5.6.3 STABLE ISOTOPES AND RADIOCHEMICALS

Isotopes and radiochemicals generally are not processed in a processing or preservation chamber, unless samples are being handled in a glove box. Wear appropriate, disposable, powderless gloves when collecting and processing samples.

Leave enough air space (at least 2 cubic centimeters) if glass bottles are used and the sample will be chilled to allow for the expansion of water samples unless instructed otherwise (Appendix A5-C). Close the polyseal cap tightly and seal with wax or plastic tape, or as directed below for the specific isotope or radiochemical.

5.6.3.A Carbon (13C/12C and 14C)

Do not let the sample contact air. Filter the sample along with other inorganic-constituent samples if particulates are visible. Samples without particulates do not require filtration or chemical treatment (NWQL Technical Memorandum 96.05). If 13 C will be collected by direct precipitation in the field using ammoniacal SrCl2, then the precipitates must be submitted as washed, dried, homogenized powders. The laboratory will not accept bottles with ammoniacal SrCl2 in solution.

Samples for 13C/12C analysis:

1. Label a 1-L glass bottle as "13C/12C , RUS," and the laboratory code or schedule number (as requested by the laboratory).
2. Filter the sample if particulates are visible. Establish a closed path from the sample source through the filter and to the bottle to exclude air from the collection system.
3. When filling the bottle, fill from the bottom of the bottle and allow an overflow of two to three bottle volumes. Cap the sample immediately.

Samples for 14C analysis:

The sample must provide a minimum of 5 mg dissolved inorganic carbon (DIC) per sample container.

1. Label bottles as " 14C, RUS/RUR," and the appropriate laboratory schedule number. Check with the laboratory and refer to NWQL Technical Memorandum 96.05 for bottling and other field requirements for samples to be analyzed for 14C.
2. Collect sample in duplicate. Filter the sample if particulates are visible. Establish a closed path from the sample source, through the filter, to the bottle to ensure that air is excluded from the collection system.
3. Fill sample container.
 

For samples with 14C content greater than or equal to 5-percent modern carbon, fill the bottle from the bottom of the bottle, allowing an overflow of 2 to 3 bottle volumes. This helps to purge air from the sample. Cap the bottle immediately. For potentially low 14C concentrations, (<10 percent modern carbon) or if relatively long filtration time is required, flush the headspace above the water sample with nitrogen while filling the bottle.

  For samples with 14C content less than 5-percent modern carbon, use a stainless steel collection vessel, such as a WhiteyTM Sample cylinder No. 304L-HDF4, with stainless steel valves on each end. Flush with several liters of filtrate from the bottom of the cylinder up and close the cylinder, leaving no headspace.
4. Archive a duplicate sample and store in the dark, chilled to 4°C or below without freezing.
 

Secure container caps with electrical tape.

  Record the sample pH and alkalinity on the sample bottle.

5.6.3.B. Hydrogen (2H/1H) and Oxygen (180/160)

Hydrogen and oxygen isotope samples can be collected together in one bottle. Use either a 60-mL clean glass bottle or a 250-mL polyethylene bottle. Label the bottles as "RUS, SH 1142." Use caps with polyseal conical inserts. To request bottles with appropriate caps, send E-mail to isotopes@usgs.gov.

 

Do not field rinse bottles.

  Do not add chemical treatment.
  Samples may be either raw or filtered.
1. Fill bottle to overflowing directly from the sampler or sample splitting device (raw sample) or from the capsule filter or other filter assembly (filtered sample).
2. If filling a glass bottle, fill to overflowing and then decant the sample until the water level is at the bottle shoulder. Cap the bottle immediately.
3. If filling a polyethylene bottle, fill to overflowing and cap the bottle immediately, leaving no headspace. Do not use polyethylene bottles if the sample will be held or archived.

5.6.3.C Nitrogen (15N/14N)

Collect filtered, untreated sample (do not use mercuric chloride) in a 1-L amber or foil-wrapped glass or high-density polyethylene bottle. Use a polyseal bottle cap. Refer to National Water Quality Technical Memorandum 95.05.

