USGS logo


Water Resources--Office of Water Quality

This document is also available in pdf format:

 Chapter 6.6.4.pdf

6.6.4
MEASUREMENT

Alkalinity, ANC, and concentrations of bicarbonate, carbonate, and hydroxide species are determined using either the inflection point titration (IPT) method or the Gran function plot (Gran) method to analyze the titration data. Be familiar with the information in 6.6.4.B (IPT method) and 6.6.4.C (Gran method) before selecting the method to be used and before starting the sample titration.

arrow The inflection point titration (IPT) method is adequate for most waters and study needs. Difficulty in identifying the inflection points using an IPT method increases as the ratio of organic acids to carbonate species increases.

arrow The Gran method is recommended for water in which the alkalinity or ANC is expected to be less than about 0.4 meq/L (20 mg/L as CaCO3), or in which conductivity is less than 100 µS/cm, or if there are appreciable noncarbonate contributors or measurable concentrations of organic acids.

The IPT and Gran methods require electrometric titration of a sample with incremental additions of H2SO4 of specified normality. Suggested combinations of titrant normality and sample volume for various ranges of alkalinity or ANC values are given in table 6.6-2. Generally, 1.600N acid is too strong for most samples and is used at alkalinity or ANC greater than 4.0 meq/L (200 mg/L as CaCO3).

Table 6.6-2

info box


6.6.4.A
TITRATION PROCEDURES

Titration procedures are identical for surface-water and ground-water determinations on filtered or unfiltered aliquots of fresh to saline water samples. Become familiar with the information and detailed instructions for the buret and digital titration systems and the IPT and Gran methods before proceeding with the titration.

TECHNICAL NOTE: Fixed endpoint titration to pH 4.5 is no longer used by the USGS for reported alkalinity values because it can be less accurate than the IPT and Gran methods, particularly at small concentrations of total carbonate species and in water with significant organic and other noncarbonate contributors to alkalinity or ANC.

Before beginning titration, select the titration system to be used.

arrow The digital system is convenient but tends to be less precise or accurate than the buret system because of mechanical inadequacies. Good technique is requisite to produce acceptable results.

arrow The buret system can be cumbersome and fragile in the field, and requires experience to execute with precision and accuracy.

arrow A micrometer buret can achieve accuracy to one-tenth of a mg/L (routine determinations are reported to whole numbers).

Select the size of the volumetric pipet (for alkalinity sample) or the graduated cylinder or digital balance (for unfiltered ANC sample) and the measurement vessel, according to the volume of sample needed.

arrow 50 mL of a sample in a 100-mL beaker is selected for most routine work.

arrow Use 100 mL of a sample in a 150-mL beaker for low concentration of alkalinity or ANC. If the total sample volume is small, you may need to use a sample volume of 25 mL or less, although concentration is low.

arrow Use 25 mL or less of a sample and a 50-mL beaker for high concentration of alkalinity or ANC.

To achieve greater accuracy, use lower normality titrant, decrease the volume of acid increments, and increase the number of increments. Figure 6.6-1 provides a general summary of the electrometric titration procedures for alkalinity or ANC.

When pipetting the sample, a small amount of contents remains in the tip of class A "TD" volumetric pipets--do not blow it out, but use the following procedure:

  1. Suspend the pipet tip vertically in a beaker, touching neither the walls nor the contents of the receiving vessel.

  2. Allow the sample to drain freely until the liquid it contains reaches the bottom of the bulb.

  3. Touch the pipet tip to the beaker wall until the flow from the pipet stops--leave the tip in contact with the beaker wall for an additional 10 seconds after the flow stops.

info box

Figure 6.6-1

When titrating, stirring helps to establish a uniform mixture of sample and titrant and an equilibrium between sensors and sample.

arrow If using a magnetic stirrer, stir the sample slowly and continuously, using the smallest stir bar; avoid creating a vortex and large streaming potentials. If using a digital titrator, keep the delivery tube immersed throughout the procedure but keep the aperture of the tube away from the stir bar to avoid bleeding acid from the tube to the sample between titrant additions.

arrow If swirling the sample to mix, make the pH measurement as the sample becomes quiescent, after each addition of titrant.

arrow Avoid splashing the sample out of the beaker or onto the beaker walls. Droplets on the beaker walls can be rinsed down with deionized water. If you splash the sample out of the beaker you must start over.

