Skip NavigationFDA Logo links to FDA home pageCenter for Devices and Radiological Health, U.S. Food and Drug AdministrationHHS Logo links to Department of Health and Human Services website
FDA Home Page | CDRH Home Page | Search | A-Z Index U.S. Food and Drug AdministrationCenter for Devices and Radiological Health Questions?
horizonal rule
Super Search
510(k) | Registration & Listing | Adverse Events | PMA | Classification | CLIA
CFR Title 21 | Advisory Committees | Assembler | Recalls | Guidance | Standards
 

New Search Back To Search Results
 

Medical Device Recalls
Class 2 Recall
Pulmonary Viewer

See Related Information
Date Recall
Initiated
September 10, 2007
Date Posted December 29, 2007
Recall Number Z-0367-2008
Product Philips Medical Systems, Cardiac Viewer or Pulmonary Viewer Application that can run on all Brilliance CT scanners using software versions 2.0 through V 2.2.5 and all EBW workstations using software versions 2.0 through V3.5.2. Philips Medical Systems, Cleveland, OH 44143. (The product is a software application that is installed in Philips Medical System CT scanners.)
Code Information For 16 slice, 16 slice Big Bore, 40 slice, and 64 slice Brilliance CT scanners the affected software application versions are: 2.0, 2.0.2, 2.2, 2.2.1, 2.2.2, and 2.2.5. For Extended Brilliance Workspace the affected software application versions are: 2.0.0, 2.0.1, 2.0.50, 2.1.0, 3.0.0, 3.0.1, 3.0.2, 3.5, and 3.5.2.
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Road
Cleveland , Ohio 44143-2131
Reason For
Recall
Incorrect Measurement: Incorrect measurement will occur when the operator uses the distance or area measurement function, such as Graphic Tool line or ROI on zoomed and batch saved images from the Cardiac Viewer or Pulmonary Viewer and then views them in the CT Viewer or on a PACS System.
Action Consignees were notified of this recall by an Urgent Device Notification letter sent on 9/10/07. The letter instructed the users to not save any zoomed images to a batch file. Make all measurements of zoomed images in the CT viewer, which provides correct measurements. Do not perform measurements on any zoomed batched images that have been stored on a PACS System. Phillips in investigating the problem and will provide updates free of charge when completed.
Quantity in Commerce 2,544
Distribution Worldwide Distribution - USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WI, WV, and WY, and country of Canada.
 