Water Resources--Office of Water Quality
This document is also available in pdf format:
- Chapter 6.6.4.pdf
6.6.4
MEASUREMENT
Alkalinity, ANC, and concentrations of bicarbonate, carbonate, and
hydroxide species are determined using either the inflection point
titration (IPT) method or the Gran function plot (Gran) method to analyze
the titration data. Be familiar with the information in 6.6.4.B (IPT method) and 6.6.4.C (Gran method) before selecting the
method to be used and before starting the sample titration.
|
The inflection point titration (IPT) method is
adequate for most waters and study needs. Difficulty in identifying the
inflection points using an IPT method increases as the ratio of organic
acids to carbonate species increases. |
| The Gran method is recommended for
water in which the alkalinity or ANC is expected to be less than about 0.4
meq/L (20 mg/L as CaCO3), or in which
conductivity is less than 100 µS/cm, or if there are appreciable
noncarbonate contributors or measurable concentrations of organic
acids. |
The IPT and Gran methods require electrometric titration of a sample with
incremental additions of H2SO4 of specified normality. Suggested combinations
of titrant normality and sample volume for various ranges of alkalinity or
ANC values are given in table 6.6-2.
Generally, 1.600N acid is too strong for most samples and is used
at alkalinity or ANC greater than 4.0 meq/L (200 mg/L as CaCO3).
6.6.4.A
TITRATION PROCEDURES
Titration procedures are identical for surface-water and ground-water
determinations on filtered or unfiltered aliquots of fresh to saline water
samples. Become familiar with the information and detailed instructions
for the buret and digital titration systems and the IPT and Gran methods
before proceeding with the titration.
TECHNICAL NOTE: Fixed endpoint
titration to pH 4.5 is no longer used by the USGS for reported alkalinity
values because it can be less accurate than the IPT and Gran methods,
particularly at small concentrations of total carbonate species and in
water with significant organic and other noncarbonate contributors to
alkalinity or ANC.
Before beginning titration, select the titration system to be used.
| The
digital system is convenient but tends to be less precise or accurate than
the buret system because of mechanical inadequacies. Good technique is
requisite to produce acceptable results. |
| The buret system can be cumbersome and fragile in
the field, and requires experience to execute with precision and
accuracy. |
| A micrometer buret can achieve accuracy to one-tenth
of a mg/L (routine determinations are reported to whole
numbers). |
Select the size of the volumetric pipet (for alkalinity sample) or the
graduated cylinder or digital balance (for unfiltered ANC sample) and the
measurement vessel, according to the volume of sample needed.
| 50 mL of a
sample in a 100-mL beaker is selected for most routine work. |
| Use 100 mL of a sample in a 150-mL beaker for low
concentration of alkalinity or ANC. If the total sample volume is small,
you may need to use a sample volume of 25 mL or less, although
concentration is low. |
| Use 25 mL or less of a sample and a 50-mL beaker for
high concentration of alkalinity or ANC. |
To achieve greater accuracy, use lower normality titrant, decrease the
volume of acid increments, and increase the number of increments. Figure 6.6-1 provides a general summary of
the electrometric titration procedures for alkalinity or ANC.
When pipetting the sample, a small amount of contents remains in the tip of
class A "TD" volumetric pipets--do not blow it out, but use the following
procedure:
- Suspend the pipet tip vertically in a beaker, touching neither the
walls nor the contents of the receiving vessel.
- Allow the sample to drain freely until the liquid it contains reaches
the bottom of the bulb.
- Touch the pipet tip to the beaker wall until the flow from the pipet
stops--leave the tip in contact with the beaker wall for an additional 10
seconds after the flow stops.
