Skip NavigationFDA Logo links to FDA home pageCenter for Devices and Radiological Health, U.S. Food and Drug AdministrationHHS Logo links to Department of Health and Human Services website
FDA Home Page | CDRH Home Page | Search | A-Z Index U.S. Food and Drug AdministrationCenter for Devices and Radiological Health Questions?
horizonal rule
Super Search
510(k) | Registration & Listing | Adverse Events | PMA | Classification | CLIA
CFR Title 21 | Advisory Committees | Assembler | Recalls | Guidance | Standards
 

New Search Back To Search Results
 

Medical Device Recalls
Class 2 Recall
SIZEWise Rentals Low-Boy Bed

See Related Information
Date Recall
Initiated
November 14, 2005
Date Posted May 25, 2006
Recall Number Z-0908-06
Product Low-Boy Bed, AC-powered adjustable hospital bed with a low position of 7'' to a high position of 30''. The beds are distributed by SIZEWise Rentals, Ellis, KS.
Code Information The serial numbers will be listed in the following format: SIZEWise number/Wheelchairs of Kansas number 22639/212666; 23243/212849; 22493/212646; 23245/212851; 21244/212428; 21254/212438; 22132/212502; 22140/212510; 22141/212511; 22142/212512; 22139/212509; 24403/213406; 22640/212667; 21251/212435; 23227/212860; 23228/212868; 23229/212869; 23257/212878/ 24384/213086; 24385/213087; 24398/213282; 24409/213412; 21252/212436; 22128/212498; 22636/212644; 23256/212879; 22502/212655; 22630/212658; 22642/212669; 21245/212429; 22129/212499; 23262/213073; 24401/213285; 22126/212496; 22496/212649; 23232/212863; 23239/212858; 23240/212846; 24402/213286; 23226/212871; 22135/212505; 22137/212507; 23246/212852; 23247/212853; 23248/212854; 23241/212847; 23260/213071; 23274/213084; 24387/213089; 23271/213081; 24396/213080; 22635/212663; 24411/213414; 24413/213416; 24414/213417; 24415/213418; 24416/213419; 24417/213420; 24418/213421; 24419/213425; 24420/213426; 24421/213427; 24422/213428; 25360/213429; 25361/213430; 25362/213431; 25363/213432; 25364/213433; 25365/213434; 25366/213435; 25367/213436; 25368/213437; 22136/212506; 22492/212631; 22638/212665; 22643/212670; 21240/212423; 21241/212424; 21246/212430; 22631/212659; 22632/212660; 22644/212874; 22646/212872; 23261/213072; 23263/213074; 23264/213075; 23265/212875; 23267/213077; 24391/213275; 24397/213281; 24399/213283; 21242/212425; 22498/212651; 23266/213076; 24400/213284; 24407/213410; 24408/213411; 21247/212431; 22495/212648; 23231/212866; 23234/212865; 24390/213123; 24389/213122; 24388/213121; 21253/212437; 23242/212848; 22134/212504; 22138/212508; 23249/212855; 23250/212856; 23251/212857; 23258/212877; 23259/212876; 22499/212652; 22500/212653; 22501/212654; 22628/212656; 22629/212657; 23233/212864; 23236/212861; 23237/212860; 24394/213278; 24395/213279; 21240/212423; 22497/212650; 24393/213277; 24405/213408; 24406/213409; 22127/212497; 22494/212647; 23230/212867; 23238/212859; 23255/212880; 24392/213276; 21249/212433; 21255/212439; 22131/212501; 22637/212671; 23252/212969; 23254/213070; 23270/213080; 24383/213085; 24386/213088; 23235/212862; 23244/212850; 22133/212503; 23253/212970; 23269/213079; 22130/212500; 22634/212662; 22645/212873; 24404/213407; 24410/213413; 22633/212661; 23272/213082; 23273/213083.
Recalling Firm/
Manufacturer
Rayes Inc
204 W 2nd Street
PO BOX 320
Ellis , Kansas 67637-0320
For Addition Information Contact Eric D. Boss
800-537-6454 Ext. 2100
Reason For
Recall
The bed may drop due to a malfunction of the actuator exceeding its stop.
Action The recalling firm met with their one customer on/about 11/14/05 explaining the reason for correction. The customer then notified their company stores and dealers via service bulletin dated 2/28/06 that they would be receiving new acutators for replacement on the beds. The actuators were sent to their company stores and dealers in early March.
Quantity in Commerce 161 beds
Distribution The beds were shipped to the nationwide customers of their one consignee who is located in KS.
 