Medical Device Recalls
Class 2 Recall
SERVO-i
|
|
Date Recall Initiated |
May 23, 2006
|
Date Posted |
January 25, 2007
|
Recall Number |
Z-0372-2007
|
Product |
Ventilator System SERVO-i V3.1 PC1777
Maquet Critical Care
|
Code Information |
Article # 64-87-800
Serial numbers 02400 - 03300
|
Recalling Firm/ Manufacturer |
MAQUET Inc.
1140 US Highway 22
Bridgewater
, New Jersey
08807-2958
|
For Addition Information Contact |
Jamie Yieh
908-947-2311
|
Reason For Recall |
Reported failures of the SERVO-i ventilators user interface system causing technical errors and increased fault frequencies. Switches may be activated unintentionally, or a function not activate when the switch is selected. Direct access knobs may not function or change the parameter in the wrong direction compared to how they are turned.
|
Action |
Maquet sent an ''Extended Warranty'' letter on May 23, 2006. On March 13, 2007, Maquet issued a Recall Notification letter to all customers stating that they will replace all PC1777 boards in the serial number range 2400-3300, which have not previously been replaced. A safety alert letter was included outlining the failure indication and effect for the user of the failures, as well as, how to identify whether the Servo-i unit has shown any failure indications.
|
Quantity in Commerce |
1, 970 units
|
Distribution |
Worldwide
|
|
|
|
|
|