Laboratory Results That Should Be Ignored
Dirk M. Elston, MD, Director
Dirk M. Elston, Department of Dermatology; Dermatopathology Program Director, Department of Pathology, Geisinger Medical Center, Danville, Pennsylvania E-mail: dmelston/at/geisinger.edu.
Disclosure: Dirk M. Elston, MD, has disclosed no relevant financial relationships.
Introduction

There are instances in which the results of a laboratory test must be reported, but should be ignored by the clinician. Some are accurate findings without clinical significance, but most are spurious results caused by an error somewhere in the testing cycle. Any abnormal laboratory result should raise 2 basic questions: First, what is the clinical relevance of the result? Second, what actions should be taken? This article examines the range of reasons why some laboratory results should be ignored. It examines the causes of spurious results and examines mechanisms to improve communication to convey such reports to the clinician in a more meaningful fashion. The ultimate goals are improved patient outcomes and better utilization of resources.

Specific issues to be addressed in this commentary include the need to expand quality programs to address the entire testing cycle, root causes of spurious lab results in clinical and anatomic pathology laboratories, the critical involvement of clinicians in the quality improvement process, and the role of interpretative reporting to reduce medical errors related to spurious results. Many spurious results will not be identified by quality programs that address only the analytic phase. Joint efforts by clinicians and laboratorians can improve detection of errors and reduce the potential for patient harm and waste.

Spurious Results Related to Breakdowns in the Testing Cycle

The testing cycle extends from the decision to perform the test through the final interpretation and response to the laboratory result (Figure 1). For decades, quality programs in laboratories have focused almost exclusively on the analytic phase of the testing cycle. Quality efforts that focus on the precision and accuracy of laboratory assays are important, but errors are more likely to arise in the preanalytic and postanalytic phases of testing.[14] Significant gains in quality can be achieved by focusing on these pre- and postanalytic phases.[59]

Figure 1Figure 1
The testing cycle.

Common sources of error in the preanalytic phase include unintelligible requests, inadequate samples, errors in patient identification and data entry, and errors during collection, transport, and accessioning of specimens. In the postanalytic phase, common errors fall into 3 major categories: (1) turnaround time (result reported too late), (2) routing (report not received by the correct clinician), and misinterpretation (correct result reported, but wrong clinical action taken).[10] Because these types of error involve both the laboratory and clinician, both must be involved in efforts to reduce errors.

Figure 2 demonstrates the portion of the testing cycle that has been traditionally addressed by laboratory quality programs. Figure 3 gives a sense of the opportunity for gains in quality if the entire testing cycle is addressed. Errors in the pre- and postanalytic phases may be attributable to the clinician, the laboratory, or the system. An effective quality program relies on active participation from all participants in the testing process. Data entry errors remain a major cause of spurious laboratory results. Every requirement for manual data entry or transfer creates an opportunity for error. A specimen may be labeled at the bedside, with the same critical information entered as a separate process on the request form and again in the laboratory data system. Failure mode analysis of the processes involved often identifies separate requirements for manual entry of 15 or more digits for the test, 5 digits for the operator, and an additional 12 for patient billing information. Bar coding at the point of care can reduce medical errors related to data entry.[11,12] Even when bar coding is instituted, scanning errors may be related to multiple wristbands and incorrect codes related to hospital transfer or previous hospitalizations. Verification of the scanned information is still required. Manual data entry may still be required when bar-coded scans are unsuccessful. Information technology systems must ensure the integration of communication throughout the testing cycle. Common standards are required for data reporting systems to ensure that they communicate with each other with seamless electronic transfer of data and the ability to capture asynchronous data. Clinical decision support can be integrated into these systems to enhance effective communication.[13] Errors in communication can also be reduced through the use of bar-coded phrase sheets and keyboard macros.[14]

Figure 2Figure 2
Quality efforts have traditionally focused only on the analytic phase.
Figure 3Figure 3
The opportunity for improvement.
False-Positive Clinical Pathology Results

