FDA

Enforcement Report

The FDA Enforcement Report is published weekly by the Food and Drug Administration, U.S. Public Health Service, Department of Health and Human Services. It contains information on actions taken in connection with agency regulatory activities.

July 14, 1999                                              99-28

RECALLS AND FIELD CORRECTIONS: FOODS -- CLASS I

________ PRODUCT Spiced Dry Tofu in 8 ounce plastic containers. Recall #F-540-9. CODE None. MANUFACTURER Water Lilies Food, Inc., Brooklyn, New York. RECALLED BY Manufacturer, by visit on March 11, 1999, and by press release. Completed field correction (relabeling) resulted from sample analysis and follow-up by the New York State Department of Agriculture and Markets. DISTRIBUTION New York. QUANTITY 10 cartons (36 containers per carton) were distributed. REASON Product contained undeclared peanuts. ________ PRODUCT DawnMist Mouth Rinse, packaged in 2 ounce and 4 ounce white plastic containers. Recall #F-543-9. CODE All lots labeled as "Made in Malaysia". MANUFACTURER HI-CITY Manufacturing, Penang, Malaysia. RECALLED BY Donovan Industries, Inc., Tampa, Florida, by letters on March 29, 1999, and April 19, 1999. Firm-initiated recall ongoing. DISTRIBUTION Nationwide. QUANTITY Approximately 1,200 cases (60 bottles per case) were distributed over the past six months in the United States. REASON Product may be contaminated with Burkholderia cepacia. ________ PRODUCT Morningland Dairy Raw Milk Mild Cheddar Cheese, packaged in random sizes ranging in weight from .50 pounds to 20 pounds. Recall #F-544-9. CODE C179. MANUFACTURER Moringland Farm, Inc., Mountain View, Missouri. RECALLED BY Manufacturer, by telephone on June 11, 1999, and by press release. Firm-initiated recall complete. DISTRIBUTION Michigan, Pennsylvania, Georgia, North Carolina, Washington state, Arkansas, Missouri, Alabama, Florida, Tennessee. QUANTITY 228 pounds were distributed. REASON Product is contaminated with Listeria monocytogenes. ________ PRODUCT "Sunshine Sprouts" Clover Sprouts, in 4-6 ounce cardboard containers. Recall #F-545-9. CODE None. MANUFACTURER Sunshine Sprouts (sprout grower), West Palm Beach, Florida. RECALLED BY Manufacturer, by letter sent on June 7, 1999, and by telephone June 8, 1999, and by press release. Firm-initiated recall complete. DISTRIBUTION Florida. QUANTITY Firm estimates none remains on the market. REASON The sprouts may have been contaminated with Salmonella s. Typhimurium. The seeds used to manufacture these sprouts were from the same lot of seeds that was implicated in a Salmonella s. Typhimurium outbreak in Colorado.