If concentrations of contributing carbonate species will be determined--Titrate to pH of about 8.1 carefully, in small increments. If concentrations of contributing carbonate species will not be determined--titrate rapidly at first, adding relatively large acid increments to bring pH to about 5.5; then titrate slowly in small increments.

arrow IPT method. Titrate to at least pH 4.0 (to pH 3.5 if the alkalinity or ANC range is unknown for the water sampled or if the sample contains high concentrations of noncarbonate contributors, such as organic acids).

--
Titrate cautiously at and beyond the expected equivalence points.

--
Let the pH value stabilize before the next addition of titrant.

arrow Gran method. Titrate to pH 3.5, or to pH 3.0 or less if the alkalinity or ANC range is unknown for the waters sampled.

--
It is not necessary to develop incremental points above about pH 5.5 for a Gran determination of the bicarbonate equivalence point.

--
A sufficient number of titration points beyond the equivalence are needed to ensure the accuracy of the calculation.

info box

Quality control (QC)--Verify your ability to reproduce the alkalinity or ANC determinations by repeating the titration periodically on duplicate or triplicate samples. The frequency and distribution of QC determinations are established by study requirements.

Rule of thumb--QC should be no less than every tenth sample. Determination on a filtered sample should be reproducible within ±5 percent when titrating a duplicate aliquot from the same batch of sample filtrate.

arrow For filtered samples with less than 0.4 meq/L (20 mg/L as CaCO3), reproducibility should be between 5 and 10 percent.

arrow If the alkalinity is about 0.02 meq/L or less, differences between duplicate samples are likely to exceed 10 percent using the standard titration methods because of rounding error alone.

When interferences are absent, titration on an unfiltered sample often results in a determination identical to or within 5 percent of the filtered sample and can be used as the QC check. If filtered and unfiltered values fail the ±5-percent criterion, repeat the titration on a replicate aliquot of filtered sample.

Reproducibility of the ANC determination to within 5 percent on duplicate unfiltered samples can be problematical when the sample has large amounts of particulate matter--extend the quality-assurance criterion to ±10 percent.

Calculation

Use the following equation to calculate the alkalinity or ANC in milliequivalents per liter:

Equation Formula


Buret titrator

When using a buret, exercise caution to ensure that the acid does not evaporate or become contaminated by extrinsic matter or moisture. The titrant temperature should be equilibrated to the sample temperature before use. Always empty the buret after each use. Never reuse the titrant solution; dispose of the solution properly.

  1. Fill a clean, dry buret with 0.01639N sulfuric acid titrant.

  2. Make sure that no air bubbles are trapped in buret or buret stopcock. Record the sulfuric acid normality and initial buret reading on field forms.

  3. Pipet the selected volume of sample to a clean beaker. Do not pipet by mouth.

  4. Rinse the pH and temperature sensors with DIW. Blot water droplets adhering to the sensors with lint-free paper (residual DIW will not affect the determination).

  5. Insert the sensors in the beaker.

  6. Measure the initial pH and temperature while gently stirring or after gently swirling the sample.

  7. Begin titration. If using a magnetic stirrer, stir the sample slowly and continuously. Measure pH after each addition of titrant, and after the acid and sample are mixed homogeneously. If a magnetic stirrer is not used, swirl to mix the sample and acid after each addition of titrant. Allow 15 to 20 seconds after each addition for equilibration, then record pH.

    TECHNICAL NOTE: 0.01 mL of a standard 0.05-mL drop of titrant tends to remain on the buret tip. To dispense this 0.01-mL titrant drop, quickly rotate the stopcock through 180 degrees (one-half turn) and then rinse the titrant from the buret tip into the filtration beaker with a small quantity of DIW (C.J. Patton, U.S. Geological Survey, written commun., 1995).

Calculation

Use the following equations to calculate alkalinity or ANC and carbonate species from inflection points with 0.01639N sulfuric acid:

Equation Formula

The presence of hydroxide is indicated when the carbonate titrant volume exceeds the bicarbonate titrant volume.


Digital titrator

It is necessary to be thoroughly familiar with the operation of the digital titrator before field use. A plunger in the digital titrator forces acid in the titrant cartridge into the delivery tube. The plunger is controlled by a main-drive screw, which in turn is controlled by rotation of the delivery knob. The delivery knob controls the volume of titrant delivered through the delivery tube, as indicated by a digital counter.