1. Label the bottles as "RUS" and with the appropriate laboratory code (LC 1717, for 15N/14N as ammonia; LC 1718, for 15N/14N as nitrate; or LC 1921, for 15N/14N as nitrate plus ammonia).
2. Fill the bottle to the shoulder directly from the capsule filter or other filter assembly.
3. Put the glass sample bottle into a foam sleeve and keep it chilled at or below 4°C without freezing. Ship overnight, per shipping instructions in NWQL Technical Memorandum 95.04.

5.6.3.D Sulfur (34S/32S)

Dissolved sulfide and sulfate are collected and processed separately, according to the directions given in Carmody and others (1998).

1. Label the bottles as "RUS" and with the appropriate laboratory code or schedule number.
2.

Send questions to the following E-mail address: <isotopes@usgs.gov>. Special equipment, chemical reagents, and training are needed in order to collect these samples properly.

Sulfate: Procedures and equipment differ for processing the sulfate sample, depending on whether sulfate concentrations are less than 20 mg/L or equal to or greater than 20 mg/L. Onsite estimation of sulfate concentration is described in Carmody and others (1998).

 

When a water sample from which dissolved sulfate will be collected for isotopic analysis also contains dissolved sulfide (greater than 0.01 mg/L), the sulfide should be removed by nitrogen stripping to avoid contamination of the sulfate sulfur by oxidized sulfide sulfur with very different isotopic composition.

  Sulfate in water with concentrations of greater than 20 mg/L sulfate is collected by precipitating BaSO4 from an acidified sample (up to 2-L volume). BaSO4 can be precipitated and filtered in the laboratory or in the field.
  Anion exchange resin can be used to collect sulfate from samples in which sulfate concentrations are less than 20 mg/L.

Sulfide: Determine if dissolved sulfide (H2S) is present by its distinctive rotten-egg odor. Measure H2S concentration with a field spectrophotometer.

 

Measurement of dissolved sulfide concentrations greater than 0.6 mg/L with a field spectrophotometer has been problematic.

  An alternative method for samples with dissolved sulfide concentration greater than about 0.5 mg/L is to collect the dissolved sulfide by direct precipitation of Ag2S by adding AgNO3 to the sample. Care must be taken to add sufficient AgNO3 to precipitate all the sulfide present, or the sulfide sample will be fractionated.
  A method for collection of dissolved sulfide by nitrogen stripping and precipitation of Ag2S in an AgNO3 trap has been tested and found to cause a small fractionation of the isotopic composition of the sulfide. Two approaches are recommended to minimize this fractionation: (1) collect sulfide by nitrogen stripping for 3 hours or more to minimize the amount of sulfide left in the carboy and, thus, the fractionation produced; and (2) determine the kinetic fractionation factor (a) for the collection apparatus and use this a to calculate the original sulfur isotopic composition of the dissolved sulfide in the ground water from Ag2S collected for a short time (about 30 minutes).

5.6.3.E Radium 226 and Radium 228

Label acid-rinsed polyethylene bottle(s) as "FAR" and add the appropriate laboratory code ("LC 794" for radon-226 and "LC 1364" for radon-228). Bottle-size requirements depend on analytical method or schedule.

1. Filter the sample using the procedures for inorganic-constituent samples, and fill the sample bottle to the shoulder directly from the capsule filter or other filter assembly.
2. Add reagent-grade HNO3 to preserve sample to pH <2. Do not substitute HCl for HNO3.

5.6.3.F Uranium (U-234, U-235, U-238)

Label 1-L acid-rinsed polyethylene bottle(s) as "FAR, SH 1130." Check with the laboratory for bottle requirements for the analysis requested.

1. Filter the sample and fill the sample bottle to the shoulder directly from the capsule filter or other filter assembly.
2. Add reagent-grade HNO3 to preserve the sample to pH <2. Do not substitute HCl for HNO3.

5.6.3.G Gross Radioactivity

Label 1-L acid-rinsed polyethylene bottle(s) as "FAR" (filtered sample) for the gross alpha and the gross beta analyses and with the appropriate laboratory schedule number. The laboratory schedule requested depends on the concentration of total dissolved solids in the sample.

1. Filter the sample.
2. Fill the sample bottle to the shoulder directly from the capsule filter or other filter assembly. .
3. Add reagent-grade HNO3 to preserve the sample to pH <2. Do not substitute HCl for HNO3.