RECALLS AND FIELD CORRECTIONS: FOODS -- CLASS III

________ PRODUCT 30 Different baked goods: a) Honey Maid Graham Pie Crust;72 pkgs/case; Net Wt 6 oz. b) Old Fashioned Ginger Snaps;6 pkgs/case; Net Wt.16 oz; c) Family Favorite Oatmeal Cookies; 12 pkgs/case; Net Wt.14 oz d) Flavor Crisp Snack Cracker Entertainment Pack; 6 pkgs/case; Net Wt. 24 oz. e) Reduced Fat Triscuit Baked Whole Wheat Wafers: 12 pkgs/case; Net Wt. 8.5 oz f) Honey Maid Honey Grahams; 12 pkgs/case; Net Wt.16 oz.and 6 pkgs/case; Net Wt.32 oz. g) Triscuit Baked Whole Wheat Wafers; 12 pkgs/case; Net Wt. 9.5 oz and 12 pkgs/case; Net Wt.13 oz. h) Low Sodium Triscuit Baked Whole Wheat Wafers; 12 pkgs/case; Net Wt. 9.5 oz. i) Premium Soup and Oyster Crackers; 12 pkgs/case; Net Wt. 10 oz. k) Nabisco Grahams; 12 pkgs/case; Net Wt. 16 oz. l) Low Fat Honey Maid Cinnamon Graham; 12 pkgs/case; Net Wt. 16 oz. m) Barnum's Animal Crackers; 24 pkgs/case; Net Wt. 2.125 oz. n) Original Premium Saltine Crackers; 12 pkgs/case; Net Wt 8 oz. o) Sweet Crisper Chocolate Crispy Baked Snacks; 6 pkgs/case; Net Wt. 7 oz. p) Sweet Crisper Honey Crispy Baked Snacks; 6 pkgs/case; Net Wt. 7 oz. q) Holiday Oreo Chocolate Sandwich Cookies; 6 pkgs/case; Net Wt. 40 oz. r) Ritz Value Pack; 6 pkgs/case; Net Wt. 40 oz. s) Oreo Chocolate Sandwich Cookies; 12 pkgs/case; Net Wt. 16 oz. t) Triscuit Thin Crisps Baked Whole Wheat Snacks Original; 12 pkgs/case; Net Wt. 8.5 oz. u) Triscuit Thin Crisps Whole Wheat Snacks French Onion Flavor; 12 pkgs/case; Net Wt. 8.5 oz. v) Chips Ahoy! Real Chocolate Chip Cookies; 12 pkgs/case; Net Wt. 12 oz. and 12 pkgs/case; Net Wt. 18 oz. w) Ritz Bits Sandwiches made with Real Cheese; 12 pkgs/case; Net Wt. 10.5 oz. x) Ritz Bits Sandwiches with Peanut Butter; 12 pkgs/case; Net Wt. 10.5 oz. y) Ritz Crackers; 6 pkgs/case; Net Wt. 32 oz. z) Deli-Style Rye Triscuit Baked Whole Wheat Wafers; 12 pkgs/case; Net Wt. 9.5 oz. aa) Snackwell's Fat Free Devil's Food Cookie Cakes; 12 pkgs/case; Net Wt. 6.75 oz. bb) Snackwell's Creme Sandwich Cookies; 12 pkgs/case; Net Wt. 7.75 oz.. Recall #F-509/535-9. CODE Old Fashioned Ginger Snaps, 9068AG, Family Favorite Oatmeal Cookies, 9060AK, 9061AK Flavor Crisps Snack Cracker Entertainment Pack, 9017RP, 9018RP, 9019RP, 9020RP, 9027RP Reduced Fat Triscuit Baked Whole Wheat Wafers, 9061AN, 9062AN Honey Maid Honey Grahams, 9059AX, 9060AV, 9060AX, 9061AV, 9061AX, 9062AV, 9062AX Honey Maid Honey Grahams, 9053AV, 9072AV Triscuit Baked Whole Wheat Wafers, 9060AN, 9062AN, 9080AN Low Sodium Triscuit Baked Whole Wheat Wafers, 9058AN Triscuit Baked Whole Wheat Wafers, 9066AN, 9068AN Premium Soup and Oyster Crackers, 9069AX Nabisco Grahams, 9070AG Low Fat Honey Maid Cinnamon Graham, 9051AM, 9052AM, 9053AM Barnum's Animals Crackers, 8360AP, 8364AP Original Premium Saltine Crackers, 9045AP Sweet Crisper Chocolate Crispy Baked Snacks, 9008AK, 9009AK, 9013AK, 9014AK Sweet Crisper Honey Crispy Baked Snacks, 9063AK Holiday Oreo Chocolate Sandwich Cookies, 9045RP, 9046RP, 9047RP Ritz Value Pack, 9045RP, 9053RP, 9055RP, 9057RP, 9059RP Oreo Chocolate Sandwich Cookies, 9066AG, 9068AG Triscuit Thin Crisps Baked Whole Wheat Snacks Original, 9062AN Triscuit Thin Crisps Baked Whole Wheat Snacks French Onion Flavor, 9054AN, 9055AN Chips Ahoy! Real Chocolate Chip Cookies, 9072AX Chips Ahoy! Real Chocolate Chip Cookies, 9067AH, 9069AH, 9070AH, 9071AH Ritz Bits Sandwiches made with Real Cheese, 9067AW, 9068AW, 9076BD, 9077BD, 9078BD Ritz Bits Sandwiches with Peanut Butter, 9074AW, 9075AW, 9076AW, 9080AW, 9081AW, 9081BD, 9082BD Ritz Crackers, 9034AV, 9040AV Deli -Style Rye Triscuit Baked Whole Wheat Wafers, 9062AN Snackwell's Fat Free Devil's Food Cookie Cakes, 9011BG Snackwell's Crème Sandwich Cookies, 9080KE MANUFACTURER Multiple sites. RECALLED BY Nabisco, Inc., East Hanover, New Jersey, by E-Mail on November 17 and 23, 1998. Firm-initiated recall complete. DISTRIBUTION Nationwide QUANTITY Approximately 183,300 cases (all products) were distributed; firm estimates none remains on the market. REASON Products were adulterated with organic volatile solvent rendering them unfit for food. ________ PRODUCT Isomil, Ready-To-Feed Liquid Infant Formula, in 1 quart (32 fluid ounce) metal cans. Recall #F-537-9. CODE The Batch Numbers: 52003RE and 52007RE, are ink-printed on the bottom of each can. The shipping cases are identified with the Code numbers: 52003RE00 and 52007RE00, which are ink-jetted on the side panels of each case. MANUFACTURER Ross Products Division, Abbott Laboratories, Casa Grande, Arizona. RECALLED BY Ross Products Division, Abbott Laboratories, Columbus, Ohio, by telephone beginning April 28, 1999, followed by fax. Firm- initiated recall complete. DISTRIBUTION Nationwide. QUANTITY 17,821 cases (106,926 cans) were distributed. REASON Product was held in cans with a low level of can lid defects allowing for post-processing contamination.

RECALLS AND FIELD CORRECTIONS: FOODS -- CLASS III

________ PRODUCT Elizabeth's Candies Milk and Dark Chocolate Deluxe Assortment, in 7 ounce boxes. Recall #F-536-9. CODE All lots. MANUFACTURER Elizabeth's Candies, Kenmore, New York. RECALLED BY Manufacturer, by visit or by telephone beginning on May 6, 1999. Completed field correction (relabeling) resulted from sample analysis and follow-up by the New York State Department of Agriculture and Markets. DISTRIBUTION New York. QUANTITY Approximately 300-400 boxes were distributed. REASON Candy is misbranded because the labeling lists only "nuts" as an ingredient but does not specify which nuts are in the product. ________ PRODUCT Crown brand: a) Radix Dioscorea (Chinese Yam) in rigid plastic containers net weight 12 ounces (340 g); b) Dried Almond in flexible plastic packages, new weight 6 ounces (170 g). Recall #F-538/539-9. CODE None. MANUFACTURER Wing Tak Fook Ltd., Sheung Wan, Hong Kong, China. RECALLED BY Murray International, Inc., Brooklyn, New York, (importer/distributor), by letter on February 23, 1999. Completed field correction (relabeling) resulted from sample analysis and follow-up by the New York State Department of Agriculture and Markets. DISTRIBUTION New York. QUANTITY a) 32 cases (50 packages per case); b) 15 cases (100 packages per case) were distributed. REASON Products contained undeclared sulfites.