  1. Record the sample volume to be titrated and the titrant normality. Equilibrate titrant temperature to sample temperature.

  2. Assemble the digital titrator.

    1. Depress the plunger-release button and retract the plunger.

    2. Insert the titrant cartridge into the titrator and twist the cartridge one-quarter turn to lock it into position.

    3. Carefully depress the plunger-release button and push the plunger forward until it makes contact with the TeflonTM seal inside the cartridge.

  3. Remove the vinyl cap from the cartridge (save the cap) and insert the straight end of the delivery tube into the cartridge.

  4. Purge the titrant through the delivery tube to ensure that no air bubbles or water are in the tube by holding the titrator with the cartridge tip up and turning the delivery knob to force a few drops of titrant through the end of the delivery tube. Rinse tube exterior with DIW and blot off acid or water droplets before inserting it into the sample.

  5. Set the digital counter to zero using the counter-reset knob, taking care not to turn the delivery knob.

  6. Pipet the selected volume of the sample to a clean beaker. If a magnetic stirrer is used, place a clean, dry, small stir bar into the beaker before pipetting the sample to the beaker. Do not use a magnetic stirrer for sample conductivity <100 µS/cm.

  7. Insert sensors into the beaker.

  8. Measure the initial pH and temperature while gently stirring or after gently swirling the sample.

  9. Immerse the end of the titrant delivery tube in the sample. To prevent bleeding of the titrant from the delivery tube, keep the aperture of the delivery tube away from the stir bar.

  10. Begin titration. If using a magnetic stirrer, stir the sample slowly and continuously. Measure pH after each addition of titrant, and after the acid and sample are mixed homogeneously. If a magnetic stirrer is not used, swirl to mix the sample and acid after each addition of the titrant. Allow 15 to 20 seconds after each addition for equilibration, then record pH.

  11. After completing the titration, depress the plunger release, retract the plunger, and remove the titrant cartridge. Immediately replace the vinyl cap on the cartridge tip. Discard the delivery tube after each use.

  12. Calculate alkalinity/ANC in the field--

    Equation Formula


6.6.4.B INFLECTION POINT TITRATION METHOD

The IPT method uses the inflection points of plotted values to select equivalence points instead of assuming equivalence points to be at pH 8.3 and 4.5 (as in the fixed endpoint method). Inflection points are points of maximum rate of change in pH per volume of titrant added. Near equivalence points, rapid pH changes occur with small additions of titrant. For this reason, titration as you near and pass the expected equivalence points must be slow and cautious, using small incremental additions of titrant. Relative error of the determinations can be within ±4 percent if the equivalence point is recognizable within ±0.3 pH unit of the true equivalence point.

Use either the buret or digital titrator. To determine the inflection point, you can either construct a titration curve by plotting the change in pH divided by the change in titrant volume against the incremental volumes of the titrant added to the sample, or tabulate a titration. Figures 6.6-2 and 6.6-3 represent the titration of a sample that has both a carbonate and a bicarbonate inflection point, whereas tables 6.6-4 and 6.6-5 represent a titration that resulted only in a bicarbonate inflection point.

arrow More than one inflection point in close proximity indicates that the true inflection point has been missed. If this occurs, titrate a duplicate sample using smaller acid increments near the inflection point or use a Gran plot.

arrow If no clear point or points can be determined easily, interferences from weak organic acids are likely--use the Gran method.

EXAMPLES:

IPT method using the buret system. Referring to table 6.6-4, pH 4.51 at a titrant volume of 8.95 mL is the point of maximum rate of change of pH per volume titrant. The actual inflection point is before the titrant volume corresponding to the maximum change in pH per unit volume of acid added; therefore, the correct value lies between 8.95 mL and the previous value, 8.90 mL. In this example, the calculated titrant volume 8.93 mL would be the correct inflection point.

IPT method using the digital titrator. Referring to table 6.6-5, pH 4.51 is the point of maximum rate of change of pH per volume of titrant. The actual inflection point, however, is between the digital-counter value (454) representing the maximum change in pH per unit volume of acid added and the previous digital-counter value (452). The correct (calculated) digital-counter value for the inflection point would be 453. The error in computing concentration from the digital-counter value 454 instead of the digital-counter value 453 is considered insignificant. Note, however, that the larger the increments used, the greater the significance of the error. Calculation of the correct inflection point is recommended.

