FDA Logo links to FDA home pageCenter for Devices and Radiological Health, U.S. Food and Drug AdministrationHHS Logo links to Department of Health and Human Services website

FDA Home Page | CDRH Home Page | Search | CDRH A-Z Index | Contact CDRH U.S. Food and Drug Administration Center for Devices and Radiological Health [Skip navigation]
horizonal rule

Super Search
510(k) | Registration & Listing | Adverse Events | PMA | Classification | CLIA
CFR Title 21 | Advisory Committees | Assembler | Recalls | Guidance | Standards
 

 

Adverse Event Report

DATASCOPE CORP. DATASCOPE INTRA AORTIC BALLOON PUMP CATHETER   back to search results
Event Description

Three different pts had an 8 fr sheathless iabp catheter (datascope) inserted with immediate clotting of the arterial lumen of the iabp, thus necessitating the insertion of another arterial line on pts who had been anticoagulated on reopro or tpa. In investigating this situation facility found that their pressure tubing was not compatible with a 2 french arterial lumen. Datascope kits include five feet of special pressurized tubing for this catheter, but you must break apart your sterile transducer set to accomplish this. Even with their tubing you must frequently flush for a dampened waveform. In talking with other hosps in the tri-state area, they report the same problems with the 8 french sheathless iabp catheter. One facility's pts with this problem did have neurological problems and ct of head revealed multiple small embolic infarcts. This was not officially attributed to the iabp.

 
Event Description

Add'l info rec'd from mfr 8/14/00: the identity of the balloon cannot be established due to the lack of reported info supplied to the fda by the rptr. It is not possible to determine if this complaint had been previously reported to datascope or if an iab was returned for evaluation.

 
Search Alerts/Recalls

  new search  |  submit an adverse event report

Brand NameDATASCOPE
Type of DeviceINTRA AORTIC BALLOON PUMP CATHETER
Baseline Device 510(K) Number
Baseline Device PMA Number
Manufacturer (Section D)
DATASCOPE CORP.
15 law dr.
fairfield NJ 07004 3206
Device Event Key277197
MDR Report Key286458
Event Key268749
Report NumberMW1019324
Device Sequence Number1
Product CodeDSP
Report Source Voluntary
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received07/18/2000
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator Health Professional
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer06/01/2000
Is the Device an Implant? No
Is this an Explanted Device? No Answer Provided

Database last updated on January 30, 2009