Although most laboratory errors relate to the pre- and postanalytic phases, the analytic phase should not be ignored as a source of errors. Modern laboratories use highly specific, sensitive methods that generally result in reliable results. Because of this, clinicians place great faith in the results that they receive, and false-positive results frequently result in a cascade of unnecessary testing and treatment. One widely publicized example involved a 22-year-old woman who underwent both surgery and chemotherapy on the basis of multiple false-positive human chorionic gonadotropin results.[15] The methodology employed was a sandwich immunoassay that utilized 2 specific monoclonal antibodies. In this assay, the target protein forms a link between the capture and signal antibodies. The assay is known for its high sensitivity and specificity. Unfortunately, heterophilic antibodies can also form links between the capture and signal antibodies, resulting in a false-positive result. Despite test modifications to minimize antibody interference, heterophilic antibody interference still occurs in a wide variety of immunoassays, and false-positive results have led to bad clinical outcomes.[1620] Because this type of interference is patient-specific, it usually escapes detection by quality assurance systems that focus mainly on the analytic phase of testing. A broader system that addresses the entire testing cycle can improve quality.

False-positive results that result in medical errors occur in a wide variety of laboratory assays. A study of 6370 specimens analyzed by a widely used bacterial latex agglutination test found that only 11 pathogens were accurately detected, whereas there were 13 false negatives and 59 false-positive results. In this study, none of the true positives had a measurable effect on patient outcomes, but several of the erroneous reports resulted in unnecessary treatment.[21] Nonculture tests for detection of Chlamydia trachomatis have advanced the detection of this serious sexually transmitted disease, but have also been complicated by false-positive results. Implicated assays include both enzyme immunoassays and DNA probe tests.[22] The issue of false-positive nucleic acid detection tests for Neisseria gonorrhoeae has prompted recommendations for the reporting of such tests.[23] A major source for such errors is horizontal genetic exchange among Neisseria, resulting in commensal Neisseria acquiring N gonorrhoeae genes.[24] As with other studies, false-positive results are more likely in populations with low disease prevalence.[25] Published recommendations with regard to reporting of nucleic acid detection tests for N gonorrhoeae include the following: Assays for the cppB gene should not be used; all positive screening assays should be confirmed by a reliable supplemental assay before a positive result is reported; and the test combination should yield a positive predictive value of at least 90% in a population with a 1% prevalence.[26] Polymerase chain reaction (PCR) assays offer sensitivity and specificity similar to culture and can be used as confirmatory tests, but may not be suitable for pharyngeal specimens.[2729]

A falsely reported outbreak of methicillin-resistant Staphylococcus aureus was linked to spurious results with an automated test system. Independent laboratory testing with oxacillin agar screen plates, broth microdilution minimal inhibitory concentrations, and assays for the presence of the mec A gene proved that almost three quarters of the organisms initially identified as methicillin-resistant S aureus were actually susceptible to oxacillin. The misclassification resulted in unnecessary vancomycin use.[30] An apparent outbreak of typhoid fever was related to false-positive tests with a hemagglutination method. Most laboratories issuing spurious reports performed the rapid slide agglutination test as opposed to the tube agglutination test.[31] Each of the above examples shares one important feature: Quality improvement programs that focused on the analytic phase of testing failed to detect unacceptable rates of false-positive results. Analysis of patient outcomes ultimately uncovered the errors.

Spurious Anatomic Pathology Results

Efforts to classify anatomic pathology errors and assess their influence on clinical outcomes have been hampered by marked differences in reporting practices and lack of a uniform classification scheme for errors in anatomic pathology.[32] As with clinical pathology, quality improvement efforts have traditionally centered on the analytic phase, with an emphasis on peer review of a percentage of cases, correlation of frozen-section and permanent-section diagnoses, case conferences, and consultations. Quality measures of the pre- and postanalytic phases have generally been limited to patient and site verification and turnaround time. Additional gains in quality may be achieved by focusing on the entire testing cycle and patient outcomes.

Interpretative anatomic pathology reports can be of value. An example is the cytologic atypia that is characteristic of so-called “ancient schwannomas.” Although the nuclei are strikingly atypical, the tumors behave in a completely benign fashion.[33] The lack of clinical significance must be clearly communicated in order to avoid unnecessarily extensive surgery. Perineuriomas represent another group of benign tumors with a wide spectrum of histologic appearance. Atypia and hypercellularity showed little correlation with clinical behavior.[34] Similar “ancient” atypia occurs in a host of other neoplasms.[3539] Unambiguous communication of the lack of significance of the finding is essential to avoid overly aggressive treatment.