See Errata for Chapter A6 for changes to Table 6.6-4.

Table 6.6-4

Figure 6.6-2

Table 6.6-5

Figure 6.6-3


6.6.4.C GRAN FUNCTION PLOT METHOD

Gran function plots commonly are used to determine alkalinity and ANC in sea water, low ionic-strength water, water with low carbonate concentrations, and water with measurable concentrations of organic compounds. The Gran function plot method also is used for calculations of the base neutralizing capacity in waters of low ionic strength such as atmospheric deposition.

The Gran function plot method uses functions that linearize titration curves, making it possible to determine alkalinity or ANC with a few points rather than relying on the inflection point from an entire titration curve (Baedecker and Cozzarelli, 1992). Four Gran functions can be calculated over the entire titration curve. The F1 function, described below, is the most commonly applicable Gran function.

arrow The F1 function requires titration data beyond the equivalence point, usually to pH between 3.0 and 3.5. For systems with measurable concentrations of organic acids, titrate to pH 2.5 (Baedecker and Cozzarelli, 1992).

arrow The buret titration system is recommended to achieve the accuracy desired when using the Gran method: substitute the equivalence points determined by the Gran function plot into the equations given in the discussion of the buret system. The digital titration system also is used under some circumstances (illustrated on table 6.6-6 and fig. 6.6-4): refer to table 6.6-3 to calculate concentrations.

To construct a Gran function plot:

  1. Titrate incrementally to about pH 3.5. It is not necessary to develop incremental points above about pH 5.5 for a Gran determination of the bicarbonate equivalence point.

  2. Plot (Vo + Vt ) 10-pH against the titrant volume (fig. 6.6-4 and table 6.6-6) using the F1 function for the bicarbonate equivalence point,

    where:

    Vo = volume of the sample

    Vt = volume of the titrant added

    Vs = volume of the titrant needed to reach the bicarbonate equivalence point

    Vw = volume of the titrant needed to reach the carbonate equivalence point.

  3. Extrapolate the straight line in the region beyond the equivalence point to (Vo+Vt ) 10-pH=0 or Vt=Vs.

    arrow Carbonate equivalence point. Plot (Vt-Vs ) 10-pH against the titrant volume (in mL or digital counts). To locate the equivalence point, extrapolate the straight line in the region beyond the equivalence point to (Vt-Vs ) 10-pH=0 or Vt=Vw.

    arrow Hydroxide equivalence point. Plot (Vo-2Vw+Vt ) 10-pH against the titrant volume (in mL or digital counts). To locate the equivalence point, extrapolate the straight line in the region beyond the equivalence point to (Vo+Vt ) 10-pH=0 or Vt=Vw.

    TECHNICAL NOTE: During alkalinity titration (carbonate system), the hydrogen ions added convert carbonate to bicarbonate and then bicarbonate to carbonic acid. The titration continues until no more species are reacting. When this process is complete, additional hydrogen ions will be in excess in the solution. The Gran function plot identifies the point at which all alkalinity has been titrated and hydrogen ions begin to be in excess. Beyond the equivalence point, the shape of the curve is determined by hydrogen ions in excess of all hydronium ion acceptors in the sample. The curvature results from the logarithmic relation between pH and hydrogen-ion activity.

    EXAMPLE:

    Gran function plot method using the digital titrator. Refer to figure 6.6-4 and table 6.6-6. In the region beyond the equivalence point in figure 6.6-4, a straight line results. Extrapolation of this straight line to (Vo=Vt)10-pH=0 or Vt=Vs locates the equivalence point. The extrapolated straight line intercept at (Vo+Vt) 10-pH=0 on figure 6.6-4 is 0.566 mL of titrant added and corresponds to an equivalence point at approximately pH of 4.58.

Figure 6.6-4

Table 6.6-6


nextSection 6.6.5
upReturn to Section 6.6.3
upReturn to Contents for 6.6--Alkalinity and Acid Neutralizing Capacity
upReturn to Chapter A6 Contents Page
upReturn to Field Manual Complete Contents
nextReturn to Water Quality Information Pages

Maintainer: Office of Water Quality
Webversion by: Genevieve Comfort
Last Modified: 7Feb02 imc