis, Missouri.
RECALLED BY    Manufacturer, by telephone on March 31, 1994.  Firm-
               initiated recall complete.
DISTRIBUTION   Missouri.
QUANTITY       1 unit.
REASON         Blood product, which was manufactured from Whole Blood that
               was exposed to unacceptable storage temperatures, was
               distributed.


RECALLS AND FIELD CORRECTIONS:  BIOLOGICS -- CLASS III
===================
_______________
PRODUCT        Granulocytes-Platelets, Pheresis.  Recall #B-247-5.
CODE           Unit #17GS65645.
MANUFACTURER   American Red Cross Blood Services, St. Paul, Minnesota.
RECALLED BY    Manufacturer, by telephone October 31, 1994, followed by
               letter November 3, 1994.  Firm-initiated recall complete.
DISTRIBUTION   Minnesota.
QUANTITY       1 unit.
REASON         Blood product, which was improperly tested for antibody to
               human immunodeficiency virus types 1 and 2, (anti-HIV) when
               an incorrect conjugate dispenser was used to add conjugate,
               was distributed. 

_______________
PRODUCT        Recovered Plasma.  Recall #B-493-5.
CODE           Unit #12FZ16639.
MANUFACTURER   American National Red Cross, Charlotte, North Carolina.
RECALLED BY    Manufacturer, by letters dated February 24, 1995.  Firm-
               initiated recall complete.
DISTRIBUTION   North Carolina, Switzerland.
QUANTITY       1 unit.
REASON         Blood products, which were collected from a donor with a
               medical history of cancer, were distributed.

_______________
PRODUCT        Recovered Plasma.  Recall #B-495-5.
CODE           Unit #32GQ91533.
MANUFACTURER   American National Red Cross, Madison, Wisconsin.
RECALLED BY    Manufacturer, by letters dated July 13 or 14, 1994.  Firm-
               initiated recall complete.
DISTRIBUTION   Missouri, Switzerland.
QUANTITY       1 unit.
REASON         Blood products, which were collected from a donor with a
               medical history of cancer, were distributed.

_______________
PRODUCT        (a) Red Blood Cells; (b) Platelets; (c) Fresh Frozen Plasma;
               (d) Recovered Plasma.  Recall #B-504/507-5.
CODE           Unit numbers:  (a) 29108-5721, 29108-5789, 29117-5514,
               29116-6810, 29115-9953; (b) 29108-5721, 29108-5789; (c)
               29117-5514, 29116-6810, 29115-9953; (d) 29108-5721, 29108-
               5789.
MANUFACTURER   United Blood Services, Chicago, Illinois.
RECALLED BY    Manufacturer, by letters dated May 24, 25, 31, 1995.  Firm-
               initiated recall ongoing.
DISTRIBUTION   Illinois, Nevada, Switzerland.
QUANTITY       (a) 5 units; (b) 2 units; (c) 3 units; (d) 2 units.
REASON         Blood products, which were incorrectly tested for antibody
               to human immunodeficiency virus types 1 and 2 (anti-HIV) or
               hepatitis B surface antigen (HBsAg), were distributed.


RECALLS AND FIELD CORRECTIONS:  DEVICES -- CLASS II
=======================
_______________
PRODUCT        Apnea Monitors (multiple models).  Recall #Z-544-5.
CODE           Units with unprotected (exposed pin style) lead wires and
               unprotected patient cables.
MANUFACTURER   Medical Data Electronics, Arleta, California.
RECALLED BY    Manufacturer, by letter on or about April 10, 1995.  Firm-
               initiated recall ongoing.
DISTRIBUTION   Nationwide and international.
QUANTITY       2,244 units.
REASON         Device labeling fails to provide adequate directions for use
               and the devices pose an unwarranted risk of injury to
               patients, because the product is not labeled to warn against
               the hazards of unprotected electrode lead wires being
               connected to electrical power sources.