Effective Use of Interpretative Reports

Interpretative reports are crucial when a laboratory result should be ignored, but even accurate results may be misinterpreted, and interpretative reports have dramatic potential to reduce medical errors and unnecessary follow-up testing.[4042] Accurate interpretation of test results depends on patient history, experience of the interpreter, and local disease prevalence. When the pretest likelihood of disease is small, there will be a larger number of false positives.[43] In the preanalytic phase, redirecting test ordering has the potential to reduce false-positive results stemming from testing patients in low-prevalence populations. In the postanalytic phase, computer graphs have been used to show post-test probability of disease as a function of pretest probability.[44]

Interpretative reports vary from graphic representation of normal ranges to elaborate mathematic models. Computer systems have facilitated interpretive reporting, especially for high-volume and high-complexity laboratory tests.[4548] Automated laboratory analyzers can present laboratory results in graphic form relative to population-based data. Graphic reports with manually added interpretative text can be of additional value in conveying results effectively.[49] Critical differences (as well as those differences that are clinically insignificant) must be clearly communicated in graphic reports. The reporting interval or bin size that is used to report numerical results influences interpretations of the critical difference between results. As the reporting interval size increases, there is a roughly linear increase in the appearance of a critical difference.[50] Standard reporting intervals must be integrated with data in regard to what difference is clinically relevant.[51]

When reporting culture results, the growth of organisms that usually represent normal flora is complicated by the fact that the same organisms can sometimes act as opportunistic pathogens.[52,53] The line between normal flora, colonist, and pathogen may be poorly defined.[54,55] It has been traditionally left to the clinician to determine the clinical significance of the organism, but as our understanding of the role of virulence factors in opportunistic infection and immune-mediated diseases evolves, laboratories will have a more defined role in communicating the possible clinical significance of various organisms. Interpretive reports are also being used to communicate bacterial susceptibility test results effectively.[56]

The wording of interpretative reports is critical, and it may be helpful to have standard comments reviewed by panels of clinicians. In one study of the quality of interpretative comments in clinical chemistry, the study authors found that a significant proportion of well-intended comments were considered to be misleading by a panel of clinicians.[57] A study of computer-based interpretative reporting of endocrine studies found that interpretative reports were valued more highly by clinicians when the interpretative comments were specific to the test result, stated as considerations rather than recommendations, and accompanied by an expanded list of differential diagnoses and a listing of drugs that are known to affect the test result.[58]

Bridges Between the Laboratory and Clinic

There are noteworthy models of whichspecialty societies representing both laboratory and clinical medicine have collaborated to create standards for reporting that define laboratory results that should be ignored, those that should be followed, and those that require action. One example is the reporting of atypical squamous or glandular cells in gynecologic cytology specimens.[5961] Another example is the interpretation of weakly or moderately positive antinuclear antibody (ANA) tests.[62] As treatment options and the technology that is used to prepare specimens for interpretation continue to evolve, such published standards must be dynamic in order to remain clinically useful.[63,64]

Novel diagnostic tests represent a particular challenge. Longitudinal algorithms that compare the new test with previous methods are required to validate new markers. Longitudinal algorithms have been successfully employed for prostate-specific antigen and CA125 assays for the detection of prostate and ovarian cancer, and the parametric empirical Bayes screening algorithm has been proposed as a method for evaluating new tumor markers.[65]

Conclusions

Laboratory results that should have been ignored commonly trigger a cascade of unnecessary testing. At the worst, they result in unnecessary surgery or the inappropriate use of a medication. Improvements in quality can be gained by addressing the entire testing cycle, both in clinical and anatomic pathology. Unambiguous communication with the clinician and common standards of language are essential in order to avoid bad outcomes. During training, clinicians are often admonished not to ignore laboratory results, and clinicians are naturally cautious when dealing with conditions that are potentially serious. A commonly cited motive for ordering additional laboratory tests is the perceived need to reassure the patient through further testing.[66] Interpretative reports and clinical decision support pathways can reduce unnecessary testing without adversely affecting patient outcomes.[43, 6774]

The College of American Pathologists continues to address questions of sensitivity, specificity, and predictive value of laboratory tests.[73] Clinical specialties have begun to address the undesirable consequences of incidental test results.[74] The time is ripe for joint efforts to improve quality. The establishment of standards will require significant investment by all stakeholders, but represents an opportunity to improve patient outcomes and reduce healthcare costs.

