IA #77-01, SEPTEMBER 01, 1998, IMPORT ALERT# 77-01, "DETENTION WITHOUT
PHYSICAL EXAMINATION OF EAR CANDLES", ATTACHMENT - 10/10/07


This import alert represents the Agency's current guidance to FDA field personnel regarding the manufacturer(s
)
and/or product(s) at issue.  It does not create or confer any rights for, or on any person, and does not opera
te to
bind FDA or the public.


TYPE OF ALERT: Detention Without Physical Examination (DWPE)

PRODUCT:       Ear Candles (a.k.a., Ear cones and Auricular candles).

PRODUCT CODE:  77[][][]

PROBLEMS:   Unapproved/unfiled premarket application
            Inadequate directions/warning
            Not Device Registered
            Not Device Listed
            No 510(k)
            Danger to Health

OASIS CHARGE CODES: NO PMA; FALSE; DANGEROUS;
                    REGISTERED; NOT LISTED; NO 510(k)

PAF:           AAP - No PMA filed
               REG - No 510(k) filed, no registration & listing
               LBL - Labeling violations

PAC:           82008

COUNTRY:         See attachment.

MANUFACTURER/
SHIPPER:         See attachment.

CHARGES:         The following charges should be used when encountering these devices:

               The articles are subject to refusal of admission pursuant to Section 801(a)(3) in that:

               (1)  The articles appear to be Class III devices  under Section 513(f) without an
                    approved application for premarket approval pursuant to Section 515(a).

                 [Adulteration - Section  501(f)(1)(B)]
                 (OASIS CHARGE CODE = NO PMA)

               (2)  The labeling for these articles appears to be false or misleading.

                 [Misbranding - Section 502(a)]
                 (OASIS CHARGE CODE = FALSE)

               (3)  The articles appear to be dangerous to health when used in the dosage or manner,
                    or with the frequency or duration, prescribed, recommended, or suggested in
                    the labeling thereof.

                 [Misbranding - Section 502(j)]
                 (OASIS CHARGE CODE = DANGEROUS)

               (4)  It appears the device establishment is subject to registration under 510(c) and the
                    initial distributor has not registered as required by 21 CFR 807.20(a)(4).

                  [Misbranding - Section 502(o)]
                     (OASIS CHARGE CODE = REGISTERED)

               (5)  It appears the devices are not included in a list  required by Section 510(j).

                 [Misbranding - Section 502(o)]
                 (OASIS CHARGE CODE = NOT LISTED)

               (6)  It appears that a notice or other information respecting the devices was not
                    provided to FDA, as required by Section 510(k), and the device was not found to
                    be substantially equivalent to a predicate device.

                 [Misbranding - Section 502(o)]
                 (OASIS CHARGE CODE = NO 510(k))

RECOMMENDING OFFICE:  Center for Devices and Radiological Health, Office of Compliance, DOEII,
                      Dental, ENT & Ophthalmics Branch, HFZ-331.

REASON FOR ALERT:     The Center for Devices and Radiological Health (CDRH), has determined that "Ear
                      Candles" are medical devices as defined by Section 201(h) of the Federal Food,
                      Drug, and Cosmetic Act (The Act).  An Ear Candle is a hollow wax cylinder
                      (about ten inches long) intended to remove excess ear wax.  This is accomplished
                      by lighting the top of the candle-like product, and allowing it to create a vacuum to
                      draw wax and other impurities from the ear.

                 The product labeling is false and misleading in that  there is no validated scientific
                 evidence to support the efficacy of the product for its intended use.  Also, the label
                 of the product contains inadequate directions for use since adequate directions
                 cannot be written for the product's purported use.  CDRH considers the product to
                 be dangerous when used according to its labeling, since the use of a lit candle in
                 the proximity of a person's face would carry a high risk of causing potentially
                 severe skin/hair burns and middle ear damage.

               Additionally, there has been no premarket notification filed (510(k)) for these products
               and the products appear to have been manufactured in establishments not duly
               registered or listed with the Agency (FDA).

GUIDANCE:      When districts encounter shipments of ear candles a recommendation for detention
               without physical examination and background information should be forwarded to
               DIOP, HFC-172 for inclusion on the attachment to this alert.  DIOP will coordinate
               with CDRH for concurrence.

FOI:           No purging required.

KEYWORDS:      Ear Candle, Auricular, Ear Cones, Misbranding,
               Adulteration, Wax

PREPARED BY:     Rosa K. Forehand, DIOP, (301) 443-6553
               Ronald L. Swann, CDRH, (301) 594-4613

DATE LOADED INTO
FIARS:         9/1/98




               ATTACHMENT TO IMPORT ALERT #77-01         10/10/07




Debbie Boehlen                                    5/23/05
  dba Queen Bee Ear Candles
26 Colebeck Drive
Welland, Canada L3C5B5
FEI# 3004910801


Kencayd Consulting (a.k.a. Candela Earcandles)
Ear Candle Supplies & Training
555 Ridley Dr.
Victoria, British Columbia, V9C 1K3
Canada
FEI#: 3000189108

Ms. Sheila Stevens (d.b.a. Superior Ear Candles
Division of J&P Holdings)
Box 137
Parson, British Columbia
Canada
FEI#: 3000189110

Europe Cosmetiques
3471 Bld. Thimens
St. Laurent, Quebec H4R 1V5
Canada
FEI#: 3000122926

Aural
Box 1236
Kaslo, British Columbia
VOG 1MO
Canada
FEI#: 3001448775

OTOSAN                             Ear Care Cone (Ear Candles)
Via Balzell                               77[][][]
75-47100 Forli, Italy                        5/27/99
FEI# 3002706370


SDC Company                          Ear Candles
A1, Block 1, Wen Hua Garden                  77[][][]
Shennan Rd. East                        8/26/04
Guangdong, China
FEI# 3004431076


Suzhou Helio Technology Co., LTD.            Ear Candles
Block 29, Dongjing Industrial Workshop,        77[][][]
Loufeng East Park, SIP                    10/10/07
Jiangsu, China
FEI #3006486221









Oxford Medical Supplies, LTD.             Ear Candles
Unit 11 Horcott Road                      77[][][]
Fairford, Gloucestershire                    10/10/07
United Kingdom Gl74BX,
FEI #3003498942


Helio Supplies Pty., Ltd.                    Ear Candles
93 Shepherd Street                      77[][][]
Chippendale NSW 2008                    10/10/07
Australia
FEI   #3003619088