RECALLS AND FIELD CORRECTIONS: DRUGS -- CLASS II

________ PRODUCT Medicap Pharmacy Peri-Colace Stool Softener With Laxative containing 100 mg of Docusate Sodium and 30 mg of Casanthranol per softgel, 100 softgels per bottle. NDC number is 56050-254- 52. Recall #D-280-9. CODE Lot #4985A EXP 7/00. MANUFACTURER Time-Cap Labs., Inc., Farmingdale, New York (Manufacturer of the bulk aspirin which was repacked and then mislabeled as Peri- Colace). RECALLED BY Apothecary Products, Inc., Burnsville, Minnesota (repacker), by telephone and fax on June 8, 1999. Firm-initiated recall complete. DISTRIBUTION Undetermined. QUANTITY 12 bottles in 2 boxes of 6 bottles each were distributed. REASON Mislabeling - Bottles actually contain 325 mg aspirin tablets.

RECALLS AND FIELD CORRECTIONS: DRUGS -- CLASS III

________ PRODUCT Calcionate Syrup, in 16 fluid ounce (pint) bottles, OTC calcium supplement. Recall #D-281-9. CODE Lot #802921. EXP 7/00. MANUFACTURER Hi-Tech Pharmacal Company, Inc., Amityville, New York. RECALLED BY MANUFACTURER, by letter dated June 14, 1999. Firm-initiated recall ongoing. DISTRIBUTION Georgia. QUANTITY 3,564 units were distributed. REASON Presence of white precipitate and cloudiness; calcium is precipitating out of formulation.