_______________
PRODUCT        Apnea Monitors (multiple models) and Patient
               Cables/Electrodes.  Recall #Z-935/936-5.
CODE           Apnea monitors with unprotected (exposed pin style) lead
               wires and unprotected patient cables.
MANUFACTURER   Nihon Kohden America, Inc., Irvine, California.
RECALLED BY    Manufacturer, by letters sent beginning June 30, 1995. 
               Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide, The Bahamas, Canada.
QUANTITY       Undetermined.
REASON         Device labeling fails to provide adequate directions for use
               and the devices pose an unwarranted risk of injury to
               patients, because the product is not labeled to warn against
               the hazards of unprotected electrode lead wires being
               connected to electrical power sources.

_______________
PRODUCT        UTS brand of the PGS-3000 Portable Pulsed Galvanic
               Stimulator, designed to aid in the healing of soft tissue
               injuries.  Recall #Z-943-5.
CODE           All serial numbers.
MANUFACTURER   Universal Technology Systems, Inc., (UTS), Jacksonville,
               Florida.
RECALLED BY    Manufacturer, by letter March 23, 1994.  Firm-initiated
               field correction complete.
DISTRIBUTION   Nationwide and Switzerland.
QUANTITY       Approximately 3000-3500 units were distributed; firm
               estimated that all units were corrected.
REASON         The "positive-negative" switch was labeled in reverse order
               on the "hot" stamped plastic lower shell housing the
               devices.

_______________
PRODUCT        Non Rigid Plastic Protective Sleeves and Promotional
               Material.  Products are accessories for the Ultra-Cam I
               Intraoral Camera System and the Ultra-Cam II Intraoral
               Camera System:  (a) Part #DV-SSU, Non Rigid Plastic
               Protective Sleeves; (b) Part #DVSSUII (1994 Promotional
               Material) or V-200006 (1995 Promotional Material) Rigid
               Plastic Sleeves; (c) Part #DV-SSUIIP (1995 Promotional
               Material), Optically Clear Rigid Sterile Sleeves, in 100
               count boxes.  
               Recall #Z-1024/1026-5.
CODE           All units.
MANUFACTURER   Dental Vision Direct, Inc., Carrolton, Texas.
RECALLED BY    Manufacturer, by letter June 6, 1995.  Firm-initiated field
               correction ongoing.
DISTRIBUTION   Nationwide and international.
QUANTITY       Undetermined.
REASON         Device labeling describes the sleeves as sterile, however,
               the sleeves are not required to be sterile.

_______________
PRODUCT        Diagnost 76A Plus, Multipurpose Fluoroscopic and
               Radiographic System with Fluoroscopic Footswitch 9804 763
               10002.  Recall #Z-1027-5.
CODE           All code.
MANUFACTURER   Philips Medezine Systeme, Hamburg, Germany.
RECALLED BY    Philips Medical Systems, Shelton, Connecticut, by conducting
               field corrections per Field Change Order number FCO 00 261
               001, dated 9/94.  Firm-initiated field correction ongoing.
DISTRIBUTION   Nationwide, Canada.
QUANTITY       94 units were distributed; 75 units had been corrected at
               time of recall initiation.
REASON         Incorrect wiring of the footswitch causes unintended
               exposure when the switch is grounded on contact and the DSI
               (spot film) is in place.

_______________
PRODUCT        Philips' Mobile C-Arm Fluoroscopic Diagnostic X-ray System
               with Mobile Viewing Station: (a) Model BV25; 
               (b) Model BV25 N/HR; (c) Model BV25 FAM N/HR;
               (d) Model BV26.  Recall #Z-1031/1034-5.
CODE           BV25 System:  Serial numbers CF 1162 through CF 1692, CP
               0467 through CP 1704, CA 0001 through CA 0581, CE 0001
               through CE 0082, any serial number starting with CM, any
               serial number with letters CF less than serial number 1162,
               and any serial number with letters CP less than serial
               number 0467; BV26 System:  Serial numbers CB 0001 through CB
               0137, and CD 0001 through CD 0024.
MANUFACTURER   Philips Medical Systems, Best, Netherlands.

