Beamline U4B
General Information
Source Type Bending Magnet Status Operational General User Beamtime 60% Energy Range 20-1200 eV
Beamline Type Facility Beamline Technique(s) X-ray reflectivity X-ray scattering, magnetic X-ray scattering, resonant Magnetic circular dichroism X-ray absorption spectroscopy X-ray fluorescence spectroscopy
Institution(s) Montana State University
Research Types Soft x-ray and VUV photo absorption spectroscopy. Techniques: X-ray absorption spectroscopy, X-ray magnetic circular dichroism, X-ray fluorescence spectroscopy, x-ray resonant magnetic and/or non-magnetic scattering.
Contact Information
Spokesperson
Dario Arena, Brookhaven National Laboratory, darena@bnl.gov, 344-5303 Yves Idzerda, Montana State University, Idzerda@physics.montana.edu, 406-994-7838
Local Contact
Dario Arena, Brookhaven National Laboratory, darena@bnl.gov, 344-5303 Joseph Dvorak, Brookhaven National Laboratory, jdvorak@bnl.gov, 344-5135
Beamtime Scheduler
Dario Arena, Brookhaven National Laboratory, darena@bnl.gov, 344-5303
Beamline Phone
631-344-7290
Instrumentation
Beamline Characteristics
Energy Range |
Mono Crystal or Grating |
Resolution (ΔE/E) |
Flux |
Spot Size (mm) |
Total Angular Acceptance (mrad) |
68 – 300 |
300 lines/mm |
10-4 - 10-3 |
2 x 1011 ph/sec/0.1%bw (@ 500 mA) |
|
15.0H x 1.5V |
136 – 600 |
600 lines/mm |
10-4 - 10-3 |
1 x 1011 ph/sec/0.1%bw (@ 500 mA) |
|
15.0H x 1.5V |
272 – 1200 |
1200 lines/mm |
10-4 - 10-3 |
1 x 1011 ph/sec/0.1%bw (@ 500 mA) |
|
15.0H x 1.5V |
13 – 50 |
600 lines/mm |
10-4 - 10-3 |
1 x 1011 ph/sec/0.1%bw (@ 500 mA) |
|
15.0H x 1.5V |
25 – 120 |
1200 lines/mm |
10-4 - 10-3 |
1 x 1011 ph/sec/0.1%bw (@ 500 mA) |
|
15.0H x 1.5V |
Source Type Bending magnet Optical System Horizontal Focusing Mirror (HFM): Spherical platinum-coated Zerodur mirror; focuses source horizontally at the sample position (14.8 meters from source); angle of incidence is 87 degrees; located 3.2 meters from the source.
Vertically Focusing Mirror (VFM): Spherical platinum-coated Zerodur mirror; focuses the source vertically onto the entrance slit of the monochromator (located 2.5 meters downstream of VFM); angle of incidence is 87.5 degrees; located 3.84 meters from the source. This mirror can be raised above or lowered below the orbital plane to select right- or left-handed elliptical polarization, for Magnetic Circular Dichroism (MCD) experiments.
Gratings: Three soft x-ray spherical gratings made of single-crystal silicon coated with gold; ruling densities 300, 600, and 1200 lines/mm, laminar type, holographically ruled and ion etched; diffract and focus vertically onto the moveable exit slit of the monochromator; the wavelength scanning mechanism is a simple sine drive type; zero order angle of incidence is 87 degrees (included angle is 174 degrees). Two VUV gratings made of ULE glass, coated with gold; ruling densities 600 and 1200 lines/mm, laminar type, holographically ruled and ion etched; diffract and focus vertically onto the movable exit slit of the monochromator via two gold-coated plane mirrors, at 12 degrees and 5 degrees grazing angles of incidence; zero order angle of incidence is 80 degrees (included angle is 160 degrees). Grating chamber is located 10.6 meters from the source.
The spot size at the first experimental chamber, located ~2 m downstream of the exit slit, is 1.5(v) x 10(h) mm2 (Can be collimated down). A second experimental chamber, for soft x-ray reflectivity, is located ~2 m downstream of the first experimental chamber, with a spot size of ~2.5 x 12 mm2.
Refocusing Optics: Bent elliptical "toroid" gold-coated electroless-nickel-on-aluminum mirror; focuses the first sample position (14.8 meters from source) to the thirdsample position (20.8 meters from source) with spot size of approx. 0.25(V) x 1.0(H) mm2. Experimental Apparatus First experimental chamber: general purpose UHV chamber with sample preparation/loading and liquid Helium capable cryostat for high resolution absorption or total electron yield measurements on solid samples. Switchable electromagnet for Magnetic Circular Dichroism (MCD) and hysteresis measurements. Maximum magnetic field (pulsed) is approximately 1000 Oe.
Second experimental chamber: soft x-ray reflectometer; sample preparation/loading and liquid nitrogen cryostat. Maximum magnetic field is approximately 300 Oe.
Third experimental chamber: available for roll-on/roll-off experimental chambers. Computer System Hardware & Software Windows 2000 PC; custom written software for data acquisition.
Links
|