When titrating, stirring helps to establish a uniform mixture of sample and
titrant and an equilibrium between sensors and sample.
| If
using a magnetic stirrer, stir the sample slowly and continuously,
using the smallest stir bar; avoid creating a vortex and large
streaming potentials. If using a digital titrator, keep the delivery tube
immersed throughout the procedure but keep the aperture of the tube away
from the stir bar to avoid bleeding acid from the tube to the sample
between titrant additions. |
| If swirling the sample to mix, make the pH
measurement as the sample becomes quiescent, after each addition of
titrant. |
| Avoid splashing the sample out of the beaker or onto
the beaker walls. Droplets on the beaker walls can be rinsed down with
deionized water. If you splash the sample out of the beaker you must start
over. |
If concentrations of contributing carbonate species will be
determined--Titrate to pH of about 8.1 carefully, in small
increments. If concentrations of contributing carbonate species will not
be determined--titrate rapidly at first, adding relatively large acid
increments to bring pH to about 5.5; then titrate slowly in small
increments.
|
IPT method. Titrate to at least pH 4.0 (to pH 3.5 if
the alkalinity or ANC range is unknown for the water sampled or if the
sample contains high concentrations of noncarbonate contributors, such as
organic acids). - --
- Titrate cautiously at and beyond
the expected equivalence points.
- --
- Let the pH value stabilize
before the next addition of titrant.
|
| Gran method. Titrate to pH 3.5, or
to pH 3.0 or less if the alkalinity or ANC range is unknown for the waters
sampled.
- --
- It is not necessary to develop incremental points
above about pH 5.5 for a Gran determination of the bicarbonate equivalence
point.
- --
- A sufficient number of titration points beyond the
equivalence are needed to ensure the accuracy of the
calculation.
|
Quality control (QC)--Verify your ability to reproduce the
alkalinity or ANC determinations by repeating the titration periodically on
duplicate or triplicate samples. The frequency and distribution of QC
determinations are established by study requirements.
Rule of thumb--QC should be no less than every tenth
sample. Determination on a filtered sample should be reproducible within
±5 percent when titrating a duplicate aliquot from the same batch of
sample filtrate.
| For
filtered samples with less than 0.4 meq/L (20 mg/L as CaCO3), reproducibility should be between 5 and 10
percent.
|
| If the alkalinity is about 0.02 meq/L or less,
differences between duplicate samples are likely to exceed 10 percent using
the standard titration methods because of rounding error
alone. |
When interferences are absent, titration on an unfiltered sample often
results in a determination identical to or within 5 percent of the filtered
sample and can be used as the QC check. If filtered and unfiltered values
fail the ±5-percent criterion, repeat the titration on a replicate
aliquot of filtered sample.
Reproducibility of the ANC determination to within 5 percent on duplicate
unfiltered samples can be problematical when the sample has large amounts
of particulate matter--extend the quality-assurance criterion to ±10
percent.
Calculation
Use the following equation to calculate the alkalinity or ANC in
milliequivalents per liter:
When using a buret, exercise caution to ensure that the acid does not
evaporate or become contaminated by extrinsic matter or moisture. The
titrant temperature should be equilibrated to the sample temperature before
use. Always empty the buret after each use. Never reuse the titrant
solution; dispose of the solution properly.
- Fill a clean, dry buret with 0.01639N sulfuric acid titrant.
- Use a 10-mL semimicroburet with 0.05-mL graduations and a
TeflonTM stopcock for samples with
alkalinity or ANC less than 4 meq/L (200 mg/L as CaCO3).
- Use a 25-mL buret with 0.1-mL graduations and a TeflonTM stopcock for samples with alkalinity or ANC of 4
meq/L (200 mg/L as CaCO3) or greater and
when the sample pH exceeds 8.1.
- If greater accuracy is desired, use a GilmontTM-type micrometer buret.
- Make sure that no air bubbles are trapped in buret or buret stopcock.
Record the sulfuric acid normality and initial buret reading on field
forms.
- Pipet the selected volume of sample to a clean beaker. Do not pipet by
mouth.
- If a magnetic stirrer is used, place a clean, dry, small stir bar into
the beaker before pipetting the sample to the beaker. Do not use a
magnetic stirrer if sample conductivity is <100
µS/cm. Place beaker on stirrer.
- Rinse the pH and temperature sensors with DIW. Blot water droplets
adhering to the sensors with lint-free paper (residual DIW will not affect
the determination).
- Insert the sensors in the beaker.
- Do not let the sensors touch the beaker bottom or wall.
- Sample depth in the beaker must be sufficient to cover the junction of
the reference electrode, the electrode bulb, and the temperature
sensor.
- Measure the initial pH and temperature while gently stirring or after
gently swirling the sample.
- Do not splash the sample onto the beaker wall or out of the beaker.
- Minimize the vortex caused by magnetic stirring.