RECALLED BY    Philips Medical Systems, Shelton, Connecticut, by
               implementing field change order numbers 07 383 001/002 and
               07 387 001.  Firm-initiated field correction ongoing.
DISTRIBUTION   Nationwide and Canada.
QUANTITY       (a) 450 units; (b) 50 units were distributed.
REASON         The trolley wheel may bend or break causing the trolley
               (cart that contains the x-ray controls and mobile viewing
               station) to tilt or topple over.

_______________
PRODUCT        Tomascan CX and CX/S Computed Tomography System, a third
               generation tomography system which measures different tissue
               characteristics and displays the image with analytical data
               on a video display screen:
               (a) Tomoscan CX CT System with software ver. 3.0A, 3.1A,
               3.1B and 3.1C;
               (b) Tomoscan CX/S CT System with software ver. 3.0A, 3.1A,
               3.1B, and 3.1C.  Recall #Z-1035/1036.
CODE           All lots.
MANUFACTURER   Philips Medical Systems Nederlands, Best, Netherlands.
RECALLED BY    Philips Medical Systems, Shelton, Connecticut, by field
               change order (FCO) 08 703 004 dated August 1993, and 
               FCO # 08 703 006 October 1994.  Firm-initiated field
               correction ongoing.
DISTRIBUTION   Nationwide and Puerto Rico.
QUANTITY       Thomoscan CX and CX/s: 135 systems; Thomoscan CX and CX/s
               with Fast Reconstruction or Plus Option:  55 units.
REASON         Patient data entry function errors may occur. 

_______________
PRODUCT        Bard Urine Meter Foley Tray.  Device consists of a tray with
               foley catheter connected to a urine meter.  
               Recall #Z-1042-5.
CODE           Lot #76HE2619 EXP 5/97.
MANUFACTURER   Productos Para El Cuidado De La Salud, Sonora, Mexico.
RECALLED BY    C.R. Bard, Inc., Covington, Georgia, by telephone October
               31, 1994, followed by letter November 15, 1994.  Firm-
               initiated field correction complete.
DISTRIBUTION   Alabama.
QUANTITY       1 case containing 10 units was distributed.
REASON         One unknown case in the lot was not sterilized.

_______________
PRODUCT        Acrysof Intraocular Lens distributed in Acrypak packages. 
               Recall #Z-1045-5.
CODE           All serial numbers in the AcryPak containers.  Units
               packaged in the "Wagon Wheel" case are not involved in the
               recall.
MANUFACTURER   Alcon Laboratories, Inc., Fort Worth, Texas.
RECALLED BY    Alcon Laboratories, Inc./CILCO, Huntington, West Virginia,
               by letters sent beginning April 3, 1995.  Firm-initiated
               recall ongoing.
DISTRIBUTION   Nationwide.
QUANTITY       Undetermined.
REASON         Opacities are observed in the lens after the lens had been
               surgically implanted.

_______________
PRODUCT        Rusch Sterile Catheter Irrigation Tray w/Piston Syringe, a
               flexible tubular device that is inserted through the urethra
               and used to pass fluid to or from the urinary tract.
               Recall #Z-1044-5.
CODE           Catalog #68895, Serial numbers:  3230A/C, 3231A/C, 3253A/c,
               3256A/c, 3196A/C.
MANUFACTURER   Rusch, Inc., Duluth, Georgia.
RECALLED BY    Manufacturer, by letter Septemer 23, 1994.  Firm-initiated
               recall complete.
DISTRIBUTION   Nationwide.
QUANTITY       43,900 trays were distributed; frim estimates none remains
               on the market.
REASON         There is a possibility of tray lid seal(s) failing,
               therefore compromising the sterility of the device.