5.6.3.H Tritium

Label a 1-L bottle as "RUR" and add the appropriate laboratory code. (High density polyethylene bottles are preferred; a glass bottle with a polyseal cap may be used. Refer to National Water Quality Technical Memorandum 92.04.)

 

Do not place tritium samples near watches or other devices with luminescent dials. Do not store sample near tritium sources (for example, glowing clocks, watches, signs).

  Do not field rinse sample bottles.
1. Fill bottle with raw, untreated sample. The bottle should be dry before being filled. It might be desirable to flush the bottle with a filtered, inert gas such as argon or nitrogen before leaving for the field--consult with the laboratory.
2. Do not allow the bottle to overflow when filling with sample.
3. Leave a slight headspace in the bottle to allow for expansion of the sample.
4. Cap the bottle securely and tape the cap to prevent it from working loose during shipping.
5. Record the date and time of sampling on the bottle label and ASR form.

For tritium samples--Keep luminescent devices far from sample collection, handling, or storage areas.

 

 

 

5.6.3.I Radon-222

Modify the sample collection setup to collect raw samples for radon analysis. Sample is collected in a glass vial containing liquid-scintillation solution (obtain radon kit from NWQL). Label radon-222 samples as "RUR-SV" and add the laboratory code.

Do not write on or put any labels on the side of the radon vial.

 

 

 

Precautions are needed when collecting samples for radon analyses to prevent introducing gas bubbles into the sample and to prevent the sample from degassing.

 

Use insulated sample tubing to prevent warming of the sample.

  Inspect sample tubing to determine whether gas bubbles are forming inside the tubing or whether any air is being drawn into the sample at any connection.
  Dislodge bubbles in sample tubing by striking the tubing firmly with a blunt object.
  Tighten connections to help prevent entrainment of air.
  To reduce degassing during sample collection, create back pressure by partially closing the valve on the radon-collection unit.
1. Collect the radon sample into a syringe directly from the pump discharge or other sampling device.
2. Rinse the syringe as follows: Insert the glass syringe needle through the septum port with the collection-unit valve partially closed. Close the valve further until there is sufficient back pressure to create an almost effortless withdrawal of sample into the syringe. Fill the syringe partially, withdraw it from the septum and invert (needle up). Eject the water to waste. Repeat at least once.
3. With the syringe plunger completely depressed (no air or water in the syringe barrel) and after the final rinse, reinsert the needle through the septum. Withdraw about 15 mL of sample into the syringe barrel slowly; avoid suction and degassing.
4. Withdraw the needle, invert the syringe (needle up), and eject the sample slowly until 10 mL remain in the syringe.
5. Tip the syringe needle downward, and insert the needle tip into the mineral oil and down to the bottom of the radon sample vial.
6. Inject the entire sample slowly. Remove the syringe and cap the vial firmly. Record the date and the exact time of sample collection on the top of the cap (do not write on or put a label on the vial).
7. Shake the vial for 30 seconds after injecting the sample. Repack it in the shipping tube and cap the tube. Complete the ASR form, wrap it around the tube, secure with a rubber band, and place the tube into a sealable plastic bag. Ship to the NWQL immediately by overnight delivery. Do not ship samples on a Friday. Do not ship radon-222 samples in coolers.

Do not allow air to contact samples for radon analysis.

 

 

 

5.6.3.J Tritium/Helium-3 (3H/3He)

Refer to NWQL Technical Memorandum 97.04S for detailed information and instructions.

Do not allow air to contact samples for 3H/3He analysis.

 

 

Water samples for the helium determination are collected, in duplicate, in special pinch-off copper tubes that hold approximately 40 mL of water. These tubes are supplied through the NWQL but are the property of the analyzing laboratory (Lamont-Doherty Earth Observatory). Unused tubes must be returned to Lamont-Doherty. Notify NWQL of the number of tubes returned.