RECALLS AND FIELD CORRECTIONS: BIOLOGICS -- CLASS II

________ PRODUCT a) Red Blood Cells; b) Platelets; c) Fresh Frozen Plasma; d) Plasma, Cryoprecipitate Reduced; e) Cryoprecipitate AHF; f) Cryoprecipitate AHF, Pooled; g) Platelets Pheresis; h) Granulocytes Pheresis; i) Recovered Plasma For Manufacturing Non-injectable Use; j) Recovered Plasma. Recall #B-905/914-9. CODE Unit numbers: a) 69715, 77277, 142284, 220419, 245114, 276682, 294841, 314070, 238484, 344834, 434670, 444967, 478032, 486373, 0577657; b) 220419, 276682, 294841, 314070, 328484, 434670; c) 294841, 344834; d) 0577657; e) 0577657; f) 9012582; g) 352213, 357294, 363313, 368209, 374208, 376026, 393606, 408547, 414029, 421271, 432676, 437417, 486501, 510584, 0526605, 0579141, 442603, 447706, 465356, 470611, 481553, 498518, 502800, 0533989, 0537182, 0545385, 0567818, 0587688, 0611765, 0559879 (parts 1 and 2), 0549243 (parts 1 and 2), 0571891 (parts 1 and 2); h) 0597507; i) 338981, 381491; j) 444967, 478032, 486373, 69715, 77277, 142284, 148659, 220419, 245114, 276682, 314070, 328484, 434670. MANUFACTURER Memorial Blood Centers of Minnesota, Minneapolis, Minnesota. RECALLED BY Manufacturer, by letters dated November 6, 20, 23, 24, 1998, December 9 and 17, 1998, January 11, 1999, and February 1, 1999. Firm-initiated recall complete. DISTRIBUTION California, Connecticut, Florida, Massachusetts, Minnesota, North Carolina, Pennsylvania, Switzerland. Firm-initiated recall complete. QUANTITY a) 15 units; b) 6 units; c) 2 units; d) 1 unit; e) 1 unit; f) 1 unit; g) 32 units; h) 1 unit; i) 2 units; j) 13 units were distributed. REASON Blood products tested negative for the antibody to the human immunodeficiency virus type 1 (anti-HIV-1), but were collected from a donor who previously tested repeatedly reactive for anti- HIV-1, Western blot indeterminate. ________ PRODUCT a) Red Blood Cells, Leukoreduced; b) Red Blood Cells. Recall #B-927/928-9. CODE Unit numbers 03FL74884 and 03LG10030. MANUFACTURER American Red Cross, Gainesville, Georgia. RECALLED BY American Red Cross, Atlanta, Georgia, by telephone on October 16, 1998, and by letter dated October 22, 1998. Firm-initiated recall complete. DISTRIBUTION Georgia. QUANTITY 1 unit of each component was distributed. REASON Blood products were collected from a donor who had traveled to an area designated as endemic for malaria. ________ PRODUCT a) Red Blood Cells; b) Recovered Plasma. Recall #B-931/932-9. CODE Unit #12GZ01593. MANUFACTURER American Red Cross Blood Services, Durham, North Carolina. RECALLED BY American Red Cross Blood Services, Charlotte, North Carolina, by letter dated July 17, 1998. Firm-initiated recall complete. DISTRIBUTION North Carolina Switzerland. QUANTITY 1 unit of each component was distributed. REASON Blood products tested negative for anti-HCV but was collected from a donor who previously tested reactive for Hepatitis C. _________ PRODUCT Platelets, Pheresis. Recall #B-933-9. CODE Unit #12GW16747 (split unit 1 and 2). MANUFACTURER American Red Cross Blood Services, Charlotte, North Carolina. RECALLED BY Manufacturer, by telephone on August 27, 1998, and by letter dated September 1, 1998. Firm-initiated recall complete. DISTRIBUTION North Carolina. QUANTITY 1 split unit was distributed. REASON Blood products had high platelet counts. ________ PRODUCT Red Blood Cells. Recall #B-934-9. CODE Unit #03LR53822. MANUFACTURER American Red Cross Blood Services, Atlanta, Georgia. RECALLED BY Manufacturer, by telephone on December 24, 1998, and by letter dated January 14, 1999. Firm-initiated recall complete. DISTRIBUTION Georgia. QUANTITY 1 unit was distributed. REASON Blood product was exposed to unacceptable shipping temperatures. ________ PRODUCT a) Red Blood Cells; b) Platelets; c) Fresh Frozen Plasma; d) Recovered Plasma for Manufacturing Non-injectable; Products; e) Recovered Plasma. Recall #B-938/942-9. CODE Unit numbers: a) 42FJ73931, 42FZ75158, 42FW43232, 42K89757, 42S58161, 42FG52747; b) 42FJ73931, 42FW43232, 42K89757, 42S58161, 42FG52747; c) 42K89757, 42S58161; d) 42FZ75158; e) 42FJ73931, 42FG52747, 42FW43232. MANUFACTURER American Red Cross Blood Services, Cleveland, Ohio. RECALLED BY Manufacturer, by letter dated December 7, 1998, and January 27, 1999. Firm-initiated recall complete. DISTRIBUTION Ohio and Massachusetts, California. QUANTITY a) 6 units; b) 5 units; c) 2 units; d) 1 unit; e) 3 units were distributed. REASON Blood products tested negative for anti-HCV, but was collected from a donor who previously tested reactive for anti-HCV. ________ PRODUCT Allergenic Extract - GS Ragweed Mix. Recall #B-943-9. CODE Lot #P1-235-5X3. MANUFACTURER Greer Laboratories, Inc., Lenoir, North Carolina. RECALLED BY Manufacturer, by telephone on April 6, 1999. Firm-initiated recall complete. DISTRIBUTION Florida. QUANTITY 1 vial was distributed. REASON Allergenic extract labeled with the incorrect concentration. ________ PRODUCT Platelets. Recall #B-944-9. CODE Unit #1558773. MANUFACTURER Central Kentucky Blood Center (CKBC), Lexington, Kentucky. RECALLED BY Manufacturer, by telephone on April 27, 1999, and by letter dated June 8, 1999. Firm-initiated recall complete. DISTRIBUTION Kentucky. QUANTITY 1 unit was distributed. REASON Platelets were prepared from a whole blood unit which was collected from a donor who had taken aspirin within three days of donation. _________ PRODUCT Red Blood Cells. Recall #B-945-9. CODE Unit #22402-9453. MANUFACTURER Blood Services, Inc., Las Vegas, Nevada. RECALLED BY Blood Systems,Inc., Scottsdale, Arizona, by telephone on October 3, 1998. Firm-initiated recall complete. DISTRIBUTION Nevada. QUANTITY 1 unit was distributed. REASON Blood product was not quarantined after receiving information concerning a post donation illness. ________ PRODUCT a) Red Blood Cells; b) Recovered Plasma. Recall #B-946/947-9. CODE Unit numbers: FK03851, FK03852, FK03853, FK03854, FK03856, FK02983, FK02984, FK02985, FK02986, FK02987, FK02988, FK02989, FK02990, FK02991, FL01488, FL01489, FL01490, FL01491, FL01492, FL01493, FL01494, FL01495, FL01497, FL01498, FL01499, FL01500, FL01501, FK04657, FK04659, FK04660, FX00499, FX00451, FX00453, FX00454, FX00455, FX00456, FX00458, FX00459, FX00463, FX00469, FX00471, FX00472, FX00476, FX00484, FX00490, FX00491, FX00495, FX00497, FX00348, FX00349, FX00350, FX00351, FX00352, FX00353, FX00354, FX00356, FX00357, FX00358, FX00360, FX00057, FX00058, FX00059. MANUFACTURER Central Illinois Community Blood Bank, Springfield, Illinois. RECALLED BY Manufacturer, by telephone on May 25, 1999, and by letter and fax on May 27, 1999, and by fax on May 26, 1999. Firm-initiated recall complete. DISTRIBUTION Illinois, California, Switzerland. QUANTITY 62 units were distributed. REASON Testing for syphilis not performed in accordance with MANUFACTURER's insert in that the age of the sample exceeded the 48 hour requirement. ________ PRODUCT Red Blood Cells. Recall #B-964-9. CODE Unit #40GT19792. MANUFACTURER American Red Cross Blood Services, Peoria, Illinois. RECALLED BY Manufacturer, by letter dated May 28, 1999. Firm-initiated recall complete. DISTRIBUTION Illinois. QUANTITY 1 unit was distributed. REASON Blood product was collected from a donor who reported travel to an area designated as endemic for malaria. ________ PRODUCT Red Blood Cells. Recall #B-965-9. CODE Unit Numbers: 4615125, 4615134, 4615178, 4615179, 4615228, 4615251, 4615329, 4615414, 4615447, 4615897, 4615919, 4616093, 4616241, 4616433, 4616909, 4616937, 4617183, 4617245, 4617299, 4617302, 4617535, 4617537, 4617575, 4617665, 4617994, 4618241, 4618297, 4618316, 4618318, 4618322, 4618373, 4618493, 4618567. MANUFACTURER Department of the Navy, Naval Hospital Blood Bank, Great Lakes, Illinois. RECALLED BY Manufacturer, by letter dated June 7, 1999. Firm-initiated recall complete. DISTRIBUTION New Jersey. QUANTITY 33 units were distributed. REASON Blood products were collected from donors who had not been properly questioned regarding malarial risk. ________ PRODUCT Platelets. Recall #B-966-9. CODE Unit #32KQ04827, MANUFACTURER American Red Cross Blood Services, Madison, Wisconsin. RECALLED BY Manufacturer, by telephone on November 12, 1998. Firm-initiated recall complete. DISTRIBUTION Wisconsin. QUANTITY 1 unit was distributed. REASON Blood product was collected from a donor who previously tested positive for human leukocyte antigen. ________ PRODUCT a) Red Blood Cells; b) Plasma. Recall #B-974/975-9. CODE Unit #11T49807. MANUFACTURER American Red Cross Blood Services, St. Louis, Missouri. RECALLED BY Manufacturer, by telephone on August 12 and 13, 1998, and by letter dated August 21, 1998. Firm-initiated recall complete. DISTRIBUTION New Jersey and California. QUANTITY 1 unit of each component was distributed. REASON Blood products tested positive for HbsAg, but were collected from a donor who previously tested reactive for Hepatitis B. ________ PRODUCT a) Red Blood Cells; b) Plasma. Recall #B-976/977-9. CODE Unit #11LY02861. MANUFACTURER American Red Cross Blood Services, St. Louis, Missouri. RECALLED BY Manufacturer, by telephone on December 31, 1997, and by letter dated January 6, 1998. Firm-initiated recall complete. DISTRIBUTION Missouri and California. QUANTITY 1 unit of each component was distributed. REASON Blood products tested negative for the antibody to the human immunodeficiency virus type 1/2 (HIV-1/2), but were collected from a donor who previously tested reactive for the antibody. ________ PRODUCT a) Red Blood Cells; b) Platelets; c) Recovered Plasma. Recall #B-991/993-9. CODE Unit #1589203. MANUFACTURER LifeSource, Glenview, Illinois. RECALLED BY Manufacturer, by letter and telephone on March 11, 1999. Firm- initiated recall complete. DISTRIBUTION Illinois and Switzerland. QUANTITY 1 unit of each component was distributed. REASON Blood products were collected from a donor with a history of cancer.