Footnotes
Sign Up now for a free monthly email that brings you the top features from MedGenMed.
Readers are encouraged to respond to George Lundberg, MD, Editor of MedGenMed, for the editor's eyes only or for possible publication via email: glundberg/at/medscape.net
References
1.
Rivers, PA; Dobalian, A; Germinario, FA. A review and analysis of the clinical laboratory improvement amendment of 1988: compliance plans and enforcement policy. Health Care Manage Rev. 2005;30:93–102. [PubMed]
2.
Johnson, PR. The contribution of proficiency testing to improving laboratory performance and ensuring quality patient care. Clin Leadersh Manag Rev. 2004;18:335–341. [PubMed]
3.
Howanitz, PJ. Errors in laboratory medicine: practical lessons to improve patient safety. Arch Pathol Lab Med. 2005;129:1252–1261. [PubMed]
4.
Stankovic, AK. The laboratory is a key partner in assuring patient safety. Clin Lab Med. 2004;24:1023–1035. [PubMed]
5.
Saxena, S; Kempf, R; Wilcox, S, et al. Critical laboratory value notification: a failure mode effects and criticality analysis. Jt Comm J Qual Patient Saf. 2005;31:495–506. [PubMed]
6.
Ricos, C; Garcia-Victoria, M; de la Fuente, B. Quality indicators and specifications for the extra-analytical phases in clinical laboratory management. Clin Chem Lab Med. 2004;42:578–582. [PubMed]
7.
Barenfanger, J; Sautter, RL; Lang, DL; Collins, SM; Hacek, DM; Peterson, LR. Improving patient safety by repeating (read-back) telephone reports of critical information. Am J Clin Pathol. 2004;121:801–803. [PubMed]
8.
Berte, LM. Patient safety: getting there from here — quality management is the best patient safety program. Clin Leadersh Manag Rev. 2004;18:311–315. [PubMed]
9.
Bachner, P. Quality assurance of the analytic process: pre- and postanalytic variation. Clin Lab Med. 1986;6:613–623. [PubMed]
10.
Saarela, U; Kivinen, T; Sarkkinen, H. External failures as descriptor of quality in clinical microbiology laboratory services. Clin Chem Lab Med. 2005;43:54–63. [PubMed]
11.
Nichols, JH; Bartholomew, C; Brunton, M, et al. Reducing medical errors through barcoding at the point of care. Clin Leadersh Manag Rev. 2004;18:328–334. [PubMed]
12.
Shaw, R; Coia, JE; Michie, J. Use of bar code readers and programmable keypads to improve the speed and accuracy of manual data entry in the clinical microbiology laboratory: experience of two laboratories. J Clin Pathol. 1999;52:54–60. [PubMed]
13.
Bates, DW; Cohen, M; Leape, LL; Overhage, JM; Shabot, MM; Sheridan, T. Reducing the frequency of errors in medicine using information technology. J Am Med Inform Assoc. 2001;8:299–308. [PubMed]
14.
Aller, RD; Friedman, W. Rapid accurate entry of microbiology results. Arch Pathol Lab Med. 1996;120:57–61. [PubMed]
15.
White, GH. Trusting numbers: uncertainty and the pathology laboratory. MJA. 2002;177:153–155. Available at: www.mja.com.au/public/issues/177_03_050802/whi10784_fm.html Accessed September 29, 2006. [PubMed]
16.
Cole, LA. Use of hCG tests for evaluating trophoblastic diseases: choosing an appropriate hCG assay, false detection of hCG, unexplained elevated hCG, and quiescent trophoblastic disease. USA hCG Reference Service Report. 2002. Available at: http://www.hcglab.com/hCG%20Ref%20Service%20Reprort%202001.htm Accessed September 25, 2006.
17.
Selby, C. Interference in immunoassay. Ann Clin Biochem. 1999;36:704–721. [PubMed]
18.
Ismail, AAA; Barth, JH. Wrong biochemistry results. BMJ. 2001;323:705–706. [PubMed]