1. Set up the sampling apparatus. Collect the sample in duplicate. These samples must also be accompanied by 500 mL of sample in a glass or high-density polyethylene bottle for tritium analysis. In addition, archive a duplicate 500-mL sample.
  a. Remove plastic caps covering the ends of the copper tube. Handle the copper tube with care--damage to the ends from scratches and bends can prevent a good vacuum connection and thus compromise sample integrity.
  b. Connect the copper tube to the submersible pump discharge tubing.
    Do not exceed 5 ft between the pump discharge and the sampling apparatus.
    The connector to the pump-discharge tubing can be plastic, rubber, or metal, but it must be airtight.
    Clear plastic tubing is preferred to allow visual inspection for possible air bubbles.
    Airtight connections must not come loose when back pressure is applied during closing of copper tubes. Secure the connections with stainless steel hose clamps, placing clamps approximately 1 in. from the end of the copper tube. (Do not damage ends of copper tube.)
    Attach a small valve to the discharge end of the copper tube and insert clear plastic (TygonTM) tubing to allow the operators to check for air bubbles (NWQL Technical Memorandum 97.04S provides photographs and diagrams).
2 Begin flow of the sample to the copper tube, checking for bubbles.
  a. Hold the copper tube at a 45-degree angle, discharge up, while flushing bubbles from the system.
  b. Forcibly tap the entire sample tubing and aluminum track that holds the copper tube with a socket wrench or other blunt object to dislodge gas bubbles. This typically takes about 1 minute.
3. Apply backpressure to prevent formation of gas bubbles. Close valve until flow is reduced, tap channel, then close completely.
4. Seal the sample container.
  a. Position the copper tube in approximate center of the pinch-off clamp.
  b. Use a socket wrench to close the bolts on the pinch-off clamps, starting at the discharge end.
  c. Turn the bolts in successive order (back and forth approximately four times until firmly closed) so that the blades of the pinch-off clamp close approximately evenly.
  d. Center the copper tube between the blades again, and close the pinch-off clamp on the inflow end as described in step c above.
  e. Double check to ensure that all bolts are tight.
5. Repeat procedure above (1-4) to collect a duplicate sample.
6. Disconnect the copper tube from the pump discharge tubing and remove the backpressure valve from the discharge end of the copper tube, taking care not to scratch or otherwise damage the ends.
  a. If the sample is saline, acidic, or otherwise corrosive, wash the ends of the copper tube with DIW.
  b. Take care not to bend the ends of the sealed copper tube.
  c. Do not replace the plastic caps on the ends of the copper tube.
  d. Place label onto the aluminum track of each sample--do not write on the tube. Include station identification, date, and time of sampling on the label.
7. Prepare the sample(s) for shipment. Fill out a form or submit a letter to the analyzing laboratory with the following information:
  Unique site identification number(s).
  Date and time of sample collection.
  Ground-water temperature at time of collection and recharge temperature, if known.
  Estimated or known tritium concentration.
  Estimated or known elevation of the recharge area for the sample.
  General description of the hydrogeologic environment, location, and well-construction information.
  Information regarding possible tritium contamination of the sample.
 

Name, fax number, E-mail address, and phone number of the person responsible for the sample(s).

5.6.3.K Chlorofluorocarbons (CFC-11, CFC-12, CFC-113)

Chlorofluorocarbons (CFCs) can be analyzed in samples of ground water, surface water, or air. The information that follows is summarized from the Office of Water Quality/Office of Ground Water Technical Memorandum 95.02 and, while generally applicable to all media, is focused toward collection of ground water for CFC analysis. Before collecting CFC samples, review this memorandum for more detailed information on field sampling, site selection, analysis requests, and applications of CFC analyses in environmental investigations. Important information for collecting CFC samples is also provided in Busenberg and Plummer (1992) and Plummer and others (1993).

Do not allow air to contact samples for CFC analysis.

 

 