RECALLS AND FIELD CORRECTIONS: BIOLOGICS -- CLASS III

________ PRODUCT Sterile Diluent for Allergenic Extract in 9.0 ml vials, Normal Saline with Phenol. Recall #B-640-9. CODE Lot #5527C. MANUFACTURER Allergy Laboratories, Inc., Oklahoma City, Oklahoma. RECALLED BY Manufacturer, by telephone on March 2, 1999, and by letter dated March 15, 1999. Firm-initiated recall complete. DISTRIBUTION Nationwide. QUANTITY 2,800 vials were distributed. REASON Discoloration was observed in some vials of sterile diluent marketed for use with allergenic extracts. ________ PRODUCT Red Blood Cells. Recall #B-901-9. CODE RH24133, RH24135, RH24136, RH24137, RH24138, RH24139, RH24142, RH24143, RH24144, RH24145, RH24148, RH24149, RH24150, RH24151. MANUFACTURER Riverview Hospital Association, Wisconsin Rapids, Wisconsin. RECALLED BY Manufacturer, by telephone on April 28 and 29, 1998. Firm- initiated recall complete. DISTRIBUTION Wisconsin and California. QUANTITY 14 units were distributed. REASON Blood products were labeled with extended expiration dates. ________ PRODUCT Red Blood Cells and Fresh Frozen Plasma. Recall #B-902/903-9. CODE Unit #32KJ41602. MANUFACTURER American Red Cross Blood Services, Madison, Wisconsin. RECALLED BY Manufacturer, by letters dated October 29, 1998 or November 6, 1998. Second notices were sent on either December 11 or 18, 1998. Firm-initiated recall complete. DISTRIBUTION Wisconsin. QUANTITY 1 unit of each component was distributed. REASON Blood products were collected from a donor whose arm inspection and skin disease status had not been accurately documented. ________ PRODUCT Allergenic Extract - Mold Mix. Recall #B-915-9. CODE Lot M0145-83-1F1. MANUFACTURER Greer Laboratories, Inc., Lenoir, North Carolina. RECALLED BY Manufacturer, by telephone on April 6, 1999. Firm-initiated recall complete. DISTRIBUTION New Jersey. QUANTITY 3 vials were distributed. REASON Allergenic extracts were labeled with an extended expiration date. ________ PRODUCT Platelets, Pheresis. Recall #B-917-9. CODE Unit #0023932 (split product) MANUFACTURER American Red Cross Blood Services, Atlanta, Georgia. RECALLED BY Manufacturer, by telephone on October 9, 1998. Firm-initiated recall complete. DISTRIBUTION Georgia. QUANTITY 1 split unit was distributed. REASON Blood products which were in transit greater than 24 hours were redistributed without swirl testing being performed. ________ PRODUCT Platelets and Fresh Frozen Plasma. Recall #B-929/930-9. CODE Unit #12FR54188. MANUFACTURER American Red Cross Blood Services, Durham, North Carolina. RECALLED BY American Red Cross Blood Services, Charlotte, North Carolina, by letters on January 30, 1999, and March 2, 1999, and by telephone on January 25, 1999. Firm-initiated recall complete. DISTRIBUTION North Carolina and Puerto Rico. QUANTITY 1 unit of each component was distributed. REASON Blood products were collected from a donor taking the drug Propecia. ________ PRODUCT a) Platelets; b) Platelets, Pheresis. Recall #B-935/936-9. CODE Contact FDA, Center for Biologics Evaluation and Research, Office of Compliance (301) 827-6220 for individual unit numbers recalled. PRIVATE MANUFACTURER American Red Cross Blood Services, Madison, Wisconsin. RECALLED BY Manufacturer, by letter dated May 13, 1998, or June 12, 1998, or July 10, 1998. Firm-initiated recall complete. DISTRIBUTION Illinois. QUANTITY a) 28 units; b) 27 units were distributed. REASON Blood products were stored at unacceptable temperatures. ________ PRODUCT Platelets, Pheresis. Recall #B-937-9. CODE Unit #03FK17429. MANUFACTURER American Red Cross Blood Services, Atlanta, Georgia. RECALLED BY Manufacturer, by telephone on February 5, 1999, and by letter dated February 9, 1999. Firm-initiated recall complete. DISTRIBUTION Georgia. QUANTITY 1 unit was distributed. REASON Blood product was distributed without having performed a platelet count. ________ PRODUCT Red Blood Cells and Recovered Plasma. Recall #B-948/949-9. CODE Unit Numbers: FK04658, FX00500, FX00457, FX00461, FX00465, FX00466, FX00467, FX00470, FX00473, FX00474, FX00477, FX00478, FX00480, FX00481, FX00482, FX00483, FX00485, FX00486, FX00487, FX00488, FX00492, FX00493, FX00494, FX00496, FX00498, FX00355. MANUFACTURER Central Illinois Community Blood Bank, Springfield, Illinois. RECALLED BY Manufacturer, by telephone on May 25, 1999. Follow-up letter were mailed and faxed on May 27, 1999. Firm-initiated recall complete. DISTRIBUTION Illinois, California, Switzerland. QUANTITY 26 units were distributed. REASON Blood products were incorrectly tested for syphilis. ________ PRODUCT Red Blood Cells, Leukocytes Reduced. Recall #B-963-9. CODE Unit numbers: 11GQ04972, 11FE32552, 11FG35193, 11FG35190, 11FG35192, 11FG40607, 11LY18691, 11GJ00819, 11GW04886, 11GW04891, 11GY01388, 11FM38222, 11LV23710. MANUFACTURER American Red Cross Blood Services, St. Louis, Missouri. RECALLED BY Manufacturer, by telephone on August 14, 1998, and by letter dated August 28, 1998. Firm-initiated recall complete. DISTRIBUTION Missouri. QUANTITY 13 units were distributed. REASON Red Blood Cells, leukoreduced by filtration greater than five days past the time of collection, were distributed for transfusion. ________ PRODUCT a) Red Blood Cells; b) Red Blood Cells, Leukocytes Reduced; c) Platelets; d) Platelets, Pooled; e) Fresh Frozen Plasma; f) Recovered Plasma for Manufacturing Injectables g) Recovered Plasma for Manufacturing Non-Injectables. Recall #B-967/973-9. CODE Unit numbers: a) 6605080, 6605083, 6605084, 6605085, 6605086, 6605088, 6605090, 6605093, 6605095, 6605099, 6605109, 6605110, 6605111, 6605115 b) 6605112, 6605116 c) 6605080, 6605083, 6605084, 6605085, 6605086, 6605088, 6605092, 6605095, 6605098, 6605099, 6605104, 6605106 d) P65682, P65687, P65685, P65690 e) 6605080 f) 6605085, 6605086, 6605088, 6605099, 6605104, 6605105, 6605106, 6605115, 6605116, 6605117 g) 6605093. MANUFACTURER Carter Blood Center, Fort Worth, Texas. RECALLED BY Manufacturer, by telephone on December 3, 1998, followed by letter dated December 16, 1998, and by fax on December 7, 1998. Firm-initiated recall complete. DISTRIBUTION Texas, California, Switzerland. QUANTITY a) 14 units; b) 2 units; c) 12 units; d) 4 units; e) 1 unit; f) 10 units; g) 1 unit. REASON Blood products were collected from donors who were offered monetary payment for donating and were mislabeled as volunteer donations. ________ PRODUCT a) Red Blood Cells; b) Platelets; c) Fresh Frozen Plasma. Recall #B-988/990-9. CODE Unit #4303896. MANUFACTURER LifeSource, Glenview, Illinois. RECALLED BY Manufacturer, by telephone on January 14, 1999. Firm-initiated recall complete. DISTRIBUTION Illinois. QUANTITY 1 unit of each component was distributed. REASON Blood products were collected from a donor who received a tissue transplant (surgery) within 1 year of donation.

