The km22 RS Energy Calibration using the RSMON corrections improves the results greatly on a run-by-run basis. However, there are a few phototubes that seem to get worse or give not what we want/expect/wish/pray for. What we want is to have the gain vs run as flat as possible. I have created 9 sets of gain files based on the following runs ranges: 49965:50013, 50014:50029, 50030:50057, 50058:50080, 50081:50120, 50121:50142, 50143:50172, 50173:50191, 50192:50218. To see the plots of gain vs run # for each tube look at the following website: http://www.phy.bnl.gov/~benjil/meetings/video/200303/ -> 2002.ps shows the gains before the RSMON corrections were applied -> 2003.ps shows the gains with the RSMON corrections applied -> rdrun.ps shows the Kentaro's run-by-run factors vs. run # **These tubes were much worse on a run-by-run basis with the **RSMON correction L2 S16 E1: runs 50121-50142: gain is low runs 50173-50218: gain is high L8 S22 E2: runs 50030:50057: gain is high runs 50143:50172: gain is low runs 50030:50057 - 50143:50172 there is an overall constant downward trend. **These tubes were improved on a run-by-run basis with the **RSMON correction L1 S1 E2: erratic L1 S4 E2: runs 50030:50057 - 50058:50080 there is a large jump in gain. L1 S7 E1: erratic L1 S10 E1: runs 50058:50080: high runs 50121:50142: low L1 S13 E2: runs 50058:50080: much to high runs 50081:50120: high runs 50058:50080 - 50173:50191 there is an approx. constant downward trend L1 S19 E2: runs 50121:50142,50173:50191,50192:50218: the gains are about constant for these 3 blocks of runs and the value of the gains for the other 6 blocks are of a different constant. L3 S4 E2: The first 6 blocks of runs are about constant with a gain of ~18. And the last 3 blocks are all ~16. So there is a discreet change between 50121:50142 - 50143:50172.