19.
Fitzmaurice, TF; Brown, C; Rifai, N, et al. False increase of cardiac troponin I with heterophilic antibodies. Clin Chem. 1998;44:2212–2214. [PubMed]
20.
Krahn, J; Parry, DM; Laroux, M; Dalton, J. High percentage of false positive cardiac troponin I results in patients with rheumatoid factor. Clin Biochem. 1999;32:477–480. [PubMed]
21.
Hayden, RT; Frenkel, LD. More laboratory testing: greater cost but not necessarily better. Pediatr Infect Dis J. 2000;19:290–292. [PubMed]
22.
Hammerschlag, MR; Ajl, S; Laraque, D. Inappropriate use of nonculture tests for the detection of Chlamydia trachomatis in suspected victims of child sexual abuse: a continuing problem. Pediatrics. 1999;104(pt1):1137–1139. [PubMed]
23.
Lum, G; Freeman, K; Nguyen, NL, et al. A cluster of culture positive gonococcal infections but with false negative cppB gene based PCR. Sex Transm Infect. 2005;81:400–402. [PubMed]
24.
Whiley, DM; Tapsall, JW; Sloots, TP. Nucleic acid amplification testing for Neisseria gonorrhoeae: an ongoing challenge. J Mol Diagn. 2006;8:3–15. [PubMed]
25.
Katz, AR; Effler, PV; Ohye, RG; Brouillet, B; Lee, MV; Whiticar, PM. False-positive gonorrhea test results with a nucleic acid amplification test: the impact of low prevalence on positive predictive value. Clin Infect Dis. 2004;38:814–819. [PubMed]
26.
Smith, DW; Tapsall, JW; Lum, G. Guidelines for the use and interpretation of nucleic acid detection tests for Neisseria gonorrhoeae in Australia: a position paper on behalf of the Public Health Laboratory Network. Commun Dis Intell. 2005;29:358–365. [PubMed]
27.
Geraats-Peters, CW; Brouwers, M; Schneeberger, PM, et al. Specific and sensitive detection of Neisseria gonorrhoeae in clinical specimens by real-time PCR. J Clin Microbiol. 2005;43:5653–5659. [PubMed]
28.
Airell, A; Lindback, E; Ataker, F; Pornull, KJ; Wretlind, B. Verification of clinical samples, positive in AMPLICOR Neisseria gonorrhoeae polymerase chain reaction, by 16S rRNA and gyrA compared with culture. Int J STD AIDS. 2005;16:415–419. [PubMed]
29.
Whiley, DM; Buda, PJ; Bayliss, J; Cover, L; Bates, J; Sloots, TP. A new confirmatory Neisseria gonorrhoeae real-time PCR assay targeting the porA pseudogene. Eur J Clin Microbiol Infect Dis. 2004;23:705–710. [PubMed]
30.
Ender, PT; Durning, SJ; Woelk, WK, et al. Pseudo-outbreak of methicillin-resistant Staphylococcus aureus. Mayo Clin Proc. 1999;74:885–889. [PubMed]
31.
Nsutebu, EF; Ndumbe, PM; Koulla, S. The increase in occurrence of typhoid fever in Cameroon: overdiagnosis due to misuse of the Widal test? Trans R Soc Trop Med Hyg. 2002;96:64–67. [PubMed]
32.
Zarbo, RJ; Meier, FA; Raab, SS. Error detection in anatomic pathology. Arch Pathol Lab Med. 2005;129:1237–1245. [PubMed]
33.
Argenyi, ZB; Balogh, K; Abraham, AA. Degenerative ("ancient") changes in benign cutaneous schwannoma. A light microscopic, histochemical and immunohistochemical study. J Cutan Pathol. 1993;20:148–153.
34.
Hornick, JL; Fletcher, CD. Soft tissue perineurioma: clinicopathologic analysis of 81 cases including those with atypical histologic features. Am J Surg Pathol. 2005;29:845–858. [PubMed]
35.
Kerl, H; Soyer, HP; Cerroni, L; Wolf, IH; Ackerman, AB. Ancient melanocytic nevus. Semin Diagn Pathol. 1998;15:210–215. [PubMed]
36.
Kamino, H; Lee, JY; Berke, A. Pleomorphic fibroma of the skin: a benign neoplasm with cytologic atypia. A clinicopathologic study of eight cases. Am J Surg Pathol. 1989;13:107–113.