- Record on the field form the pH and temperature values, the counter
reading (it should read "0000"), the titrant normality, the time, and the
sample size.
- Begin titration. If using a magnetic stirrer, stir the sample slowly
and continuously. Measure pH after each addition of titrant, and after the
acid and sample are mixed homogeneously. If a magnetic stirrer is not
used, swirl to mix the sample and acid after each addition of titrant.
Allow 15 to 20 seconds after each addition for equilibration, then record
pH.
- pH 8.1--To
determine the bicarbonate inflection point, add the titrant drop by drop in
0.01-mL increments with a 25-mL buret until pH is less than 8.0.
- pH < 8.1--Titrate rapidly to pH of about 5.0,
using several large acid increments (to pH of 5.5 for sample alkalinity or
ANC of 0.4 meq/L (<20 mg/L as CaCO3) or
conductivity of sample <100 µS/cm).
- pH <5.0--Cautiously add the titrant drop by drop
in 0.01-mL increments from pH 5.0 to 4.0 or less (the most sensitive part
of the titration curve is usually between pH 4.8 and 4.3). Extend
titration to pH 3.5 or less if using the Gran method or for a sample high
in organic acid concentration.
TECHNICAL NOTE: 0.01 mL of a
standard 0.05-mL drop of titrant tends to remain on the buret tip. To
dispense this 0.01-mL titrant drop, quickly rotate the stopcock through 180
degrees (one-half turn) and then rinse the titrant from the buret tip into
the filtration beaker with a small quantity of DIW (C.J. Patton, U.S.
Geological Survey, written commun., 1995).
Calculation
Use the following equations to calculate alkalinity or ANC and carbonate
species from inflection points with 0.01639N sulfuric acid:
The presence of hydroxide is indicated when the carbonate titrant volume
exceeds the bicarbonate titrant volume.
It is necessary to be thoroughly familiar with the operation of the digital
titrator before field use. A plunger in the digital titrator forces acid
in the titrant cartridge into the delivery tube. The plunger is controlled
by a main-drive screw, which in turn is controlled by rotation of the
delivery knob. The delivery knob controls the volume of titrant delivered
through the delivery tube, as indicated by a digital counter.
- Record the sample volume to be titrated and the titrant normality.
Equilibrate titrant temperature to sample temperature.
- Assemble the digital titrator.
- Depress the plunger-release button and retract the plunger.
- Insert the titrant cartridge into the titrator and twist the cartridge
one-quarter turn to lock it into position.
- Carefully depress the plunger-release button and push the plunger
forward until it makes contact with the TeflonTM seal inside the cartridge.
- Remove the vinyl cap from the cartridge (save the cap) and insert the
straight end of the delivery tube into the cartridge.
- Do not push the delivery tube beyond the cartridge tip.
- Do not alter the delivery tube.
- Tag the delivery tube to avoid cross contamination. Do not interchange
delivery tubes between cartridges with different titrant normality.
- Purge the titrant through the delivery tube to ensure that no air
bubbles or water are in the tube by holding the titrator with the cartridge
tip up and turning the delivery knob to force a few drops of titrant
through the end of the delivery tube. Rinse tube exterior with DIW and
blot off acid or water droplets before inserting it into the sample.
- Set the digital counter to zero using the counter-reset knob, taking
care not to turn the delivery knob.
- Pipet the selected volume of the sample to a clean beaker. If a
magnetic stirrer is used, place a clean, dry, small stir bar into the
beaker before pipetting the sample to the beaker. Do not use a magnetic
stirrer for sample conductivity <100 µS/cm.
- Insert sensors into the beaker.
- Rinse the pH and temperature sensors with DIW. Blot off water droplets
adhering to the sensors.
- Do not let sensors touch the bottom or wall of the beaker.
- The amount of sample in the beaker must be sufficient to cover the
junction of the reference electrode, the electrode bulb, and the
temperature sensor.
- Measure the initial pH and temperature while gently stirring or after
gently swirling the sample.
- Do not splash sample onto beaker wall or out of the beaker.
- Minimize the vortex caused by magnetic stirring.
- Record the pH and temperature values, the counter reading (it should
read "0000"), the titrant normality, the time, and the sample size on field
forms.
