Medical Device Recalls
Class 2 Recall
LumiWave Infrared lamp
|
|
Date Recall Initiated |
June 14, 2006
|
Date Posted |
August 24, 2006
|
Recall Number |
Z-1419-06
|
Product |
LumiWave 1X4 Infrared Therapy Device, Catalog # LW1X4, BioCare Systems, Inc.
|
Code Information |
Catalog # LW1X4, all units, all lots
|
Recalling Firm/ Manufacturer |
Biocare Systems Inc.
9853 Paperflower Drive
Parker
, Colorado
80138
|
For Addition Information Contact |
Ms. Lanaya Reiter
303-796-7043
|
Reason For Recall |
Electrical Safety Hazard-when pulling on the cord of the AC adapter, to remove it from an outlet, the prongs on the AC adapter, may detach from the unit, damaging it and exposing the user to an electrical safety hazard
|
Action |
Consignees were notified by telephone beginning 06/08/2006, followed by a letter on 06/23/2006.
|
Quantity in Commerce |
175 units
|
Distribution |
Nationwide distribution --- including the states of CA, CO, IL, MA, MD, MI, NC, FL, PA, TX, UT, WA.
|
|
|
|
|
|