37.
Tamada, S; Ackerman, AB. Dermatofibroma with monster cells. Am J Dermatopathol. 1987;9:380–387. [PubMed]
38.
Layfield, LJ; Bentley, RC; Mirra, JM. Pseudoanaplastic giant cell tumor of bone. Arch Pathol Lab Med. 1999;123:163–166. [PubMed]
39.
Goodman, WT; Bang, RH; Padilla, RS. Giant dermatofibroma with monster cells. Am J Dermatopathol. 2002;24:36–38. [PubMed]
40.
Laposata, ME; Laposata, M; Van Cott, EM; Buchner, DS; Kashalo, MS; Dighe, AS. Physician survey of a laboratory medicine interpretive service and evaluation of the influence of interpretations on laboratory test ordering. Arch Pathol Lab Med. 2004;128:1424–1427. [PubMed]
41.
Laposata, M. Patient-specific narrative interpretations of complex clinical laboratory evaluations: who is competent to provide them? Clin Chem. 2004;50:471–472. [PubMed]
42.
Covell, DG; Uman, GC; Manning, PR. Information needs in office practice: are they being met? Am J Med. 2000;109:218–223. [PubMed]
43.
Bernstein, RM; Hollingworth, GR; Wood, WE. Prompting physicians for cost-effective test ordering in the low prevalence conditions of family medicine. Proc Annu Symp Comput Appl Med Care. 1994:824–828. [PubMed]
44.
Tandberg, D; Deely, JJ; O'Malley, AJ. Generalized likelihood ratios for quantitative diagnostic test scores. Am J Emerg Med. 1997;15:694–699. [PubMed]
45.
McNeely, MD. Computerized interpretation of laboratory tests: an overview of systems, basic principles and logic techniques. Clin Biochem. 1983;16:141–146. [PubMed]
46.
Swezey, CB; Jacobson, W. Computer-based diagnostic reporting for serum electrolytes. Am J Clin Pathol. 1980;74:812–819. [PubMed]
47.
Trendelenburg, C; Colhoun, O; Wormek, A; Massey, KL. Knowledge-based test result in interpretation in laboratory medicine. Clin Chim Acta. 1998;278:229–242. [PubMed]
48.
Pribor, HC. Expert systems in laboratory medicine: a practical consultative application. J Med Syst. 1989;13:103–109. [PubMed]
49.
Regeniter, A; Steiger, JU; Scholer, A; Huber, PR; Siede, WH. Windows to the ward: graphically oriented report forms. Presentation of complex, interrelated laboratory data for electrophoresis/immunofixation, cerebrospinal fluid, and urinary protein profiles. Clin Chem. 2003;49:41–50.
50.
Jones, GR. Effect of the reporting-interval size on critical difference estimation: beyond “2.77.” Clin Chem. 2006;52:880–885. [PubMed]
51.
Holvoet, P; Macy, E; Landeloos, M; Jones, D; Nancy, JS; Van de Werf, F; Tracy, RP. Analytical performance and diagnostic accuracy of immunometric assays for the measurement of circulating oxidized LDL. Clin Chem. 2006;52:760–764. [PubMed]
52.
Waghorn, DJ. Actinobaculum massiliae: a new cause of superficial skin infection. J Infect. 2004;48:276–277. [PubMed]
53.
Berg, G; Eberl, L; Hartmann, A. The rhizosphere as a reservoir for opportunistic human pathogenic bacteria. Environ Microbiol. 2005;7:1673–1685. [PubMed]
54.
Kuipers, EJ; Michetti, P. Bacteria and mucosal inflammation of the gut: lessons from Helicobacter pylori. Helicobacter. 2005;10(suppl1):66–70. [PubMed]
55.
Peipert, JF. Actinomyces: normal flora or pathogen? Obstet Gynecol. 2004;104(pt2):1132–1133. [PubMed]
56.
Waites, KB. Effective communication of antimicrobial susceptibility data by pathologists to clinicians. Adv Exp Med Biol. 2005;563:165–177. [PubMed]
57.
Lim, EM; Sikaris, KA; Gill, J, et al. Quality assessment of interpretative commenting in clinical chemistry. Clin Chem. 2004;50:632–637. [PubMed]
58.