- Immerse the end of the titrant delivery tube in the sample. To prevent
bleeding of the titrant from the delivery tube, keep the aperture of the
delivery tube away from the stir bar.
- Begin titration. If using a magnetic stirrer, stir the
sample slowly and continuously. Measure pH after each addition of titrant,
and after the acid and sample are mixed homogeneously. If a magnetic
stirrer is not used, swirl to mix the sample and acid after each addition
of the titrant. Allow 15 to 20 seconds after each addition for
equilibration, then record pH.
- pH8.1--Slowly add
the titrant in replicate increments no greater than two to three digital
counts until pH of the sample is about 8.0, to determine the carbonate
inflection point. Record the pH and digital counter reading after each
addition of the titrant. Larger increments can be used for samples
containing high carbonate concentrations.
- pH <8.1--Titrate rapidly with several large acid
increments to pH of about 5.0 (to pH 5.5 for sample alkalinity at 0.4 meq/L
(<20 mg/L as CaCO3) or sample
conductivity <100 µS/cm).
- pH <5.0--Add titrant cautiously, in increments of
one to three digital counts, from pH 5.0 to 4.0 or less. (The most
sensitive part of the titration curve is between pH 4.8 and 4.3 for many
natural waters.) If using the Gran method, extend the titration to pH 3.5
or less and to pH 3.0 or less for samples high in organic acids and other
noncarbonate contributors, or when the alkalinity or ANC range is
unknown.
- After completing the titration, depress the plunger release, retract
the plunger, and remove the titrant cartridge. Immediately replace the
vinyl cap on the cartridge tip. Discard the delivery tube after each
use.
- Calculate alkalinity/ANC in the field--
- To determine carbonate alkalinity concentrations and concentrations of
contributing species, plot change in pH divided by change in digital counts
against digital counts of the titrant or tabulate change in pH divided by
change in digital counts.
- Refer to table 6.6-3 for digital titration factors.
- The factors and equations used for the 0.1600N or 1.600N titrant
cartridges are as follows (these calculations are based on the same
equations described under the buret titration procedures, except that
milliliters of acid used is shown as digital counts for the HachTM titrator; 800 counts = 1 mL):
6.6.4.B INFLECTION POINT TITRATION METHOD
The IPT method uses the inflection points of plotted values to select
equivalence points instead of assuming equivalence points to be at pH 8.3
and 4.5 (as in the fixed endpoint method). Inflection points are points of
maximum rate of change in pH per volume of titrant added. Near
equivalence points, rapid pH changes occur with small additions of titrant.
For this reason, titration as you near and pass the expected equivalence
points must be slow and cautious, using small incremental additions of
titrant. Relative error of the determinations can be within
±4 percent if the equivalence point is recognizable within ±0.3
pH unit of the true equivalence point.
Use either the buret or digital titrator. To determine the
inflection point, you can either construct a titration curve by plotting
the change in pH divided by the change in titrant volume against the
incremental volumes of the titrant added to the sample, or tabulate a
titration. Figures 6.6-2 and 6.6-3 represent the titration of a sample
that has both a carbonate and a bicarbonate inflection point, whereas tables 6.6-4 and 6.6-5 represent a titration that
resulted only in a bicarbonate inflection point.
| More than
one inflection point in close proximity indicates that the true inflection
point has been missed. If this occurs, titrate a duplicate sample using
smaller acid increments near the inflection point or use a Gran
plot. |
| If no clear point or points can be determined
easily, interferences from weak organic acids are likely--use the Gran
method. |
EXAMPLES:
IPT method using the buret system.
Referring to table 6.6-4, pH 4.51 at a titrant volume of
8.95 mL is the point of maximum rate of change of pH per volume titrant.
The actual inflection point is before the titrant volume corresponding to
the maximum change in pH per unit volume of acid added; therefore, the
correct value lies between 8.95 mL and the previous value, 8.90 mL. In this
example, the calculated titrant volume 8.93 mL would be the correct
inflection point.
IPT method using the digital titrator. Referring to table 6.6-5, pH 4.51 is the point of
maximum rate of change of pH per volume of titrant. The actual inflection
point, however, is between the digital-counter value (454) representing the
maximum change in pH per unit volume of acid added and the previous
digital-counter value (452). The correct (calculated) digital-counter
value for the inflection point would be 453. The error in computing
concentration from the digital-counter value 454 instead of the
digital-counter value 453 is considered insignificant. Note, however, that
the larger the increments used, the greater the significance of the error.