The collection, transport, and storage of water samples without contamination is critical in order to reliably age-date waters containing CFCs.
Contact the CFC laboratory (USGS, Reston, Va.) to arrange for training and to plan for CFC sampling. Remain in contact with the CFC laboratory throughout the planning, sampling, and analysis phases.
1. Schedule CFC sampling several months in advance, using the CFC request form (fig. 5-4), in order to
  Reserve the sampling equipment. Special sampling apparatus is loaned to USGS study personnel who will collect water samples for CFC analyses and who have had the training required for CFC sampling.
  Obtain the required number of borosilicate-glass ampoules (five ampoules per ground-water sample). The borosilicate-glass ampoules are supplied by the USGS CFC laboratory and are included in the price of analysis.
  Obtain cylinders of compressed ultra-pure nitrogen gas and welding-grade oxygen.
2. Assemble additional tools, materials, and equipment needed (table 5-8; Office of Water Quality/Office of Ground Water Technical Memorandum 95.02).
  Discuss the type and modification of pumping equipment with the CFC laboratory. USGS personnel should contact the laboratory in Reston, Va., by phone (703-648-5847), fax (703-648-5832), or E-mail (cfc@usgs.gov). The CFC laboratory will provide the guidance needed to prevent cross-contamination of samples collected contemporaneously for CFCs and trace elements.
  Replace the sample discharge tubing with the appropriate tubing (table 5-8): refrigeration-grade copper or aluminum tubing are recommended; nylon tubing can be used but should be analyzed by the CFC laboratory before use; chromatographic-grade 304 stainless steel tubing also can be used. Do not use fluorocarbon polymer tubing.
3. Collect an equipment blank and submit for CFC analysis at least 1 week before collecting environmental samples.
  For CFC sampling only, the matrix of the blank typically is water collected from a relatively deep well tapping an aquifer recharged only with pre-1940 water.
  Ship the sample by overnight delivery to the USGS CFC Laboratory in Reston, Va.
  Review QC-data results and make equipment changes accordingly. Proceed with sampling only if the equipment blanks are clean.
4. Collect preliminary samples for radiochemicals and VOC analyses at the well(s) selected (steps 5a through c below) and review the results before collecting or submitting the samples for CFC analyses. Select wells that are open hole or have metal or threaded (not glued) PVC casing. Do not submit samples for CFC analyses that contain the following:
  Hazardous radioactive substances.
  More than 0.5 mg/L of any CFC or other halocarbon (vinyl chloride, methyl chloride, methyl chloroform, methyl bromide, methylene chloride, chloroform, trichloroethylene, carbon trichloride, and tetrachloroethylene).
5. Sample collection.
  a. Review sampling plans.
  b. Measure the water level and prepare to purge the well.
  c. Purge the well, recording field measurements for specific electrical conductance, pH, temperature, and dissolved oxygen at approximately 5-minute intervals (NFM 4.2; NFM 6).
    The measurement of dissolved-oxygen concentrations is particularly important.
    Record on field forms whether or not a hydrogen sulfide odor is detected. Collect the samples needed for other organic and inorganic analyses. Collection of samples for analysis of VOCs and tritium is strongly recommended when collecting samples for CFC analysis.
  d. Set up the CFC sampling equipment. Replace the sample tubing, if necessary, or make adjustments to direct the sample flow through copper lines or other appropriate tubing.
    CFC samples must not come in contact with air. Use only the CFC apparatus specifically designed for this purpose that is supplied by the CFC Laboratory.
    Review the precautions listed for radon sampling (section 5.6.3.I) to prevent sample degassing and air contact.
  e. Collect ground water into a borosilicate-glass ampoule--you will need five ampoules per well. Flame-seal the ampoule. Repeat for each of the remaining ampoules. (It takes about 5 minutes to collect sample into an ampoule.)
    Follow the instructions given in training for site evaluation and use of the CFC sampling apparatus.
   

After the last ampoule is filled and sealed, measure specific electrical conductance, pH, temperature, and dissolved oxygen and record the measurements on the field form.

6. Determine, if feasible, concentrations of hydrogen sulfide and methane, tritium/helium-3, and dissolved nitrogen and argon gases.
  a. Measure hydrogen sulfide concentrations (Baedecker and Cozzarelli, 1992) if there is a hydrogen sulfide odor.
  b. Collect sample for tritium/helium-3 age dating.
  c. Collect sample for analysis of dissolved nitrogen and argon gases (to determine temperature of recharge water) and arrange for dissolved gas analyses. USGS personnel should contact the Northeastern Region Common Use Laboratory at (703) 648-6234.
7. Label the sample ampoules as directed by the CFC Laboratory. Pack the ampoules in the boxes supplied by the CFC Laboratory (place padding on top of the ampoules). Be sure to pack the field form (fig. 5-4) in the box with the ampoules.
8. Ship the CFC samples to arrive the next day, from Monday through Thursday. Do not ship on a Friday; there must be no weekend overlay of samples.