_______________
PRODUCT        4.5 "Imagecath" Coronary Angioscope, Model COV45, indicated
               for the pre-procedural or post-procedural visualization and
               diagnosis of vascular disease within the peripheral and
               coronary vessels.  Recall #Z-1046-5.
CODE           Lot numbers:  4G059, 4H1118, 4I1267, 4J1414, 4L1627, 4L1640,
               5A0089.
MANUFACTURER   Edwards Interventional Cardiology Division, Irvine,
               California.
RECALLED BY    Manufacturer, by letter dated February 22, 1995.  Firm-
               initiated recall complete.
DISTRIBUTION   Arizona, California, Florida, Georgia, Illinois, Louisiana,
               Michigan, Minnesota, New Jersey, New York, Ohio,
               Pennsylvania, Tennessee, Texas, Wisconsin, Washington, D.C.,
               France, Germany, India, Japan.
QUANTITY       360 units were distributed; no defective units are expected
               to remain in commerce.
REASON         Inter-lumenal tear causes leakage resulting in over-
               inflation of the occlusion cuff.

_______________
PRODUCT        Inflow Valve Conduit of the HeartMate Implantable Pneumatic
               Left Ventricular Assist System (IP LVAS), for use in
               patients on the cardiac transplant list, as a temporary
               mechanical circulatory support for nonreversible left
               ventricular failure as a bridge to cardiac transplantation. 
               Recall #Z-1047-5.
CODE           All codes and serial numbers.
MANUFACTURER   Thermo Cardiosystems, Inc. (TCI), Woburn, Massachusetts.
RECALLED BY    Manufacturer, by letters of June 7, 1995, and July 14, 1995,
               and by telephone between May 31, 1995 and June 23, 1995. 
               Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide and international.
QUANTITY       281 units were distributed.
REASON         Abrasion between the inflow valve conduit and the metal
               graft cage portions of the device may result in the
               development of holes in the valve portion of the device.

_______________
PRODUCT        Ceiling Suspension unit 4153 240 0640 used in conjunction
               with SL Series Multi-Energy Linear Accelerator, for use in
               radiotherapy with ceiling suspension unit 4513 240 0640:
               (a) Localizer/Simulator; (b) SL75-5; (c) SL75-20.  
               Recall #Z-1048/1050-5.
CODE           (a) All serial numbers lower than 10.010, and all K series;
               (b) Serial numbers 300-364; (c) Serial numbers 200-274.
MANUFACTURER   Philips Medical Systems Radiotherapy, West Sussex, England.
RECALLED BY    Philips Medical Systems, Shelton, Connecticut, by visit on
               or about August 1994.  Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide.
QUANTITY       (a) 2 systems; (b) 5 systems; (c) 9 systems were
               distributed.
REASON         A weld failure on the ceiling suspension unit may cause the
               ceiling suspension to fall part way to the floor.

_______________
PRODUCT        (a) Ponsky "Pull" Peg Tray, 28 Fr; (b) Bard Guidewire System
               PEG, 28 Fr; (c) PEG Support Tray.  Recall #Z-1051/1053-5.
CODE           (a) Catalog No. 000730, Lot Nos. 881E0231, 88KE0105,
               88LE0080;
               (b) Catalog No. 000731, Lot Nos. 88JE0181, 88LE0081;
               (c) Catalog No. 000300, Lot Nos. 88GE0096,
               88LE0285, 88CF0218, 88HE0189, 88AF0180, 88IE0233,
               88AF0290, 88JE0381, 88BF0199.
MANUFCTURER    Bard Interventional Products, Mentor, Ohio.
RECALLED BY    Bard Interventional Products, C.R. Bard Inc., Products Div.,
               Tewksbury, Massachusetts, by FeDex letter May 22, 1995. 
               Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide.
QUANTITY       1,866 units.
REASON         The tray seals may not be adequately sealed, therefore
               compromising the sterility of the devices.