RECALLS AND FIELD CORRECTIONS: DEVICES -- CLASS II

________ PRODUCT CVIS Insight Imaging Catheter for intraluminal coronary interventional procedures; 2.6French/30 MHz and 2.6French/40 MHz, Model C3020 and C3005, indicated for ultrasound examination of the coronary intravascular pathology only. Intravascular ultrasound is indicated in-patients who are candidates for transluminal coronary interventional procedures. Recall #Z-1051/1052-9. CODE UPN/Material #H749C30200, Catalog #C3020, Lots: All Lots UPN/Material #H749C30050, Catalog #C3005, Lots: All Lots. MANUFACTURER Boston Scientific Corporation, San Jose, California. RECALLED BY Manufacturer, by letter on June 15, 1999. Firm-initiated recall ongoing. DISTRIBUTION Nationwide and international. QUANTITY 44,841 units were distributed. REASON The product may separate from the proximal segment. ________ PRODUCT Midmark 73018 Attachment Accessories for the Midmark Magnum General Table, hydraulic, sold as kits and are designed for use in the support of patients during orthopedic surgery procedures. Recall #Z-1053-9. CODE Serial Numbers: TFY1010 through TFY1025. MANUFACTURER Midmark Corporation, Versailles, Ohio. RECALLED BY Manufacturer, by telephone on June 14-15, 1999, and by letter mailed on June 24, 1999. Firm-initiated recall ongoing. DISTRIBUTION New Jersey, Florida, Michigan, Louisiana, Minnesota, Texas, Missouri, District of Columbia. QUANTITY 14 units. REASON Defects of table accessories cause unintended motion during orthopedic surgery. ________ PRODUCT Henley Healthcare Fluidotherapy Unit, air-fluidized beds employing dry heat that is intended to treat local pain, range of motion and blood flow insufficiency in the body extremities: a) Henley Healthcare Fluidotherapy, Model 110D; b) Henley Healthcare Fluidotherapy, Model 110DE, Recall #Z-1054/1055-9. CODE a) Serial Numbers 2-00001 - 2-00369 b) Serial Numbers 2-00001 - 2-00006. MANUFACTURER Henley Healthcare, Inc., Sugarland, Texas. RECALLED BY Manufacturer, by letter dated January 15, 1998. Firm-initiated recall ongoing. DISTRIBUTION Nationwide and international. QUANTITY a) 369 units; b) 6 units were distributed. REASON Excessively long metal bolts securing metal legs to the wooden base of the device protruded too far into the housing, contacting the electrical connection to the heater after collapsing a metal heater schroud, causing 13 amps of electrical current to be delivered to the metal legs. ________ PRODUCT Doctors Review System, OBLink software releases prior to 9.0. Recall #Z-1056-9. CODE All units. MANUFACTURER Digisonics, Inc., Houston, Texas. RECALLED BY Manufacturer, by telephone, fax, and letter on March 29, 1999. Firm-initiated field correction ongoing. DISTRIBUTION Nationwide. QUANTITY 39 units were distributed. REASON Y2K software error may produce an incorrect estimated delivery date (EDD). ________ PRODUCT QTEST(R) STREP, an IVD kit for P.O.C. testing to detect Group A streptococcus, using a sample throat swab, Catalog 494776 - 40 tests/kit; 494780 - 80 tests/kit. Recall #Z-1058-9. CODE SERIAL NOS. 5648, 5663, 5664, 5673, 5665. MANUFACTURER Becton Dickinson Microbiology Systems, Cockeysville, Marland. RECALLED BY Manufacturer, by letter on June 10, 1999, and by letter on June 22, 1999. Firm-initiated recall ongoing. DISTRIBUTION Nationwide. QUANTITY 1,044 kits were distributed. REASON The Vol-pak pouch sealer failed to provide an adequate seal, and the Quali-pak leak tester failed to detect the inadequate seal. ________ PRODUCT DIRECTIGEN 1-2-3 GROUP A STREP TEST, DEVICE LISTING 702228, an IVD kit used by hospitals for detection of Group A streptococcus using a sample of throat swabs or bacterial colonies. Recall #Z-1059-9. CODE CATALOG # 852540, Serial Nos. 5638, 5662, 5670. MANUFACTURER Becton Dickinson Microbiology Systems, Cockeysville, Maryland. RECALLED BY Manufacturer, by letter on June 10, 1999, and by letters beginning June 22, 1999. Firm-initiated recall ongoing. DISTRIBUTION Washington state, Tennessee, Florida, Georgia, Connecticut, Utah, Kentucky, New Jersey, Pennsylvania, Indiana, Alaska, Illinois, Minnesota, Arkansas, California, Virginia, District of Columbia. QUANTITY 1,230 kits were distributed. REASON The VolPak pouch sealer failed to provide an adequate seal, and the QualiPak leak tester failed to detect the inadequate seal. An inadequate seal may lead to false negative results. ________ PRODUCT Quest Myocardial Protection System (MPS), a sterile, single-use, prescription device: a) Model No. 5001101; b) Model No. 5001101 NE; c) Model No. 5001102. Recall #Z-1061/1063-9. CODE a) Lot No. 9306.05K; b) Lot No. 9373.08K; c) Lot Nos. 9267.08J, 290.03K, 9340.01L, 9357.03L, 9394-03L, 9403.04L, 404.07L, 9405.09L, 9415.03M, 9434.05M. MANUFACTURER Quest Medical, Inc., (QMI), an Atrion company, Allen, Texas. RECALLED BY Manufacturer, by telephone on May 6, 1999, followed by letter. Firm-initiated recall ongoing. DISTRIBUTION Nationwide. QUANTITY 3,430 units were distributed. REASON The delivery sets may develop a leak in the main pump cassette during use. ________ PRODUCT Ampilcor HBV Monitor(tm) Kit, an IVD test that utilizes PCR nucleic acid amplification and DNA hybridization for the detection and quantitative measurement for Hepatitis B viral DNA in human serum or plasma: a) Catalog No. 87521/0757160 (Domestic), Amplicor HBV Monitor(tm) Test "For Research Use Only"; b) Catalog No. 83271/0757152 (International), Amplicor HBV Monitor(tm) Test "For In Vitro Diagnostic Use". Recall #Z-1064/1065-9. CODE Lot numbers: a) Lot No. 9803, EXP 5/99; b) Lot No. 9804, EXP 5/99. MANUFACTURER AB Sangtec Medical, Bromma, Sweden. RECALLED BY Roche Molecular Systems, Inc., Somerville, New Jersey, by field correction dated February 8, 1999. Firm-initiated recall complete. DISTRIBUTION New Jersey, Pennsylvania, Maryland, North Carolina, Indiana, Michigan, Texas, California. QUANTITY 91 kits were distributed. REASON The IS values for the standards are low which results in HBV/IS OD ratios which are high and outside of the kit-specified acceptance criteria causing invalid runs.

