Mechanical Characterization of Multiplicity Vertex Detector Truss Structure

Debbie Clark, Jan Boissevain, Jehanne Simon-Gillo, Jennifer Lock
P-25, Los Alamos National Laboratory
(submitted: March 9, 1995)
PHENIX-MVD-95-3
PHENIX Note #222

We built a prototype of the MVD truss structure to test the mechanical stability, measure the deflection characteristics and review the design concept. Our studies led to a a new design of the mounting scheme, major changes in the endplate design and a selection of thin-walled aluminum tubing.

1. Objectives

We obtained models of the endplates and tubing in order to build a prototype that resembled a Multiplicity Vertex Detector (MVD) assembly. We wanted to measure the sagging of the structure as a whole. At the same time, because of background considerations, we wanted to minimize the mass of the truss structure while maintaining mechanical stability. We also wanted to minimize any measured deflections by either modifying the design of the truss structure and/or the mounting scheme of the MVD. We compared our deflection measurements with calculations using standard engineering methods.

2. Test Set-up

We measured the deflection of various tube sizes, both individually and within the prototype truss structure. We chose to start with aluminum tubing to see if the aluminum was a possible technology choice in terms of mechanical stability and as an affordable material for prototype tests. We tested the following tubes:

Threaded inserts were glued into each tubing size using a 24 hour epoxy (3M Scotch Brand), and allowed to cure for two days before the tests were started. The items were then bolted to an immobile structure (steel I beam). Various weights were placed on the structure and the resulting deflection was measured with a 1" gauge (0.001" accuracy). We also measured torque-induced displacement of the assembly by hanging weights from the outside radius of the endplate. The test set-up for the torque and deflection tests are shown in Figure 1.

Figure 1

Figure 1: Part (a) shows a torque setup with a weight placed on the outside radius of the endplate and the displacement is measured with perpendicular gauges. (b) A deflection setup for a singele tube is shown. Deflections measures were made for two tubes coupled with a metal stip as show in part (c).

The maximum deflection, y, was calculated and compared to lab measurements:

	Equation 1

where w is the uniform load (lb/inches), l is the length (inches), E is the modulus of elasticity, and I is the moment of inertia.

3. Measurements

The following measurements were taken with the test setup and proceedure described in Section 2:

3.1. Test deflection of single tubes a & b in cantilever.

The deflection tests varied from the engineering calculations. There was an unexplained long-term fade component; that is, the deflection measurement increased with time. Elasticity in the glue joint is suspected to be responsible for the fade.

We removed the inserts from the tubes and then repeated the measurements. The measured results agree with the calculations for the tubing without the inserts. Then we removed the initial set of inserts from the first set of tubes and reapplied the adhesive (same brand). This time we allowed the tube/ inserts glue joints to cure one week before repeating the original tests. The results did improve but still contained a much-reduced fade component.

Figure 2 a and b show the results of these tests; vertical displacement is plotted as a function of the weight applied to the structure. Both measurements and calculation of plain tube (27" in length) as well as the measurement from the tube with inserts (29" in length) after a one week cure time are shown in the plot.

The tube with inserts show significantly more deflection than the tubes without insert. The thicker-walled tubing is more mechanically stable than the thin-walled tube.

3.2. Test torque of truss structure with two perpendicular gauges.

It was found that torque-induced vibrations damp slowly on a thin-walled tube; the thick-wall tube had a better response. We could combat any excess torque vibrations by removing material from the outside radius of the endplates and lightening the aluminum pieces wherever possible. The center of gravity of the initial version of the endplate is 73.5 mm from the tube centerline; this could be changed to approximately 50 mm. Figure 3 shows a vertical versus a horizontal torque displacement for various weights and using a thick-wall tube. The plot indicates the sensitivity of the truss structure to torque-induced displacement and further supports the desire to make the endplate an open structure.

3.3. Test deflection of truss structure with double tube a&b in cantilever.

The measured deflection of the assembly is shown in Figure 4 a & b. These tests were performed with a metal strip attaching the tubes instead of full endplates (as shown in figure 1) as a result of the tests described in section 3.2. The tubes include inserts which have been glued on and allowed to cure for one week. There is no significant difference in using a thick-walled tubing over a thin-walled tubing.

4. Conclusions

Weld inserts rather than glue to eliminate stretch in joints.
With the exception of the added deflection due to the weight of the inserts, we expect the welded tube to behave like a single tube without inserts.
Reduce torsion by removing outside radius material from end plates and moving center of gravity as close as possible to the axis of the support tubes.
This is a major change in the design of the truss structure and also reduces the mass of the assembly. The improved design of the truss structure is shown in Figure 5.
Support MVD at both ends to eliminate deflection due to cantilever.
This represents a major design change in the mounting scheme. Figure 6 shows the effect of the new mounting scheme design; the magnitude of any deflection should be reduced by a factor of ten.
Thin-walled aluminum tubing is an acceptable material choice for the truss structure tubing in terms of mechnical stability.
Its effect on background still needs to be investigated, and may not be acceptable. Regardless, the tests indicate that a thick-walled tube is not necessary to maintain mechanical stability, especially in lieu of the new mounting scheme design which minimizes deflections.
Figure 2a

Figure 2a) Deflection for 0.020" tube.

Figure 2b

Figure 2b) Deflection measurements for 0.060" tube.

Figure 3

Figure 3. Torque-induced vibration test.

Figure 4

Figure 4. Deflection tests of double-tube truss structure

Figure 5

Figure 5. The conceptual design of the truss structure is shown at the top of the figure. The improved design with the endplate as an open structure is shown at the bottom.

Figure 6

Figure 6. Demonstration of cantilever beam versus a beam supported at both ends.