_______________
PRODUCT        Bi-Ventricular Support System 5000, a mechanical circulatory
               support system for use in patients suffering from
               postcardiotomy ventricular dysfunction.  Recall #Z-1054-5.
CODE           Catalog #0005-000.  All serial numbers.
MANUFACTURER   Abiomed, Inc., Danvers, Massachusetts.
RECALLED BY    Manufacturer, by letter June 21, 1995.  Firm-initiated
               recall ongoing.
DISTRIBUTION   Nationwide and international.
QUANTITY       230 units.
REASON         High levels of electrostatic discharge may cause
               inappropriate initiation of the fixed rate back-up system or
               othr anomalous console behavior.  The console may start
               pumping in emergency system, a "System Failure" message may
               appear on the display panel, or it may cease pumping.

_______________
PRODUCT        Chest Drain In-Line Connector used in 5 models of the
               Atrium-Water Seal Chest Drain:  
               (a) Single Collection Drain Model numbers 2002-000, 2002-
               100, 2002-070; (b) Dual Collection Drain Model numbers 2020-
               000, 2020-003; (c) Blood Recovery Chest Drain, Model numbers
               2050-000, 2050-100, 2050-070; (d) Dual Blood Recovery Chest
               Drain, Model numbers 2052-000; (e) Multipurpose Chest Drain,
               Model numbers 2060-000, 2060-100.  Recall #Z-1063-5.
CODE           Model #s                             Lot #s.
              (a) 2002-000, 2002-100, 2002-070      502065-000
                                                    502091-000
                                                    502098-000
                                                    502123-000
                                                    502125-000
                                                    502133-000
                                                    502134-000
                                                    502068-100
                                                    502104-100
                                                    502127-100

               (b) 2020-000, 2020-003               520089-000
                                                    520130-000
                                                    520131-003

               (c) 2050-000, 2050-100, 2050-070     550082-000
                                                    550083-070
                                                    550095-000
                                                    550096-100
                                                    550101-070
                                                    550121-100
                                                    550119-000
                                                    550141-000

               (d) 2052-100                         552079-000
                                                    552108-000

               (e) 2060-000, 2060-100               560085-000
                                                    560105-000
                                                    560106-100

MANUFACTURER   Atrium Medical Corporation, Hudson, New Hampshire.
RECALLED BY    Atrium Medical Corporation, Hudson, New Hampshire, by letter
               June 22, 1995.  Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide and international
QUANTITY       39,372 units were distributed.
REASON         The affected lots were assembled with a defective silicone
               o-ring which may allow room air to leak into the drainage
               tube between the patient and the chest drain.

_______________
PRODUCT        Optional Tract Mounting Light Systems manufctured with
               support arms from lot G0741AW:
               (a) Polaris Surgical Light System;
               (b) Gemini Surgical Lights;
               (c) Examiner 10 Lights.  Recall #Z-1069/1071-5.
CODE           Systems manufactured with support arms from lot G074AW.
MANUFACTURER   American Sterilizer Company, Montgomery, Alabama.
RECALLED BY    American Sterilizer Company, Erie, Pennsylvania, by
               telephone December 29, 1994.  Firm-initiated recall
               complete.
DISTRIBUTION   Nationwide and international.
QUANTITY       75 units were distributed.  None remain in commerce.
REASON         There is a potential for the lights to fall due to failure
               of the support arms from lot G074AW to meet tensile strength
               specifications.


RECALLS AND FIELD CORRECTIONS:  DEVICES -- CLASS III
======================
_______________
PRODUCT        Exactech brand 6.5 x 40 mm Titanium Bone Screw, Catalog
               #SC65-40.  Recall #Z-923-5.
CODE           Lot #824, Serial numbers:  QLJV-QLJZ, QLKA-QLKC, QLKG, QLKI,
               QLKP.
MANUFACTURER   Exactech, Inc., Gainesville, Florida.
RECALLED BY    Manufacturer, by telephone April 16, 1993, followed by
               letter November 9, 1993.  Firm-initiated recall complete.
DISTRIBUTION   Florida.
QUANTITY       21 screws were distributed; firm estimaes none remains on
               the market.
REASON         Bone screw may fracture.