Calculation of the correct inflection point is
recommended.
See Errata for Chapter A6 for
changes to Table 6.6-4.
6.6.4.C GRAN FUNCTION PLOT METHOD
Gran function plots commonly are used to determine alkalinity and ANC in
sea water, low ionic-strength water, water with low carbonate
concentrations, and water with measurable concentrations of organic
compounds. The Gran function plot method also is used for calculations of
the base neutralizing capacity in waters of low ionic strength such as
atmospheric deposition.
The Gran function plot method uses functions that linearize titration
curves, making it possible to determine alkalinity or ANC with a few points
rather than relying on the inflection point from an entire titration curve
(Baedecker and Cozzarelli, 1992). Four Gran functions can be calculated
over the entire titration curve. The F1 function, described below, is the
most commonly applicable Gran function.
| The F1
function requires titration data beyond the equivalence point, usually to
pH between 3.0 and 3.5. For systems with measurable concentrations of
organic acids, titrate to pH 2.5 (Baedecker and Cozzarelli,
1992). |
| The buret titration system is recommended to achieve
the accuracy desired when using the Gran method: substitute the equivalence
points determined by the Gran function plot into the equations given in the
discussion of the buret system. The digital titration system also is used
under some circumstances (illustrated on table 6.6-6 and fig. 6.6-4): refer to table 6.6-3 to calculate
concentrations. |
To construct a Gran function plot:
- Titrate incrementally to about pH 3.5. It is not necessary to develop
incremental points above about pH 5.5 for a Gran determination of the
bicarbonate equivalence point.
- Plot (Vo + Vt
) 10-pH against the titrant
volume (fig. 6.6-4 and table 6.6-6) using the F1 function for the bicarbonate equivalence
point,
where:
Vo = volume of the sample
Vt = volume of the titrant added
Vs = volume of the titrant needed to
reach the bicarbonate equivalence point
Vw = volume of the titrant needed to
reach the carbonate equivalence point.
- Extrapolate the straight line in the region beyond the equivalence
point to (Vo+Vt
) 10-pH=0 or Vt=Vs.
|
Carbonate equivalence point. Plot (Vt-Vs ) 10-pH against the titrant volume (in mL or digital
counts). To locate the equivalence point, extrapolate the straight line in
the region beyond the equivalence point to (Vt-Vs ) 10-pH=0 or Vt=Vw. |
| Hydroxide equivalence point. Plot
(Vo-2Vw+Vt ) 10-pH against the titrant volume (in mL or digital
counts). To locate the equivalence point, extrapolate the straight line in
the region beyond the equivalence point to (Vo+Vt ) 10-pH=0 or Vt=Vw. |
TECHNICAL NOTE: During alkalinity
titration (carbonate system), the hydrogen ions added convert carbonate to
bicarbonate and then bicarbonate to carbonic acid. The titration continues
until no more species are reacting. When this process is complete,
additional hydrogen ions will be in excess in the solution. The Gran
function plot identifies the point at which all alkalinity has been
titrated and hydrogen ions begin to be in excess. Beyond the equivalence
point, the shape of the curve is determined by hydrogen ions in excess of
all hydronium ion acceptors in the sample. The curvature results from the
logarithmic relation between pH and hydrogen-ion activity.
EXAMPLE:
Gran function plot method using the digital titrator.
Refer to figure 6.6-4 and table 6.6-6. In the region beyond the
equivalence point in figure 6.6-4, a
straight line results. Extrapolation of this straight line to (Vo=Vt)10-pH=0 or Vt=Vs locates the equivalence point. The extrapolated
straight line intercept at (Vo+Vt) 10-pH=0 on figure 6.6-4 is 0.566 mL of titrant added
and corresponds to an equivalence point at approximately pH of
4.58.
Section 6.6.5
Return to Section 6.6.3
Return to Contents for 6.6--Alkalinity and Acid
Neutralizing Capacity
Return to Chapter A6 Contents Page
Return to Field Manual Complete Contents
Return to Water Quality Information Pages Maintainer: Office of Water Quality
Webversion by: Genevieve Comfort
Last Modified: 7Feb02 imc