RECALLS AND FIELD CORRECTIONS: DEVICES -- CLASS III

________ PRODUCT Steris Process Biological Indicator; the kit includes 20 biological indicators (B.stearothermophilus), 20 vials filled with growth medium, and a Certificate of Performance and Instruction Card, to be used in conjunction with STERIS Chemical Monitoring Strips. The product is shipped frozen. Recall #Z-1048-9. CODE Catalog Numbers: S3032, S3036, and S3036-2. The lot number is a 6-digit number. All lots which end with '613', '815', or '816' are subject to this recall. The expiration dates and ending lot numbers are as follows: '613'/exp 7/99; '815'/exp 8/99; and '816'/exp 3/99. MANUFACTURER Steris Corporation, Mentor, Ohio. RECALLED BY Manufacturer, by memorandum dated August 26, 1998. Firm- initiated recall complete. DISTRIBUTION Nationwide and international. QUANTITY 8,553 kits were distributed. REASON Product failed stability testing at three (3) months. The spore counts dropped below labeled specification. ________ PRODUCT Kodak Digital Science Medical Film EIR-11 Laser Imaging Film, intended for secondary imaging of radiographs, packed 150 sheets per box. Recall #Z-1057-9 CODE Lot/Serial Nos. 4597-0128-003-11/-12/-13/-14/-15. MANUFACTURER Eastman Kodak Company, Health Imaging Division, Rochester, New York. RECALLED BY Manufacturer, by telephone on April 26 and 27, 1999. Firm- initiated recall ongoing. DISTRIBUTION Arizona, Kansas, North Carolina, Nebraska, Ohio, Pennsylvania. QUANTITY 32 cases (450 sheets per case) were distributed. REASON The film was exposed to light during the rewind process, and has an elevated background fog. The normal level of fog has a Dmin of 0.18, while the identified product has a Dmin of 0.30. ________ PRODUCT ARCHITECT Estradiol Reagent Kit, list 6C22-20 - 4 x 100 tests and list 6C22-25 - 1 x 100 tests; an in-vitro diagnostic Chemiluminescent Microparitcle Immunoassay (CMIA) for the quantitative determination of estradiol in human serum and plasma. Recall #Z-1060-9. CODE List 6C22-20, lot 49994M100 and List 6C22-25, lot 49994M101. MANUFACTURER Abbott Laboratories, Abbott Park, Illinois. RECALLED BY Manufacturer, by letter dated June 29, 1999. Firm-initiated recall ongoing. DISTRIBUTION Nationwide and international. QUANTITY 245 400-test kits and 68 100-test kits were distributed. REASON The Architect Estradiol Reagents may show under-recovery of estradiol with undiluted patient specimens.

ALLS AND FIELD CORRECTIONS: VETERINARY PRODUCTS -- CLASS II

________ PRODUCT Agri-Cillin, Penicillin G Procaine Injectable Suspension, USP, Injectable Antibiotic in Aqueous Suspension 300,000 units per mL, OTC. Recall #V-070-9. CODE Lot #8501-02 EXP 12/00. MANUFACTURER Norbrook Laboratories Ltd., Newry, Northern Ireland. RECALLED BY Agri Laboratories, Ltd., St. Joseph, Missouri, by letter dated May 26, 1999. Firm-initiated recall ongoing. DISTRIBUTION Nebraska, South Dakota, California, Idaho, Oklahoma, Ohio, Texas, Louisiana, Kansas, New York, Colorado. QUANTITY 3,618 bottles were distributed. REASON Product contains glass, unidentified foreign material fiber clumps and hair. END OF ENFORCEMENT REPORT FOR JULY 14, 1999. BLANK PAGES MAY FOLLOW.

[FDA HOME PAGE]