_______________
PRODUCT        SR1 Estradiol, an in-vitro diagnostic for the quantitative
               determination of estradiol in human serum or heparinized
               plasma without extraction.  Recall #Z-1043-5.
CODE           Lot #1238 EXP 8/95.
MANUFACTURER   Biodata S.P.A., Rome, Italy.
RECALLED BY    Biochem Immunosystems (U.S.), Inc., Allentown, Pennsylvania,
               by letter dated May 11, 1995, and by telephone on May 11 and
               12, 1995.  Firm-initiated recall ongoing.
DISTRIBUTION   Nationwide and Canada.
QUANTITY       365 boxes (32 cartridges per box) were distributed; firm
               estimaed that 50 boxes remained on market at time of recall
               initiation.
REASON         There is a potential for aberrant (low) patient results,
               which may be significantly contrary to the expected clinical
               condition of the female patient.


RECALLS AND FIELD CORRECTIONS:  VETERINARY PRODUCTS -- CLASS II
===========
_______________
PRODUCT        Bulk Beef Pro 6SBH Feedlot Premix, Category I Type B
               Medicated Article containing lasalocid/melengestrol acetate
               for increased weight gain, feed efficiency and suppression
               of estrus in heifers.  Recall #V-038-5.
CODE           Lot #PB1005-3.
MANUFACTURER   The Miller Company, Omaha, Nebraska.
RECALLED BY    Manufacturer, by telephone February 13 and 14, 1995.  Firm-
               initiated recall complete.
DISTRIBUTION   Iowa.
QUANTITY       2 tons were distributed.
REASON         Product is superpotent.


SEIZURES:
=================================================================
_______________
PRODUCT        Lubricated condoms (94-718-522 et al).
CHARGE         Adulterated - The articles are class III medical devices for
               which there is no approved premarket approval application in
               effect; and, the articles' purity falls below that which
               they purport and are represented to possess.  Misbranded -
               The articles' packaging fails to bear a label containing the
               name and place of business of the manufacturer, packer, or
               distributor; and no notices respecting the devices were pro-
               vided to FDA as required.
LOCATION       Handy Care, Paramount, California.
FILED          May 8, 1995; U.S. District Court for the Central District of
               California; Civ. #CV95-3074 RSWL; FDC #67077.
SEIZED         June 8, 1995 - goods valued at approximately $173,126.

_______________
PRODUCT        Raw Frozen Whole Tuna (95-714-095).
CHARGE         Adulterated - The article consists in part of decomposed
               tuna.
LOCATION       Sea-Land Service, Inc., Long Beach, California.
FILED          June 26, 1995; U.S. District Court for the Central District
               of California; Civ. #95-4257 KN(JRx); FDC #67095.
SEIZED         July 20, 1995 - goods valued at approximately $10,000.

_______________
PRODUCT        Condensed Milk (LECHERA leche condensada azucarada) (95-741-
               379).
CHARGES        Misbranded - The article's label lacks an English language
               statement of identity, declaration of net quantity of
               contents, and statement of ingredients prominently and
               conspicuously placed.  The label further lacks:  the name
               and place of business of the manufacturer, packer, or
               distributor in the required size of type; and, the required
               nutrition information.
LOCATION       La Bodega, Inc., Chicago, Illinois.
FILED          July 18, 1995; U.S. District Court for the Northern District
               of Illinois, Eastern Division; Civ. #95C 4148; FDC 67098.
SEIZED         July 20, 1995 - goods valued at approximately $2,692.

_______________
PRODUCT        Dinnerware (94-718-522 et al).
CHARGE         Adulterated - The articles contain lead, an unsafe food
               additive.
LOCATION       Mid-America Tablewares, Inc., Eau Claire, Wisconsin.
FILED          May 23, 1995; U.S. District Court for the Western District
               of Wisconsin; Civ. #95C 0357C; FDC #67086.
SEIZED         June 8, 1995 - goods valued at approximately $173,126.


END OF ENFORCEMENT REPORT FOR AUGUST 9, 1995.  BLANK PAGES MAY
FOLLOW.

                                   ####