Dons, RF. Interpretive reporting of laboratory data: proposed criteria to qualify as a high-quality, limited clinical pathology consultation. Endocr Pract. 1996;2:255–260. [PubMed]
59.
American College of Obstetricians and Gynecologists Committee on Quality Assessment. ACOG criteria set. Atypical squamous cells of undetermined significance (ASCUS). Int J Gynaecol Obstet. 1996;52:215–216.
60.
Gurbuz, A; Karateke, A; Kabaca, C; Kir, G. Atypical glandular cells: improvement in cytohistologic correlation by the 2001 Bethesda system. Int J Gynecol Cancer. 2005;15:903–910. [PubMed]
61.
Gonzalez, D; Hernandez, E; Anderson, L; Heller, P; Atkinson, BF. Clinical significance of a cervical cytologic diagnosis of atypical squamous cells of undetermined significance. Favoring a reactive process or low grade squamous intraepithelial lesion. J Reprod Med. 1996;41:719–723. [PubMed]
62.
Reichlin, M. Washington, DC: Lupus Foundation of America; 2005. Laboratory tests used in the diagnosis of lupus. Available at: http://www.lupus.org/education/brochures/labtests.html Accessed September 25, 2006.
63.
Schledermann, D; Ejersbo, D; Hoelund, B. Significance of atypia in conventional Papanicolaou smears and liquid-based cytology: a follow-up study. Cytopathology. 2004;15:148–153. [PubMed]
64.
Levi, AW; Kelly, DP; Rosenthal, DL; Ronnett, BM. Atypical squamous cells of undetermined significance in liquid-based cytologic specimens: results of reflex human papillomavirus testing and histologic follow-up in routine practice with comparison of interpretive and probabilistic reporting methods. Cancer. 2003;99:191–197. [PubMed]
65.
McIntosh, MW; Urban, N; Karlan, B. Generating longitudinal screening algorithms using novel biomarkers for disease. Cancer Epidemiol Biomarkers Prev. 2002;11:159–166. [PubMed]
66.
van der Weijden, T; van Bokhoven, MA; Dinant, GJ; van Hasselt, CM; Grol, RP. Understanding laboratory testing in diagnostic uncertainty: a qualitative study in general practice. Br J Gen Pract. 2002;52:974–980. [PubMed]
67.
Board, N; Brennan, N; Caplan, G. Use of pathology services in re-engineered clinical pathways. J Qual Clin Pract. 2000;20:24–29. [PubMed]
68.
Dons, RF. Interpretive reporting of laboratory data: proposed criteria to qualify as a high-quality, limited clinical pathology consultation. Endocr Pract. 1996;2:255–260. [PubMed]
69.
Yamauchi, K; Fukatsu, T. A decision support system for diagnostic consultation in laboratory tests. Medinfo. 1995;8(pt2):1034. [PubMed]
70.
Boon-Falleur, L; Sokal, E; Nightingale, PG. Utilization of laboratory resources: developments in knowledge-based ordering systems. Int J Biomed Comput. 1995;40:17–30. [PubMed]
71.
Plapp, FV; Essmyer, CE; Byrd, AB; Zucker, ML. How to successfully influence laboratory test utilization. Clin Leadersh Manag Rev. 2000;14:253–260. [PubMed]
72.
Sucov, A; Bazarian, JJ; deLahunta, EA; Spillane, L. Test ordering guidelines can alter ordering patterns in an academic emergency department. J Emerg Med. 1999;17:391–397. [PubMed]
73.
Anita, J; Slomski, AJ. Northfield, Ill: College of American Pathologists; Innovations in pathology: improving laboratory use. Available at: http://www.cap.org/apps/docs/cap_today/feature_stories/innovations_lab_use_feature.html Accessed September 25, 2006.
74.
Sonnenberg, A. Personal view: 'don't ask, don't tell' — the undesirable consequences of incidental test results in gastroenterology. Aliment Pharmacol Ther. 2004;20:381–387. [PubMed]